Crosstalk in x-ray framing cameras: Effect on voltage, gain, and timing (invited).

We present evidence that electromagnetic crosstalk between independent strips in gated x-ray framing cameras can affect relative gains by up to an order of magnitude and gate arrival times up to tens of picoseconds when strip separation times are less then ∼1 ns. Crosstalk is observed by multiple methods, and it is confirmed by direct measurements of voltage on the active surface of the detector and also by indirect voltage monitors in routine operation. The voltage measurements confirm that crosstalk is produced not only in the active regions of the microchannel plate, but also along the entire input path of the voltage pulses.

[1]  C. J. Anderson Electrical properties of an inputoutput cable for Josephson applications , 1982 .

[2]  Jeong-Cheol Lee,et al.  Empirical equations on electrical parameters of coupled microstrip lines for crosstalk estimation in printed circuit board , 2001 .

[3]  R Tommasini,et al.  Spectroscopic determination of temperature and density spatial profiles and mix in indirect-drive implosion cores. , 2007, Physical review. E, Statistical, nonlinear, and soft matter physics.

[4]  A. Feller,et al.  Crosstalk and reflections in high-speed digital systems , 1965, AFIPS '65 (Fall, part I).

[5]  J. Kilkenny High speed proximity focused X-ray cameras , 1991 .

[6]  H. Kroger,et al.  Behavior of Short Pulses on Tightly Coupled Microstrip Lines and Reduction of Crosstalk by Using Overlying Dielectric , 2007, IEEE Transactions on Advanced Packaging.

[7]  G. Rochau,et al.  Measurement and modeling of pulsed microchannel plate operation (invited). , 2008, The Review of scientific instruments.

[8]  A study of crosstalk applied to ultra-high-speed parallel lossy microstrip and strip lines with shielding effectiveness , 1999, 1999 International Symposium on Electromagnetic Compatibility (IEEE Cat. No.99EX147).

[9]  C Cerjan,et al.  Radiation hardening of gated x-ray imagers for the National Ignition Facility (invited). , 2010, The Review of scientific instruments.

[10]  Neal R. Pederson,et al.  Gated x-ray detector for the National Ignition Facility , 2006 .

[11]  Nobuhiko Izumi,et al.  X-ray imaging in an environment with high-neutron background on National Ignition Facility , 2011, Optical Engineering + Applications.

[12]  S R Nagel,et al.  Methods for characterizing x-ray detectors for use at the National Ignition Facility. , 2012, The Review of scientific instruments.

[13]  P Bell,et al.  A hardened gated x-ray imaging diagnostic for inertial confinement fusion experiments at the National Ignition Facility. , 2010, The Review of scientific instruments.