Acceptance testing prone stereotactic breast biopsy units.
暂无分享,去创建一个
When the Mammography Quality Standards Act becomes law in October, 1994, stereotactic breast biopsy units may require yearly physicist calibration. Upright stereotactic units can be easily tested using conventional mammography procedures and a gelatin phantom containing simulated calcifications, but prone units are difficult to assess because of the under-table tube configuration. The two current manufacturers of these units have made different design decisions which affect each unit's calibration. There are a number of important distinctions between screening and prone biopsy units. For the two currently available prone units, a pronounced heel effect makes ion chamber position critical. Focal spot measurements are particularly difficult on one unit because there is no light field. The fixed grid on the other unit must be tested with a flood film. Physicists who inspect these units before their clinical use should be aware of variations needed by this equipment for specific acceptance tests.