PC-based instrumentation for electrodermal activity measurement

A PC-based EDA measuring system is presented. The system is composed of a laptop with a PCMCIA DAQ-card running LabVIEW® software, a small front-end with a dual op-amp IC and a few passive components, and three skin-surface electrodes. The electrode system gives a monopolar measurement below the measuring electrode regardless of the electrode sizes, unless very small. Usage of the system is demonstrated by measurements from a mental stress experiment on 17 volunteers. There was a significant correlation (R=0.51, p>0.001) between the self-assessed stress-level and the EDA response frequency. The system allows easy on-site customization in software of measuring parameters, signal-quality monitoring and non-linearity detection in real time. We believe that the most suited use for the system is for stationary experimental purposes where this flexibility is desired. The system is easy to reproduce by engineers interested in doing EDA research.

[1]  S. Grimnes Impedance measurement of individual skin surface electrodes , 1983, Medical and Biological Engineering and Computing.

[2]  Sverre Grimnes,et al.  Measuring depth depends on frequency in electrical skin impedance measurements , 1999 .

[3]  Sverre Grimnes,et al.  Embedded instrumentation for skin admittance measurement , 2008, 2008 30th Annual International Conference of the IEEE Engineering in Medicine and Biology Society.

[4]  Sverre Grimnes,et al.  Electrical measurement of sweat activity. , 2008, Physiological measurement.

[5]  W. Boucsein Electrodermal activity, 2nd ed. , 2012 .

[6]  Sverre Grimnes,et al.  Noise properties of the 3-electrode skin admittance measuring circuit , 2009 .