Considerations for Thermal Injury Analysis for RF Ablation Devices

Background: The estimation of lesion size is an integral part of treatment planning for the clinical applications of radiofrequency ablation. However, to date, studies have not directly evaluated the impact of different computational estimation techniques for predicting lesion size. In this study, we focus on three common methods used for predicting tissue injury: (1) iso-temperature contours, (2) Cumulative equivalent minutes, (3) Arrhenius based thermal injury. Methods: We created a geometric model of a multi-tyne ablation electrode and simulated thermal and tissue injury profiles that result from three calculation methods after 15 minutes exposure to a constant RF voltage source. A hybrid finite element technique was used to calculate temperature and tissue injury. Time-temperature curves were used in the assessment of iso-temperature thresholds and the method of cumulative equivalent minutes. An Arrhenius-based formulation was used to calculate sequential and recursive thermal injury to tissues. Results: The data demonstrate that while iso-temperature and cumulative equivalent minute contours are similar in shape, these two methodologies grossly over-estimate the amount of tissue injury when compared to recursive thermal injury calculations, which have previously been shown to correlate closely with in vitro pathologic lesion volume measurement. In addition, Arrhenius calculations that do not use a recursive algorithm result in a significant underestimation of lesion volume. The data also demonstrate that lesion width and depth are inadequate means of characterizing treatment volume for multi-tine ablation devices. Conclusions: Recursive thermal injury remains the most physiologically relevant means of computationally estimating lesion size for hepatic tumor applications. Iso-thermal and cumulative equivalent minute approaches may produce significant errors in the estimation of lesion size.

[1]  D B Plewes,et al.  Quantifying tissue damage due to focused ultrasound heating observed by MRI , 1999, Magnetic resonance in medicine.

[2]  Uyen D. Nguyen,et al.  Thermal modeling of lesion growth with radiofrequency ablation devices , 2004, Biomedical engineering online.

[3]  E. J. Woo,et al.  Thermal—electrical finite element modelling for radio frequency cardiac ablation: Effects of changes in myocardial properties , 2000, Medical and Biological Engineering and Computing.

[4]  Henriques Fc,et al.  Studies of thermal injury; the predictability and the significance of thermally induced rate processes leading to irreversible epidermal injury. , 1947 .

[5]  K. Henle,et al.  TIME‐TEMPERATURE RELATIONSHIPS FOR HEAT‐INDUCED KILLING OF MAMMALIAN CELLS * , 1980, Annals of the New York Academy of Sciences.

[6]  Dieter Haemmerich,et al.  Automatic control of finite element models for temperature-controlled radiofrequency ablation , 2005, Biomedical engineering online.

[7]  J. R. Roti Roti,et al.  Time-temperature conversions in biological applications of hyperthermia. , 1980, Radiation research.

[8]  John G. Webster,et al.  Hepatic bipolar radio-frequency ablation between separated multiprong electrodes , 2001, IEEE Transactions on Biomedical Engineering.

[9]  Francis A. Duck,et al.  Physical properties of tissue : a comprehensive reference book , 1990 .

[10]  Dieter Haemmerich,et al.  Sequential Activation of a Segmented Ground Pad Reduces Skin Heating During Radiofrequency Tumor Ablation: Optimization via Computational Models , 2008, IEEE Transactions on Biomedical Engineering.

[11]  S. Baldwin,et al.  A Heat Transfer Model of Thermal Balloon Endometrial Ablation , 2001, Annals of Biomedical Engineering.

[12]  Steven L. Jacques,et al.  Nonlinear finite-element analysis of the role of dynamic changes in blood perfusion and optical properties in laser coagulation of tissue , 1996 .

[13]  S. Weininger,et al.  Factors to consider in a risk analysis for safe surface temperature , 2005, IEEE Symposium on Product Safety Engineering, 2005..

[14]  Hong Cao,et al.  Three-dimensional finite-element analyses for radio-frequency hepatic tumor ablation , 2002, IEEE Trans. Biomed. Eng..

[15]  John G. Webster,et al.  Thermal dose versus isotherm as lesion boundary estimator for cardiac and hepatic radio-frequency ablation , 2003, Proceedings of the 25th Annual International Conference of the IEEE Engineering in Medicine and Biology Society (IEEE Cat. No.03CH37439).

[16]  Xiaoyi Min,et al.  Comparison of methods in approximation of blood flow in finite element models on temperature profile during RF ablation , 1998, Proceedings of the 20th Annual International Conference of the IEEE Engineering in Medicine and Biology Society. Vol.20 Biomedical Engineering Towards the Year 2000 and Beyond (Cat. No.98CH36286).

[17]  P D Wolf,et al.  Effect of Skin Electrode Location on Radiofrequency Ablation Lesions: An In Vivo and a Three‐Dimensional Finite Element Study , 1998, Journal of cardiovascular electrophysiology.

[18]  P. Savard,et al.  A finite element model for radiofrequency ablation of the myocardium , 1994, IEEE Transactions on Biomedical Engineering.

[19]  H. H. Pennes Analysis of tissue and arterial blood temperatures in the resting human forearm. , 1948, Journal of applied physiology.

[20]  W. Dewey,et al.  Thermal dose determination in cancer therapy. , 1984, International journal of radiation oncology, biology, physics.

[21]  D. Schutt,et al.  Effects of variation in perfusion rates and of perfusion models in computational models of radio frequency tumor ablation. , 2008, Medical physics.

[22]  John G. Webster,et al.  Using ANSYS for three-dimensional electrical-thermal models for radio-frequency catheter ablation , 1997, Proceedings of the 19th Annual International Conference of the IEEE Engineering in Medicine and Biology Society. 'Magnificent Milestones and Emerging Opportunities in Medical Engineering' (Cat. No.97CH36136).

[23]  R. W. Lau,et al.  The dielectric properties of biological tissues: II. Measurements in the frequency range 10 Hz to 20 GHz. , 1996, Physics in medicine and biology.

[24]  Fernando Hornero,et al.  Thermal-electrical modeling for epicardial atrial radiofrequency ablation , 2004, IEEE Transactions on Biomedical Engineering.

[25]  John C. Bischof,et al.  The Kinetics of Thermal Injury in Human Renal Carcinoma Cells , 2005, Annals of Biomedical Engineering.

[26]  P. Wolf,et al.  Temperature-controlled and constant-power radio-frequency ablation: what affects lesion growth? , 1999, IEEE Transactions on Biomedical Engineering.

[27]  Hong Cao,et al.  Guidelines for predicting lesion size at common endocardial locations during radio-frequency ablation , 2001, IEEE Transactions on Biomedical Engineering.

[28]  Yu-Chi Lai,et al.  Lesion size estimator of cardiac radiofrequency ablation at different common locations with different tip temperatures , 2004, IEEE Transactions on Biomedical Engineering.

[29]  John G. Webster,et al.  Effects of changes in electrical and thermal conductivities on radiofrequency lesion dimensions , 1997, Proceedings of the 19th Annual International Conference of the IEEE Engineering in Medicine and Biology Society. 'Magnificent Milestones and Emerging Opportunities in Medical Engineering' (Cat. No.97CH36136).

[30]  A. Eriksson,et al.  Influence of electrical and thermal properties on RF ablation of breast cancer: is the tumour preferentially heated? , 2005, Biomedical engineering online.

[31]  John G. Webster,et al.  Hepatic radiofrequency ablation with internally cooled probes: effect of coolant temperature on lesion size , 2003, IEEE Transactions on Biomedical Engineering.

[32]  Fernando Hornero,et al.  A cooled intraesophageal balloon to prevent thermal injury during endocardial surgical radiofrequency ablation of the left atrium: a finite element study , 2005, Physics in medicine and biology.

[33]  Patrick D. Wolf,et al.  Effect of electrode contact on lesion growth during temperature controlled radio frequency ablation , 1998, Proceedings of the 20th Annual International Conference of the IEEE Engineering in Medicine and Biology Society. Vol.20 Biomedical Engineering Towards the Year 2000 and Beyond (Cat. No.98CH36286).

[34]  Michael C. Kolios,et al.  The effects of dynamic optical properties during interstitial laser photocoagulation. , 2000, Physics in medicine and biology.

[35]  J. G. Webster,et al.  Hepatic bipolar radiofrequency ablation creates coagulation zones close to blood vessels: A finite element study , 2003, Medical and Biological Engineering and Computing.

[36]  D. Wyman,et al.  Dynamic modeling of interstitial laser photocoagulation: Implications for lesion formation in liver in vivo , 1999, Lasers in surgery and medicine.

[37]  P. Wolf,et al.  A Three-Dimensional Finite Element Model of Radiofrequency Ablation with Blood Flow and its Experimental Validation , 2000, Annals of Biomedical Engineering.

[38]  S. B. Field,et al.  An Introduction to the Practical Aspects of Clinical Hyperthermia , 1990 .

[39]  Fernando Hornero,et al.  Esophageal Temperature During Radiofrequency‐Catheter Ablation of Left Atrium: A Three‐Dimensional Computer Modeling Study , 2006, Journal of cardiovascular electrophysiology.

[40]  Jay Gopalakrishnan A Mathematical Model for Irrigated Epicardial Radiofrequency Ablation , 2002, Annals of Biomedical Engineering.

[41]  S. Labonte,et al.  Numerical model for radio-frequency ablation of the endocardium and its experimental validation , 1994, IEEE Transactions on Biomedical Engineering.