Aberration measurement in HRTEM: implementation and diagnostic use of numerical procedures for the highly precise recognition of diffractogram patterns.

The precise characterisation of the instrumental imaging properties in the form of aberration parameters constitutes an almost universal necessity in quantitative HRTEM, and is underlying most hardware and software techniques established in this field. We focus in this paper on the numerical analysis of individual diffractograms as a first preparatory step for further publications on HRTEM aberration measurement. The extraction of the defocus and the 2-fold astigmatism from a diffractogram is a classical pattern recognition problem, which we believe to have solved in a near-optimum way concerning precision, speed, and robustness. The newly gained measurement precision allows us to resolve fluctuations of the defocus and the 2-fold astigmatism and to assess thereby the optical stability of electron microscopes. Quantitative stability criteria are elaborated, which may serve as helpful guidelines for daily work as well as for microscope acceptance tests.

[1]  Stephan Uhlemann,et al.  A spherical-aberration-corrected 200 kV transmission electron microscope , 1998 .

[2]  D. Van Dyck,et al.  Ultimate resolution and information in electron microscopy: general principles , 1992 .

[3]  R. Wade,et al.  Electron microscope transfer functions for partially coherent axial illumination and chromatic defocus spread , 1977 .

[4]  C. Jia,et al.  High-resolution imaging with an aberration-corrected transmission electron microscope. , 2002, Ultramicroscopy.

[5]  O. Krivanek A method for determining the coefficient of spherical aberration from a single electron micrograph , 1976 .

[6]  A. Kirkland,et al.  “Indirect” High-Resolution Transmission Electron Microscopy: Aberration Measurement and Wavefunction Reconstruction , 2004, Microscopy and Microanalysis.

[7]  A. Thust,et al.  Numerical correction of lens aberrations in phase-retrieval HRTEM , 1996 .

[8]  Kazuo Ishizuka,et al.  Coma-free alignment of a high-resolution electron microscope with three-fold astigmatism , 1994 .

[9]  F. Zemlin,et al.  Coma-free alignment of high-resolution electron microscopes with the aid of optical diffractograms , 1978 .

[10]  A. Thust High-resolution transmission electron microscopy on an absolute contrast scale. , 2009, Physical review letters.

[11]  A. Thust,et al.  Maximum-likelihood method for focus-variation image reconstruction in high resolution transmission electron microscopy , 1996 .

[12]  Atomic structure of a Sigma=3, {111} twin-boundary junction in a BaTiO3 thin film , 1999 .

[13]  William H. Press,et al.  Numerical recipes , 1990 .

[14]  K. Dierksen,et al.  Determination of Image Aberrations in High-resolution Electron Microscopy using Diffractogram and Cross-correlation Methods , 1995 .

[15]  K. Urban,et al.  Studying Atomic Structures by Aberration-Corrected Transmission Electron Microscopy , 2008, Science.

[16]  Maximilian Haider,et al.  Residual wave aberrations in the first spherical aberration corrected transmission electron microscope , 1998 .

[17]  H. Rose,et al.  Geometrical Charged-Particle Optics , 2008 .

[18]  Pierre Stadelmann,et al.  Effect of three-fold astigmatism on high resolution electron micrographs , 1995 .

[19]  D. Van Dyck,et al.  Ultimate resolution and information in electron microscopy II. The information limit of transmission electron microscopes , 1993 .

[20]  G. G. Stokes "J." , 1890, The New Yale Book of Quotations.

[21]  M. Haider,et al.  Correction of the spherical aberration of a 200 kV TEM by means of a hexapole-corrector , 1995 .

[22]  J. Barthel,et al.  Quantification of the information limit of transmission electron microscopes. , 2008, Physical review letters.

[23]  W. O. Saxton,et al.  A new way of measuring microscope aberrations , 2000, Ultramicroscopy.

[24]  F. Zemlin A practical procedure for alignment of a high resolution electron microscope , 1979 .

[25]  O. Scherzer The Theoretical Resolution Limit of the Electron Microscope , 1949 .

[26]  A. Thust,et al.  Focal-series reconstruction in HRTEM: simulation studies on non-periodic objects , 1996 .

[27]  R. Egerton Electron energy-loss spectroscopy in the TEM , 2008 .

[28]  C. Jia,et al.  Atomic-scale analysis of the oxygen configuration at a SrTiO3 dislocation core. , 2005, Physical review letters.

[29]  L. Houben,et al.  Atomic-precision determination of the reconstruction of a 90 degree tilt boundary in YBa2Cu3O7-delta by aberration corrected HRTEM. , 2006, Ultramicroscopy.

[30]  W. Coene,et al.  Improved methods for the determination of the spherical aberration coefficient in high-resolution electron microscopy from micrographs of an amorphous object , 1991 .

[31]  J. Frank,et al.  Signal-to-noise ratio of electron micrographs obtained by cross correlation , 1975, Nature.

[32]  Ondrej L. Krivanek,et al.  Towards sub-Å electron beams , 1999 .

[33]  Andreas Thust,et al.  Quantitative high-speed matching of high-resolution electron microscopy images , 1992 .

[34]  Kazuo Ishizuka,et al.  Contrast transfer of crystal images in TEM , 1980 .