Fracture of a Sivash-Range of Motion Femoral Stem at the Diaphyseal Spline

We discuss one of the four reported cases involving the fracture of a spline of the Sivash-range of motion (S-ROM) femoral prosthesis. It occurred in a 71-year-old female patient and was fully discovered during stem extraction in revision total hip arthroplasty (THA). The fractured spline was successfully removed using a reverse curette and fluoroscopic guidance. The placement of a new S-ROM femoral prosthesis was successful, and there were no other orthopedic complications.

[1]  H. Matar,et al.  Long-Term Survivorship of Modular Cementless Femoral Stem in Complex Primary Total Hip Arthroplasty: A Concise Minimum 15-Year Follow-Up Report. , 2021, The Journal of arthroplasty.

[2]  T. Wright,et al.  Fracture of an S-ROM stem at the sleeve-stem junction , 2018, Arthroplasty today.

[3]  R. Shah,et al.  Alarmingly High Rate of Implant Fractures in One Modular Femoral Stem Design: A Comparison of Two Implants. , 2017, The Journal of arthroplasty.

[4]  D. Dennis,et al.  Fracture of a Spline of the Femoral Stem in an Ingrown Modular Hip Implant , 2016, Hip international : the journal of clinical and experimental research on hip pathology and therapy.

[5]  J. Laffosse Removal of well-fixed fixed femoral stems. , 2016, Orthopaedics & traumatology, surgery & research : OTSR.

[6]  O. Huk,et al.  Stem-Sleeve Junction Failure of a Modular Femoral Hip System: a Retrieval Analysis , 2015, HSS journal : the musculoskeletal journal of Hospital for Special Surgery.

[7]  Young-Min Kwon,et al.  Modular hip implant fracture at the stem-sleeve interface. , 2015, Orthopedics.

[8]  W. Walter,et al.  The S-ROM hydroxyapatite proximally-coated modular femoral stem in revision hip replacement: results of 397 hips at a minimum ten-year follow-up. , 2014, The bone & joint journal.

[9]  W. Walter,et al.  Spontaneous fracture of diaphyseal stem of S-ROM femoral prosthesis , 2014, BMJ Case Reports.

[10]  N. Mehran,et al.  Failure of a modular hip implant at the stem-sleeve interface. , 2013, Orthopedics.

[11]  J. Currier,et al.  Femoral stem fracture and in vivo corrosion of retrieved modular femoral hips. , 2012, The Journal of arthroplasty.

[12]  E. Swarts,et al.  Proximal Component Modularity in THA—At What Cost?: An Implant Retrieval Study , 2012, Clinical orthopaedics and related research.

[13]  R. Calfee,et al.  Modular femoral stem-sleeve junction failure after primary total hip arthroplasty. , 2009, The Journal of arthroplasty.

[14]  Alan M Kop,et al.  Corrosion of a hip stem with a modular neck taper junction: a retrieval study of 16 cases. , 2009, The Journal of arthroplasty.

[15]  R. Jones,et al.  Primary total hip arthroplasty with use of the modular S-ROM prosthesis. Four to seven-year clinical and radiographic results. , 1999, The Journal of bone and joint surgery. American volume.

[16]  J. Lawrence,et al.  Acetabular defect classification and surgical reconstruction in revision arthroplasty. A 6-year follow-up evaluation. , 1994, The Journal of arthroplasty.