Dynamic field shaping for stereotactic radiosurgery: a modeling study.

PURPOSE This work assesses the relative field shaping advantages of dynamic field shaping devices for stereotactic radiosurgery using a linear accelerator. METHODS AND MATERIALS We selected 43 intracranial tumors (2.0-4.2 cm maximum dimension, 1.5-25.5 cc tumor volume) out of the first 64 intracranial tumors treated with radiosurgery at the Joint Center for Radiation Therapy. We modeled five field shaping devices, each including a fixed auxiliary circular collimator: (a) fixed circular collimator alone; (b) two independent parallel jaws; (c) four independent rectangular jaws; (d) four independent rotatable jaws; and (e) "ideal" multileaf collimator. We adjusted the model parameters until the minimum target isodose was 80% of the dose delivered to isocenter. We defined the treatment volume ratio as the target volume divided by the treatment volume (volume receiving at least the minimum target dose). We used the treatment volume ratio to compare the five models and the actual patient treatments. RESULTS For 34 tumors originally treated with one isocenter, the median Treatment Volume Ratio was higher for all of the device models except the fixed circular collimator compared to the actual patient treatments. For the nine tumors originally treated with multiple isocenters, the median Treatment Volume Ratio for the actual multiple isocenter treatments was similar to that for two parallel jaws, four rectangular jaws and four rotatable jaws. Only the median "ideal" collimator treatment volume ratio was higher for these nine tumors. CONCLUSION Simple field shaping devices have approximately 50% of the conformal advantage of an "ideal" multileaf collimator. Approximately 50% of typical radiosurgical tumors between 2 and 4 cm have field shaping advantages which exceed the geometrical uncertainties inherent in linear accelerator radiosurgery treatments. The three models, two parallel, four rectangular, or four rotatable independent jaws would improve current linear accelerator technology by providing homogeneous doses with equivalent field shaping for most tumors originally treated with inhomogeneous multiple isocenter plans (6/9 tumors in the current series).

[1]  William H. Press,et al.  Numerical Recipes in FORTRAN - The Art of Scientific Computing, 2nd Edition , 1987 .

[2]  J. Moeller,et al.  Optimization of electron arc therapy doses by multi-vane collimator control. , 1987, International journal of radiation oncology, biology, physics.

[3]  K. Nakagawa,et al.  An integrated radiotherapy treatment system and its clinical application. , 1987, Radiation medicine.

[4]  Anders Brahme ization of stationar , 1988 .

[5]  A. Montelius,et al.  Neutron beam characteristics from 50 MeV protons on beryllium using a continuously variable multi-leaf collimator. , 1983, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[6]  R L Siddon,et al.  Prism representation: a 3D ray-tracing algorithm for radiotherapy applications , 1985, Physics in medicine and biology.

[7]  L D Lunsford,et al.  Stereotactic radiosurgery for arteriovenous malformations of the brain. , 1991, Journal of neurosurgery.

[8]  A. Niemierko,et al.  Random sampling for evaluating treatment plans. , 1990, Medical physics.

[9]  K. Winston,et al.  Radiosurgery for arteriovenous malformations of the brain using a standard linear accelerator: rationale and technique. , 1988, International journal of radiation oncology, biology, physics.

[10]  R L Siddon,et al.  Solution to treatment planning problems using coordinate transformations. , 1981, Medical physics.

[11]  F. A. Seiler,et al.  Numerical Recipes in C: The Art of Scientific Computing , 1989 .

[12]  L D Lunsford,et al.  Treatment planning for gamma knife radiosurgery with multiple isocenters. , 1990, International journal of radiation oncology, biology, physics.

[13]  J. Schwade,et al.  Improved linac dose distributions for radiosurgery with irregularly shaped fields , 1990 .

[14]  J A Brace,et al.  Conformation therapy at the Royal Free Hospital. A progress report on the tracking cobalt project. , 1981, The British journal of radiology.

[15]  K. Winston,et al.  A system for stereotactic radiosurgery with a linear accelerator. , 1986, International journal of radiation oncology, biology, physics.

[16]  A. Brahme,et al.  Optimization of stationary and moving beam radiation therapy techniques. , 1988, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[17]  D D Leavitt,et al.  Dynamic field shaping to optimize stereotactic radiosurgery. , 1991, International journal of radiation oncology, biology, physics.

[18]  W Schlegel,et al.  Treatment planning for conformation therapy using a multi-leaf collimator. , 1988, Strahlentherapie und Onkologie : Organ der Deutschen Rontgengesellschaft ... [et al].

[19]  H. Kooy,et al.  Dose-volume histogram computations for small intracranial volumes. , 1993, Medical physics.

[20]  E J Holupka,et al.  Treatment planning for stereotactic radiosurgery of intra-cranial lesions. , 1991, International journal of radiation oncology, biology, physics.

[21]  H. Kooy,et al.  Variables associated with the development of complications from radiosurgery of intracranial tumors. , 1991, International journal of radiation oncology, biology, physics.

[22]  R L Siddon,et al.  Stereotactic radiosurgery for intracranial arteriovenous malformations using a standard linear accelerator. , 1989, International journal of radiation oncology, biology, physics.

[23]  S Sakuma,et al.  Computer-assisted rotation and multiple stationary irradiation technique. Newly designed overrunning multi-leaf collimators for conformation radiotherapy. , 1988, European journal of radiology.

[24]  G K Svensson,et al.  Measurements of dose distributions in small beams of 6 MV x-rays. , 1987, Physics in medicine and biology.

[25]  Computer controlled, multileaf collimator for rotataional radiation therapy. , 1979, AJR. American journal of roentgenology.