Discriminating between normal and glaucomatous eyes using the Heidelberg Retina Tomograph, GDx Nerve Fiber Analyzer, and Optical Coherence Tomograph.

OBJECTIVE To compare the ability of 3 instruments, the Heidelberg Retina Tomograph (HRT), the GDx Nerve Fiber Analyzer (GDx), and the Optical Coherence Tomograph (OCT), to discriminate between healthy eyes and eyes with early to moderate glaucomatous visual field loss. SUBJECTS AND METHODS Forty-one patients with early to moderate glaucomatous visual field loss and 50 healthy subjects were included in the study. The HRT, GDx, and OCT imaging and visual field testing were completed on 1 eye from each subject within a 6-month interval. Statistical differences in sensitivity at fixed specificities of 85%, 90%, and 95% were evaluated. In addition, areas under the receiver operating characteristic (ROC) curve were compared. RESULTS No significant differences were found between the area under the ROC curve and the best parameter from each instrument: OCT thickness at the 5-o'clock inferior temporal position (mean +/- SE, 0.87 +/- 0.04), HRT mean height contour in the nasal inferior region (mean +/- SE, 0.86 +/- 0.04), and GDx linear discriminant function (mean +/- SE, 0.84 +/- 0.04). Twelve HRT, 2 GDx, and 9 OCT parameters had an area under the ROC curve of at least 0.81. At a fixed specificity of 90%, significant differences were found between the sensitivity of OCT thickness at the 5-o'clock inferior temporal position (71%) and parameters with sensitivities less than 52%. Qualitative assessment of stereophotographs resulted in a sensitivity of 80%. CONCLUSION Although the area under the ROC curves was similar among the best parameters from each instrument, qualitative assessment of stereophotographs and measurements from the OCT and HRT generally had higher sensitivities than measurements from the GDx.

[1]  L. Zangwill,et al.  Reproducibility of nerve fiber layer thickness measurements by use of optical coherence tomography. , 2000, Ophthalmology.

[2]  L. Zangwill,et al.  A comparison of optical coherence tomography and retinal nerve fiber layer photography for detection of nerve fiber layer damage in glaucoma. , 2000, Ophthalmology.

[3]  L. Zangwill,et al.  The retinal nerve fiber layer thickness in ocular hypertensive, normal, and glaucomatous eyes with optical coherence tomography. , 2000, Archives of ophthalmology.

[4]  R N Weinreb,et al.  Is neuroprotection a viable therapy for glaucoma? , 1999, Archives of ophthalmology.

[5]  J G Fujimoto,et al.  Evaluation of focal defects of the nerve fiber layer using optical coherence tomography. , 1996, Ophthalmology.

[6]  L. Zangwill,et al.  Imaging technologies for assessing neuroprotection in glaucomatous optic neuropathy. , 1999, European journal of ophthalmology.

[7]  L. Zangwill,et al.  Detection of glaucoma with scanning laser polarimetry. , 1998, Archives of ophthalmology.

[8]  A. Dreher,et al.  Differentiating patients with glaucoma from glaucoma suspects and normal subjects by nerve fiber layer assessment with scanning laser polarimetry. , 1998, Ophthalmology.

[9]  J Caprioli,et al.  Slope of the peripapillary nerve fiber layer surface in glaucoma. , 1998, Investigative ophthalmology & visual science.

[10]  G. Wollstein,et al.  Identification of early glaucoma cases with the scanning laser ophthalmoscope. , 1998, Ophthalmology.

[11]  S M Drance,et al.  The ability of scanning laser ophthalmoscopy to identify various glaucomatous optic disk appearances. , 1998, American journal of ophthalmology.

[12]  P A Sample,et al.  Detection of early glaucomatous structural damage with confocal scanning laser tomography. , 1998, Journal of glaucoma.

[13]  N. Swindale,et al.  Sector‐Based Analysis of Optic Nerve Head Shape Parameters and Visual Field Indices in Healthy and Glaucomatous Eyes , 1997, Journal of glaucoma.

[14]  S M Drance,et al.  ROC analysis of Heidelberg Retina Tomograph optic disc shape measures in glaucoma. , 1997, Canadian journal of ophthalmology. Journal canadien d'ophtalmologie.

[15]  F W Fitzke,et al.  Variation of nerve fibre layer thickness measurements with age and ethnicity by scanning laser polarimetry , 1997, The British journal of ophthalmology.

[16]  S. Drance,et al.  The effect of optic disc size on diagnostic precision with the Heidelberg retina tomograph. , 1997, Ophthalmology.

[17]  H. Lemij,et al.  The sensitivity and specificity of nerve fiber layer measurements in glaucoma as determined with scanning laser polarimetry. , 1997, American journal of ophthalmology.

[18]  J. Caprioli,et al.  Detection of structural damage from glaucoma with confocal laser image analysis. , 1996, Investigative ophthalmology & visual science.

[19]  J. Fujimoto,et al.  Reproducibility of nerve fiber layer thickness measurements using optical coherence tomography. , 1996, Ophthalmology.

[20]  L Brigatti,et al.  Neural networks to identify glaucoma with structural and functional measurements. , 1996, American journal of ophthalmology.

[21]  J. Schuman Imaging Optic Nerve and Nerve Fiber Layer: Optical Coherence Tomography , 1996 .

[22]  N. Swindale,et al.  Ability of the Heidelberg Retina Tomograph to Detect Early Glaucomatous Visual Field Loss , 1995, Journal of glaucoma.

[23]  E A Swanson,et al.  Quantification of nerve fiber layer thickness in normal and glaucomatous eyes using optical coherence tomography. , 1995, Archives of ophthalmology.

[24]  L. Zangwill,et al.  Scanning laser polarimetry to measure the nerve fiber layer of normal and glaucomatous eyes. , 1995, American journal of ophthalmology.

[25]  J. Fujimoto,et al.  Optical coherence tomography: A new tool for glaucoma diagnosis , 1995, Current opinion in ophthalmology.

[26]  C E Phelps,et al.  Estimating Diagnostic Test Accuracy Using a "Fuzzy Gold Standard" , 1995, Medical decision making : an international journal of the Society for Medical Decision Making.

[27]  T Zeyen,et al.  Comparisons of methods to detect glaucomatous optic nerve damage. , 1993, Ophthalmology.

[28]  R N Weinreb,et al.  Effect of repetitive imaging on topographic measurements of the optic nerve head. , 1993, Archives of ophthalmology.

[29]  M Schulzer,et al.  Reproducibility of topographic parameters obtained with the heidelberg retina tomograph. , 1993, Journal of glaucoma.

[30]  J Caprioli,et al.  Progression of disc and field damage in early glaucoma. , 1993, Archives of ophthalmology.

[31]  A. Sommer,et al.  Clinically detectable nerve fiber atrophy precedes the onset of glaucomatous field loss. , 1991, Archives of ophthalmology.

[32]  J. Jonas,et al.  Histomorphometry of the human optic nerve. , 1990, Investigative ophthalmology & visual science.

[33]  M Schulzer,et al.  The normal human optic nerve. Axon count and axon diameter distribution. , 1989, Ophthalmology.

[34]  H A Quigley,et al.  The effect of age on normal human optic nerve fiber number and diameter. , 1989, Ophthalmology.

[35]  E. DeLong,et al.  Comparing the areas under two or more correlated receiver operating characteristic curves: a nonparametric approach. , 1988, Biometrics.

[36]  W. Green,et al.  Optic nerve damage in human glaucoma. III. Quantitative correlation of nerve fiber loss and visual field defect in glaucoma, ischemic neuropathy, papilledema, and toxic neuropathy. , 1982, Archives of ophthalmology.

[37]  D. R. Anderson,et al.  The mode of progressive disc cupping in ocular hypertension and glaucoma. , 1980, Archives of ophthalmology.

[38]  A. Sommer,et al.  Optic disc parameters and onset of glaucomatous field loss. II. Static screening criteria. , 1979, Archives of ophthalmology.

[39]  A. Sommer,et al.  The nerve fiber layer in the diagnosis of glaucoma. , 1977, Archives of ophthalmology.