Stanford 3D hyperthermia treatment planning system. Technical review and clinical summary.

In the field of deep regional hyperthermia cancer therapy the Sigma 60 applicator of the BSD-2000 Hyperthermia System is one of the most widely used devices. This device employs four independent sources of radiofrequency electromagnetic energy to heat tumour sites deep within the body. The difficulty in determining the input parameters for the four sources has motivated the development of a computer-based three-dimensional (3D) treatment planning system. The Stanford 3D Hyperthermia Treatment Planning System has been in clinical use at Stanford Medical Center for the past 2 years. It utilizes a patient-specific, three-dimensional computer simulation to determine safe and effective power deposition plans. An optimization programme for the selection of the amplitudes, phases and frequency for the sources has been developed and used in the clinic. Examples of the application of the treatment planning for hyperthermia treatment of pulmonary, pelvic, and mediastinal tumours are presented. Methods for quantifying the relative effectiveness of various treatment plans are reviewed.

[1]  B.J. James,et al.  Creation of three-dimensional patient models for hyperthermia treatment planning , 1992, IEEE Transactions on Biomedical Engineering.

[2]  C. Taylor,et al.  Electromagnetic pulse scattering in time-varying inhomogeneous media , 1969 .

[3]  Om P. Gandhi,et al.  Numerical simulation of annular phased arrays for anatomically based models using the FDTD method , 1989 .

[4]  P. Turner,et al.  BSD-2000 approach for deep local and regional hyperthermia: clinical utility. , 1989, Strahlentherapie und Onkologie : Organ der Deutschen Rontgengesellschaft ... [et al].

[5]  Dennis M. Sullivan,et al.  Frequency-dependent FDTD methods using Z transforms , 1992 .

[6]  A. Guy,et al.  Nonionizing electromagnetic wave effects in biological materials and systems , 1972 .

[7]  D M Sullivan,et al.  Potential hazards of radiative electromagnetic hyperthermia in the presence of multiple metallic surgical clips. , 1992, International journal of hyperthermia : the official journal of European Society for Hyperthermic Oncology, North American Hyperthermia Group.

[8]  P. Deuflhard,et al.  Strategies for optimized application of annular-phased-array systems in clinical hyperthermia. , 1991, International journal of hyperthermia : the official journal of European Society for Hyperthermic Oncology, North American Hyperthermia Group.

[9]  Dennis M. Sullivan,et al.  A frequency-dependent FDTD method for biological applications , 1992 .

[10]  S. Knox,et al.  Hyperthermia and radiation therapy in the treatment of recurrent merkel cell tumors , 1988, Cancer.

[11]  Allen Taflove,et al.  Review of the formulation and applications of the finite-difference time-domain method for numerical modeling of electromagnetic wave interactions with arbitrary structures , 1988 .

[12]  J. Lyman Complication Probability as Assessed from Dose-Volume Histograms , 1985 .

[13]  V. Sathiaseelan,et al.  Theoretical Analysis and Clinical Demonstration of the Effect of Power Pattern Control Using the Annular Phased-Array Hyperthermia System , 1986 .

[14]  R. B. Standler,et al.  A frequency-dependent finite-difference time-domain formulation for dispersive materials , 1990 .

[15]  Bladder cooling in patients treated with regional hyperthermia of the pelvis using an annular phased array. , 1988, International journal of radiation oncology, biology, physics.

[16]  Dennis M. Sullivan,et al.  Mathematical methods for treatment planning in deep regional hyperthermia , 1991 .

[17]  Om P. Gandhi,et al.  Use of the Finite-Difference Time-Domain Method in Calculating EM Absorption in Human Tissues , 1987, IEEE Transactions on Biomedical Engineering.

[18]  D. Sullivan,et al.  Direct use of CT scans for hyperthermia treatment planning , 1992, IEEE Transactions on Biomedical Engineering.

[19]  Dennis M. Sullivan,et al.  Comparison of measured and simulated data in an annular phased array using an inhomogeneous phantom , 1992 .

[20]  D. Sullivan,et al.  Peripheral neuropathy and myonecrosis following hyperthermia and radiation therapy for recurrent prostatic cancer: correlation of damage with predicted SAR pattern. , 1992, International journal of hyperthermia : the official journal of European Society for Hyperthermic Oncology, North American Hyperthermia Group.

[21]  D. Kapp,et al.  Local hyperthermia and radiation therapy in the retreatment of superficially located recurrences in Hodgkin's disease. , 1990, International journal of radiation oncology, biology, physics.

[22]  Dennis M. Sullivan,et al.  Three-dimensional computer simulation in deep regional hyperthermia using the finite-difference time-domain method , 1990 .

[23]  E. H. Curtis,et al.  Optimization of the absorbed power distribution for an annular phased array hyperthermia system. , 1989, International journal of radiation oncology, biology, physics.

[24]  K. Yee Numerical solution of initial boundary value problems involving maxwell's equations in isotropic media , 1966 .

[25]  G T Chen,et al.  Dose volume histogram analysis of liver radiation tolerance. , 1986, International journal of radiation oncology, biology, physics.