A weighted rebinned backprojection‐filtration algorithm from partially beam‐blocked data for a single‐scan cone‐beam CT with hybrid type scatter correction

PURPOSE Scatter contamination constitutes a dominant source of degradation of image quality in cone-beam computed tomography (CBCT). We have recently developed an analytic image reconstruction method with a scatter correction capability from the partially blocked cone-beam data out of a single scan. Despite its easy implementation and its computational efficiency, the developed method may result in additional image artifacts for a large cone angle geometry due to data inconsistency. To improve the image quality at a large cone angle, we propose a weighted rebinned backprojection-filtration (wrBPF) algorithm in conjunction with a hybrid type scatter correction approach. METHODS The proposed method uses a beam-blocker array that provides partial data for scatter correction and image reconstruction and that only blocks the beam within a limited cone angle. This design allows a chance to keep the image quality at larger cone angles by use of data redundancy since the projection data corresponding to larger cone angles are not blocked. However, the scatter correction would not be straightforward. In order to correct for the scatter in the projections at larger cone angles, we propose a novel scatter correction method combining a measurement-based and a convolution-based method. We first estimated the scatter signal using a measurement-based method in the partially beam-blocked regions, and then optimized the fitting parameters of a convolution-kernel that can be used for scatter correction in the projections at larger cone angles. For image reconstruction, we developed a wrBPF with butterfly filtering. We have conducted an experimental study to validate the proposed algorithm for image reconstruction and scatter correction. RESULTS The experimental results revealed that the developed reconstruction method makes full use of the benefits of partial beam-blocking for scatter correction and image reconstruction and at the same time enhances image quality at larger cone angles by use of an optimized convolution-based scatter correction. CONCLUSIONS The proposed method that enjoys the advantages of both measurement-based and convolution-based methods for scatter correction has successfully demonstrated its capability of reconstructing accurate images out of a single scan in circular CBCT.

[1]  Seungryong Cho,et al.  Analytic image reconstruction from partial data for a single-scan cone-beam CT with scatter correction. , 2015, Medical physics.

[2]  Henrik Turbell,et al.  Cone-Beam Reconstruction Using Filtered Backprojection , 2001 .

[3]  Willi A Kalender,et al.  A fast and pragmatic approach for scatter correction in flat-detector CT using elliptic modeling and iterative optimization , 2010, Physics in medicine and biology.

[4]  J. Star-Lack,et al.  Improved scatter correction using adaptive scatter kernel superposition , 2010, Physics in medicine and biology.

[5]  K. Klingenbeck-Regn,et al.  Efficient object scatter correction algorithm for third and fourth generation CT scanners , 1999, European Radiology.

[6]  Lei Xing,et al.  Scatter correction in cone-beam CT via a half beam blocker technique allowing simultaneous acquisition of scatter and image information. , 2012, Medical physics.

[7]  Ning Wen,et al.  Combining scatter reduction and correction to improve image quality in cone-beam computed tomography (CBCT). , 2010, Medical physics.

[8]  W. Kalender,et al.  Technical note: comparing coherent and incoherent scatter effects for cone-beam CT , 2008, Physics in medicine and biology.

[9]  Lei Zhu,et al.  Scatter correction for full-fan volumetric CT using a stationary beam blocker in a single full scan. , 2011, Medical physics.

[10]  Willi A Kalender,et al.  X-ray scatter data for flat-panel detector CT. , 2007, Physica medica : PM : an international journal devoted to the applications of physics to medicine and biology : official journal of the Italian Association of Biomedical Physics.

[11]  D. Jaffray,et al.  Efficient scatter distribution estimation and correction in CBCT using concurrent Monte Carlo fitting. , 2014, Medical physics.

[12]  Xi Chen,et al.  Robust moving-blocker scatter correction for cone-beam computed tomography using multiple-view information , 2017, PloS one.

[13]  L. Feldkamp,et al.  Practical cone-beam algorithm , 1984 .

[14]  Wei Zhao,et al.  Patient-specific scatter correction for flat-panel detector-based cone-beam CT imaging. , 2015, Physics in medicine and biology.

[15]  Ruola Ning,et al.  X-ray scatter correction algorithm for cone beam CT imaging. , 2004, Medical physics.

[16]  J H Siewerdsen,et al.  Cone-beam computed tomography with a flat-panel imager: initial performance characterization. , 2000, Medical physics.

[17]  Bin Li,et al.  Optimization of the geometry and speed of a moving blocker system for cone‐beam computed tomography scatter correction , 2017, Medical physics.

[18]  Freek J. Beekman,et al.  Efficient Monte Carlo based scatter artifact reduction in cone-beam micro-CT , 2006, IEEE Transactions on Medical Imaging.

[19]  Lei Zhu,et al.  Scatter Correction Method for X-Ray CT Using Primary Modulation: Theory and Preliminary Results , 2006, IEEE Transactions on Medical Imaging.

[20]  Klaus Klingenbeck,et al.  A general framework and review of scatter correction methods in x-ray cone-beam computerized tomography. Part 1: Scatter compensation approaches. , 2011, Medical physics.

[21]  Lei Zhu,et al.  Scatter correction for cone-beam CT in radiation therapy. , 2009 .

[22]  Lei Xing,et al.  Binary moving-blocker-based scatter correction in cone-beam computed tomography with width-truncated projections: proof of concept , 2017, Physics in medicine and biology.

[23]  Xiaochuan Pan,et al.  A rebinned backprojection-filtration algorithm for image reconstruction in helical cone-beam CT , 2007, Physics in medicine and biology.

[24]  Changhwan Kim,et al.  Data consistency-driven scatter kernel optimization for x-ray cone-beam CT , 2015, Physics in medicine and biology.

[25]  H Zaidi,et al.  Measurement of scattered radiation in a volumetric 64-slice CT scanner using three experimental techniques , 2010, Physics in medicine and biology.

[26]  S. Richard,et al.  A simple, direct method for x-ray scatter estimation and correction in digital radiography and cone-beam CT. , 2005, Medical physics.

[27]  Marc Kachelrieß,et al.  Hybrid scatter correction for CT imaging. , 2012, Physics in medicine and biology.

[28]  Xiaochuan Pan,et al.  A unified analysis of FBP-based algorithms in helical cone-beam and circular cone- and fan-beam scans. , 2004, Physics in medicine and biology.

[29]  Jing Wang,et al.  4D cone-beam computed tomography (CBCT) using a moving blocker for simultaneous radiation dose reduction and scatter correction , 2018, Physics in medicine and biology.

[30]  W. Kalender,et al.  Combining deterministic and Monte Carlo calculations for fast estimation of scatter intensities in CT , 2006, Physics in medicine and biology.

[31]  N. R. Bennett,et al.  Scatter correction method for x-ray CT using primary modulation: phantom studies. , 2010, Medical physics.

[32]  Michael Unser,et al.  Interior Tomography Using 1D Generalized Total Variation. Part II: Multiscale Implementation , 2015, SIAM J. Imaging Sci..

[33]  Willi A Kalender,et al.  Ring artifact correction for high-resolution micro CT , 2009, Physics in medicine and biology.

[34]  Wei Zhao,et al.  A model-based scatter artifacts correction for cone beam CT. , 2016, Medical physics.

[35]  Radhe Mohan,et al.  Scatter kernel estimation with an edge-spread function method for cone-beam computed tomography imaging , 2008, Physics in medicine and biology.

[36]  Emil Y. Sidky,et al.  Region of interest reconstruction from truncated data in circular cone-beam CT , 2006, IEEE Transactions on Medical Imaging.

[37]  Freek J. Beekman,et al.  Accelerated simulation of cone beam X-ray scatter projections , 2004, IEEE Transactions on Medical Imaging.

[38]  Shaojie Tang,et al.  Axial Cone-Beam Reconstruction by Weighted BPF/DBPF and Orthogonal Butterfly Filtering , 2016, IEEE Transactions on Biomedical Engineering.

[39]  William H. Press,et al.  Numerical Recipes 3rd Edition: The Art of Scientific Computing , 2007 .

[40]  Xiaochuan Pan,et al.  Exact image reconstruction on PI-lines from minimum data in helical cone-beam CT. , 2004, Physics in medicine and biology.

[41]  K. Klingenbeck,et al.  A general framework and review of scatter correction methods in cone beam CT. Part 2: scatter estimation approaches. , 2011, Medical physics.

[42]  M. Kachelriess,et al.  Robust primary modulation-based scatter estimation for cone-beam CT. , 2015, Medical physics.