Do ‘passive’ medical titanium surfaces deteriorate in service in the absence of wear?

Globally, more than 1000 tonnes of titanium (Ti) is implanted into patients in the form of biomedical devices on an annual basis. Ti is perceived to be ‘biocompatible’ owing to the presence of a robust passive oxide film (approx. 4 nm thick) at the metal surface. However, surface deterioration can lead to the release of Ti ions, and particles can arise as the result of wear and/or corrosion processes. This surface deterioration can result in peri-implant inflammation, leading to the premature loss of the implanted device or the requirement for surgical revision. Soft tissues surrounding commercially pure cranial anchorage devices (bone-anchored hearing aid) were investigated using synchrotron X-ray micro-fluorescence spectroscopy and X-ray absorption near edge structure. Here, we present the first experimental evidence that minimal load-bearing Ti implants, which are not subjected to macroscopic wear processes, can release Ti debris into the surrounding soft tissue. As such debris has been shown to be pro-inflammatory, we propose that such distributions of Ti are likely to effect to the service life of the device.

[1]  A. Sanz-Medel,et al.  Titanium release in serum of patients with different bone fixation implants and its interaction with serum biomolecules at physiological levels , 2011, Analytical and bioanalytical chemistry.

[2]  A. Reid,et al.  Fluid Exudates From Inflamed Bone-Anchored Hearing Aids Demonstrate Elevated Levels of Cytokines and Biomarkers of Tissue and Bone Metabolism , 2010, Otology & neurotology : official publication of the American Otological Society, American Neurotology Society [and] European Academy of Otology and Neurotology.

[3]  H. Simmen,et al.  Titanium uptake, induction of RANK‐L expression, and enhanced proliferation of human T‐lymphocytes , 2010, Journal of Orthopaedic Research.

[4]  N. Jacobsen,et al.  Role of oxidative damage in toxicity of particulates , 2010, Free radical research.

[5]  L. Filgueira,et al.  Metal is not inert: role of metal ions released by biocorrosion in aseptic loosening--current concepts. , 2009, Journal of biomedical materials research. Part A.

[6]  Stuart A. Cavill,et al.  I18--the microfocus spectroscopy beamline at the Diamond Light Source. , 2009, Journal of synchrotron radiation.

[7]  L. Filgueira,et al.  Uptake and intracellular distribution of various metal ions in human monocyte-derived dendritic cells detected by Newport Green™ DCF diacetate ester , 2009, Journal of Neuroscience Methods.

[8]  Kevin Ong,et al.  The epidemiology of revision total hip arthroplasty in the United States. , 2009, The Journal of bone and joint surgery. American volume.

[9]  L. Yin,et al.  Dynamic performance of the beam position monitor support at the SSRF. , 2009, Journal of synchrotron radiation.

[10]  P. Revell,et al.  The combined role of wear particles, macrophages and lymphocytes in the loosening of total joint prostheses , 2008, Journal of The Royal Society Interface.

[11]  A. Sanz-Medel,et al.  High-resolution ICP–MS determination of Ti, V, Cr, Co, Ni, and Mo in human blood and urine of patients implanted with a hip or knee prosthesis , 2008, Analytical and bioanalytical chemistry.

[12]  J. W. House,et al.  Bone-Anchored Hearing Aids: Incidence and Management of Postoperative Complications , 2007, Otology & neurotology : official publication of the American Otological Society, American Neurotology Society [and] European Academy of Otology and Neurotology.

[13]  J. Hunt,et al.  Host inflammatory response to NiCr, CoCr, and Ti in a soft tissue implantation model. , 2006, Journal of biomedical materials research. Part A.

[14]  P. Korovessis,et al.  Metallosis after contemporary metal-on-metal total hip arthroplasty. Five to nine-year follow-up. , 2006, The Journal of bone and joint surgery. American volume.

[15]  D. Gillett,et al.  Bone-anchored hearing aids: results of the first eight years of a programme in a district general hospital, assessed by the Glasgow benefit inventory , 2006, The Journal of Laryngology & Otology.

[16]  R. Tuan,et al.  Particulate endocytosis mediates biological responses of human mesenchymal stem cells to titanium wear debris , 2006, Journal of orthopaedic research : official publication of the Orthopaedic Research Society.

[17]  K. Mikecz,et al.  Effects of soluble metals on human peri-implant cells. , 2005, Journal of biomedical materials research. Part A.

[18]  M Newville,et al.  ATHENA, ARTEMIS, HEPHAESTUS: data analysis for X-ray absorption spectroscopy using IFEFFIT. , 2005, Journal of synchrotron radiation.

[19]  G. Truskey,et al.  Effects of titanium particle size on osteoblast functions in vitro and in vivo. , 2005, Proceedings of the National Academy of Sciences of the United States of America.

[20]  Ricardo M. Souto,et al.  The effect of temperature on the nucleation of corrosion pits on titanium in Ringer's physiological solution. , 2005, Biomaterials.

[21]  J. Cobb,et al.  Pain in the well-fixed, aseptic titanium hip replacement. The role of corrosion. , 2004, The Journal of bone and joint surgery. British volume.

[22]  S. Ichinose,et al.  The study of metal ion release and cytotoxicity in Co–Cr–Mo and Ti–Al–V alloy in total knee prosthesis – scanning electron microscopic observation , 2003, Journal of materials science. Materials in medicine.

[23]  J. Fisher,et al.  Biological reactions to wear debris in total joint replacement , 2000, Proceedings of the Institution of Mechanical Engineers. Part H, Journal of engineering in medicine.

[24]  J. C. Griess Crevice Corrosion of Titanium in Aqueous Salt Solutions , 1968 .