Assessment of new small-field detectors against standard-field detectors for practical stereotactic beam data acquisition.

Two new detectors (0.015 cm3 ion chamber from PTW, 0.6 mm diameter diode from Scanditronix AB) designed specifically for use in small stereotactic fields were compared against similar, more routine, detectors (0.125 cm3 ion chamber, parallel plate chamber, shielded and unshielded diodes and film). Percentage depth doses, tissue maximum ratios, off-axis ratios and relative output factors were compared for circular fields in the 40-12.5 mm diameter range, with a view to identifying the optimum detector for stereotactic beam data acquisition. Practical suggestions for beam data collection and analysis are made, with an emphasis on what is achievable practically in radiotherapy departments where the primary demand is to provide a routine service. No single detector was found to be ideal, and neither of the two new measurement devices had any significant advantages over more routine devices, in the situations measured. Although the new 0.015 cm3 ion chamber was an improvement on a 0.125 cm3 ion chamber in the measurement of profiles, it was still too large when compared with a diode. The new small diode had a low signal to noise ratio which made reliable data difficult to extract and its only advantage is possibly improved resolution in fields smaller than the range tested. The use of a larger unshielded diode is recommended for all measurements, with the additional cross-checking of data against at least one small ion chamber and film. A simple method of obtaining reliable output data from the detectors used is explained.

[1]  P. Almond,et al.  Silicon diode detectors used in radiological physics measurements. Part I: Development of an energy compensating shield. , 1977, Medical physics.

[2]  S C Sharma,et al.  Film dosimetry of megavoltage photon beams: a practical method of isodensity-to-isodose curve conversion. , 1981, Medical physics.

[3]  J. VanBuren,et al.  Dosimetry of small radiation fields for 10-MV x rays. , 1983, Medical physics.

[4]  D. J. Dawson,et al.  Analysis of physical parameters associated with the measurement of high-energy x-ray penumbra. , 1984, Medical physics.

[5]  Dose measurements and calculations of small radiation fields for 9-MV x rays. , 1985, Medical physics.

[6]  G. Rikner,et al.  Selective shielding of a p-Si detector for quality independence. , 1985, Acta radiologica. Oncology.

[7]  Penumbral measurements in water for high-energy x rays. , 1986, Medical physics.

[8]  G K Svensson,et al.  Measurements of dose distributions in small beams of 6 MV x-rays. , 1987, Physics in medicine and biology.

[9]  B. Bjärngard,et al.  Doses on the central axes of narrow 6-MV x-ray beams. , 1990, Medical physics.

[10]  D. F. Lewis,et al.  GafchromicTM dosimetry media: A new high dose, thin film routine dosimeter and dose mapping tool , 1990 .

[11]  C. Sibata,et al.  Influence of detector size in photon beam profile measurements. , 1991, Physics in medicine and biology.

[12]  G. Hartmann,et al.  Tissue maximum ratios (and other parameters) of small circular 4, 6, 10, 15 and 24 MV x-ray beams for radiosurgery. , 1992, Physics in medicine and biology.

[13]  Physics for radiosurgery with linear accelerators. , 1992, Neurosurgery clinics of North America.

[14]  T Kron,et al.  Dosimetry of 6-MV x-ray beam penumbra. , 1993, Medical physics.

[15]  Analysis of dosimetric measurements in linac radiosurgery calibration. , 1993, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[16]  E B Podgorsak,et al.  Buildup region of high-energy x-ray beams in radiosurgery. , 1993, Medical physics.

[17]  Beam profile analysis using GafChromic films. , 1993, Stereotactic and functional neurosurgery.

[18]  Imaging of radiation dose for stereotactic radiosurgery. , 1993, Medical dosimetry : official journal of the American Association of Medical Dosimetrists.

[19]  P. O'Brien,et al.  Absorbed dose perturbation caused by diodes for small field photon dosimetry. , 1994, Medical physics.

[20]  E. McCullough,et al.  The use of a radiochromic detector for the determination of stereotactic radiosurgery dose characteristics. , 1994, Medical physics.

[21]  S. Rustgi Evaluation of the dosimetric characteristics of a diamond detector for photon beam measurements. , 1995, Medical physics.

[22]  C. Sibata,et al.  Deconvolution of detector size effect for small field measurement. , 1995, Medical physics.

[23]  P W Hoban,et al.  A comparison of dosimetry techniques in stereotactic radiosurgery. , 1996, Physics in medicine and biology.

[24]  Dosimetry of very-small (5-10 mm) and small (12.5-40 mm) diameter cones and dose verification for radiosurgery with 6-MV X-ray beams. , 1996, Stereotactic and functional neurosurgery.

[25]  B W Corn,et al.  Characteristics of a dedicated linear accelerator-based stereotactic radiosurgery-radiotherapy unit. , 1996, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[26]  M. Karlsson,et al.  Semi-conductor detectors in output factor measurements. , 1997, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[27]  H. Minn,et al.  Dose characteristics of in-house-built collimators for stereotactic radiotherapy with a linear accelerator. , 1998, Physics in medicine and biology.

[28]  S Cora,et al.  Use of a new type of radiochromic film, a new parallel-plate micro-chamber, MOSFETs, and TLD 800 microcubes in the dosimetry of small beams. , 1998, Medical physics.

[29]  J. Galvin,et al.  Radiochromic film dosimetry: recommendations of AAPM Radiation Therapy Committee Task Group 55. American Association of Physicists in Medicine. , 1998, Medical physics.

[30]  G. Wickman,et al.  Liquid ionization chamber measurements of dose distributions in small 6 MV photon beams. , 1998, Physics in medicine and biology.

[31]  C. Fiorino,et al.  Dosimetry of Gamma Knife and linac-based radiosurgery using radiochromic and diode detectors. , 1999, Physics in medicine and biology.