Variations in aortic pressure affect the mechanics of the intra-aortic balloon: an in vitro investigation.

This study aims to investigate the mechanics of the intra-aortic balloon (IAB) under different aortic pressure (P(ao)) and inclination (0-75 degrees). Pressure and flow were measured in an artificial aorta during IAB pumping with a frequency of 1:3. Volume displaced toward the "coronary arteries" during inflation (V(prox)) and "intra-aortic" pressure reduction during deflation (P(r)) were derived. IAB duration of inflation and deflation was determined with a high-speed camera visualization. When the aorta was horizontal, P(ao) raised from 45 mm Hg to 115 mm Hg, V(prox) reduced by 18% (25.0 +/- 1.0 mL vs. 30.4 +/- 1.9 mL) and P(r) increased by 117% (106.4 +/- 0.3 mm Hg vs. 48.9 +/- 0.6 mm Hg). When the aorta was inclined, at low P(ao) of 45 mm Hg, V(prox) was reduced by 30% from 0 degrees to 45 degrees (19.8 +/- 2.3 mL vs. 28.3 +/- 1.7 mL) and P(r) was reduced by 66% (16.5 +/- 0.1 mm Hg vs. 48.9 +/- 0.6 mm Hg). However, at high P(ao) of 115 mm Hg, V(prox) remained unchanged with increasing angle (20.0 +/- 1.0 mL) and P(r) was reduced by 24% (80.6 +/- 0.8 mm Hg vs. 106.4 +/- 0.3 mm Hg). Increasing P(ao) increased duration of inflation. At low P(ao), increasing angle resulted in increasing duration of inflation, but at high P(ao), increasing angle had the opposite effect. Duration of deflation generally decreased with P(ao) and increased with increasing angle. The IAB pump is affected by both P(ao) and angle, indicating that non-normotensive patients or patients in the semi-recumbent position might not receive the full benefits of IAB counterpulsation.

[1]  W J KOLFF,et al.  Diastolic balloon pumping (with carbon dioxide) in the aorta--a mechanical assistance to the failing circulation. , 1962, American heart journal.

[2]  C. Stefanadis,et al.  Basic Principles of the Intraaortic Balloon Pump and Mechanisms Affecting Its Performance , 2005, ASAIO journal.

[3]  J. Paratz,et al.  Positioning practices for ventilated intensive care patients: current practice, indications and contraindications. , 2006, Australian critical care : official journal of the Confederation of Australian Critical Care Nurses.

[4]  R. Stenz Intra-aortic balloon counterpulsation , 2006 .

[5]  J. Rello,et al.  Evidence on measures for the prevention of ventilator-associated pneumonia , 2007, European Respiratory Journal.

[6]  Susan J. Quaal,et al.  Comprehensive intraaortic balloon counterpulsation , 1993 .

[7]  K. Grady,et al.  Intraaortic balloon pumping through the common iliac artery: management of the ambulatory intraaortic balloon pump patient. , 2000, Progress in cardiovascular nursing.

[8]  D. A. Young,et al.  Intra-aortic balloon pumping: does posture matter? , 2005, Artificial organs.

[9]  O. Barnea,et al.  Cardiac energy considerations during intraaortic balloon pumping , 1990, IEEE Transactions on Biomedical Engineering.

[10]  J. Pepper,et al.  Pressure and flow-volume distribution associated with intra-aortic balloon inflation: an in vitro study. , 2007, Artificial organs.

[11]  S. Moulopoulos The Limits of Counterpulsation , 1993, The International journal of artificial organs.

[12]  H. Gewirtz,et al.  The Effect of Intraaortic Balloon Counterpulsation on Regional M4yocardial Blood Flow and Oxygen Consumption in the Presence of Coronary Artery Stenosis in Patients with Unstable Angina , 1982, Circulation.

[13]  Semi-recumbent Position in ICU , 2008 .

[14]  J. Nanas,et al.  Arterial compliance is a main variable determining the effectiveness of intra-aortic balloon counterpulsation: quantitative data from an in vitro study. , 2002, Medical engineering & physics.

[15]  C. Charitos,et al.  The efficacy of the high volume counterpulsation technique at very low levels of aortic pressure. , 1998, The Journal of cardiovascular surgery.

[16]  Kimon S Stamatelopoulos,et al.  New aspects on the role of blood pressure and arterial stiffness in mechanical assistance by intra-aortic balloon pump: in-vitro data and their application in clinical practice. , 2004, Artificial organs.