Clinical evaluation of Er:YAG laser caries treatment

To prepare the enamel, the energy used was mainly 345 mJ and repetition rate 2 Hz, for dentine the optimal energy of Er:YAG drilling machine was 200 mJ and repetition rate from 1 to 2 Hz, depending on cavity depth. Subject of treatment were caries of enamel and dentine and it was possible to remove the old insufficient fillings. The average number of pulses was 111.22, ranging from 16 to 489. During preparation, vibrations of microexplosions were felt by 8 patients, however, neither pain or unpleasant sensations were experienced. The filling materials used were composite resins and glassionomer cements. Their clinical evaluation 6 months post insertion was similar to that of the classical drilling system.

[1]  Peter Rechmann,et al.  Caries selective ablation: effects of water cooling , 1994, Other Conferences.

[2]  Raimund Hibst,et al.  Heat effect of pulsed Er:YAG laser radiation , 1990, Photonics West - Lasers and Applications in Science and Engineering.

[3]  Raimund Hibst,et al.  Experimental studies of the application of the Er:YAG laser on dental hard substances: I. Measurement of the ablation rate , 1989, Lasers in surgery and medicine.

[4]  N M King,et al.  A pilot study of the marginal adaptation and surface morphology of glass-cermet cements. , 1996, Quintessence international.

[5]  C M Cobb,et al.  The effect of an air-powder abrasive system on in vitro root surfaces. , 1984, Journal of periodontology.

[6]  M. Wolbarsht,et al.  Wet versus dry enamel ablation by Er:YAG laser. , 1992, The Journal of prosthetic dentistry.