The external magnetic fields produced by artificial current dipoles placed in conducting-gel filled spheres, partial spheres and human skulls have been measured. Using a homogeneously conducting sphere model, the dipole parameters have been estimated. Although good quality data fitting is obtainable for a range of positions of the centre of the sphere used in the modelling, the predictions for the dipole obtained with those different assumed positions differ considerably. It is found that the most reliable method of obtaining accurate estimates of the dipole parameters is by taking an average of the predictions of several of the best-fit analyses. The information about the accuracy of the predictions that is obtainable from the fitting routine's residual error parameter is discussed.
[1]
L. Kaufman,et al.
Application of SQUID sensors to the investigation of neural activity in the human brain
,
1983
.
[2]
B. Neil Cuffin,et al.
Magnetic Fields of a Dipole in Special Volume Conductor Shapes
,
1977,
IEEE Transactions on Biomedical Engineering.
[3]
J Beatty,et al.
Magnetic localization of a dipolar current source implanted in a sphere and a human cranium.
,
1986,
Electroencephalography and clinical neurophysiology.
[4]
B. Neil Cuffin,et al.
Effects of fissures in the brain on electroencephalograms and magnetoencephalograms
,
1985
.