Computational evaluation of aortic occlusion and the proposal of a novel, improved occluder: Constrained endo‐aortic balloon occlusion

Because aortic occlusion is arguably one of the most dangerous aortic manipulation maneuvers during cardiac surgery in terms of perioperative ischemic neurological injury, the purpose of this investigation is to assess the structural mechanical impact resulting from the use of existing and newly proposed occluders. Existing (clinically used) occluders considered include different cross-clamps (CCs) and endo-aortic balloon occlusion (EABO). A novel occluder is also introduced, namely, constrained EABO (CEABO), which consists of applying a constrainer externally around the aorta when performing EABO. Computational solid mechanics are employed to investigate each occluder according to a comprehensive list of functional requirements. The potential of a state of occlusion is also considered for the first time. Three different constrainer designs are evaluated for CEABO. Although the CCs were responsible for the highest strains, largest deformation, and most inefficient increase of the occlusion potential, it remains the most stable, simplest, and cheapest occluder. The different CC hinge geometries resulted in poorer performance of CC used for minimally invasive procedures than conventional ones. CEABO with a profiled constrainer successfully addresses the EABO shortcomings of safety, stability, and positioning accuracy, while maintaining its complexities of operation (disadvantage) and yielding additional functionalities (advantage). Moreover, CEABO is able to achieve the previously unattainable potential to provide a clinically determinable state of occlusion. CEABO offers an attractive alternative to the shortcomings of existing occluders, with its design rooted in achieving the highest patient safety. Copyright © 2016 John Wiley & Sons, Ltd.

[1]  Feasibility Study of a Direct Endo-Aortic Clamp Balloon , 2007 .

[2]  Ansar Hassan,et al.  Minimally invasive mitral valve surgery: a systematic review and meta-analysis. , 2008, European journal of cardio-thoracic surgery : official journal of the European Association for Cardio-thoracic Surgery.

[3]  B. Geršak,et al.  Direct measurement of clamping forces in cardiovascular surgery , 2006, Medical and Biological Engineering and Computing.

[4]  Manuel Doblaré,et al.  An uncoupled directional damage model for fibred biological soft tissues. Formulation and computational aspects , 2007 .

[5]  M. Dujovny,et al.  Minimum vascular occlusive force. , 1979, Journal of neurosurgery.

[6]  Gerhard A Holzapfel,et al.  A three-dimensional finite element model for arterial clamping. , 2001, Journal of biomechanical engineering.

[7]  R. Suri Minimally invasive video-assisted mitral valve surgery: A 12-year, 2-center experience in 1178 patients , 2010 .

[8]  F. Musumeci,et al.  Endovascular balloon versus transthoracic aortic clamping for minimally invasive mitral valve surgery: impact on cerebral microemboli. , 2006, Interactive cardiovascular and thoracic surgery.

[9]  G M McKhann,et al.  Adverse cerebral outcomes after coronary bypass surgery. , 1997, The New England journal of medicine.

[10]  C. Huber,et al.  Ascending aorta measurements as assessed by ECG-gated multi-detector computed tomography: a pilot study to establish normative values for transcatheter therapies , 2009, European Radiology.

[11]  A. Pappalardo,et al.  Surgical management of the atherosclerotic ascending aorta: is endoaortic balloon occlusion safe? , 2006, The Annals of thoracic surgery.

[12]  A. Khera,et al.  Agreement between methods of measurement of mean aortic wall thickness by MRI , 2009, Journal of magnetic resonance imaging : JMRI.

[13]  Gerhard A. Holzapfel,et al.  Nonlinear Solid Mechanics: A Continuum Approach for Engineering Science , 2000 .

[14]  Gerhard A Holzapfel,et al.  Computational stress-deformation analysis of arterial walls including high-pressure response. , 2007, International journal of cardiology.

[15]  Françoise Thivolet,et al.  Comparison between 3 aortic clamps for video-assisted cardiac surgery: a histological study in a pig model. , 2006, The heart surgery forum.

[16]  Wolfgang A. Wall,et al.  A computational strategy for prestressing patient‐specific biomechanical problems under finite deformation , 2010 .

[17]  Jay D. Humphrey,et al.  Theory of small on large: Potential utility in computations of fluid–solid interactions in arteries , 2007 .

[18]  M. Mack,et al.  Mitral valve surgery: comparison of outcomes in matched sternotomy and port access groups. , 2010, The Journal of heart valve disease.

[19]  Thomas Boland,et al.  In vitro evaluation of phosphonylated low-density polyethylene for vascular applications. , 2002, Journal of biomedical materials research.

[20]  A. Galloway,et al.  First report of the Port Access International Registry. , 1999, The Annals of thoracic surgery.

[21]  M. Cengiz,et al.  Pressure-controlled Vascular Clamp: A Novel Device for Atraumatic Vessel Occlusion , 2004, Annals of vascular surgery.

[22]  Y. Bobryshev,et al.  Artery Wall Damage and Platelet Uptake from So-Called Atraumatic Arterial Clamps: An Experimental Study , 1997 .

[23]  W. Crone,et al.  The Effects of Prolonged Ethanol Exposure on the Mechanical Properties of Polyurethane and Silicone Catheters Used for Intravascular Access , 2005, Infection Control & Hospital Epidemiology.

[24]  M. Hansson,et al.  Embolic material generated by multiple aortic crossclamping: a perfusion model with human cadaveric aorta. , 2003, The Journal of thoracic and cardiovascular surgery.

[25]  D. Glower,et al.  Transaortic Endoclamp for Mitral Valve Operation through Right Minithoracotomy in 369 Patients , 2010, Innovations.

[26]  Urs Utzinger,et al.  Microstructural and biomechanical alterations of the human aorta as a function of age and location , 2010, Biomechanics and modeling in mechanobiology.

[27]  J D Humphrey,et al.  Quantification of regional differences in aortic stiffness in the aging human. , 2014, Journal of the mechanical behavior of biomedical materials.

[28]  Wolfgang A. Wall,et al.  Finite deformation frictional mortar contact using a semi‐smooth Newton method with consistent linearization , 2010 .

[29]  Toshiyuki Takagi,et al.  The application of superelastic SMAs in less invasive haemostatic forceps , 2007 .

[30]  A. Chambers,et al.  The Effect of Increasing Clamping Forces on Endothelial and Arterial Wall Damage: An Experimental Study in the Sheep , 1999 .

[31]  A. Pezzella,et al.  Global Aspects of Cardiothoracic Surgery with Focus on Developing Countries , 2010, Asian cardiovascular & thoracic annals.

[32]  G. Holzapfel,et al.  Arterial clamping: finite element simulation and in vivo validation. , 2012, Journal of the mechanical behavior of biomedical materials.

[33]  J. van der Linden,et al.  Institutional report - Cardiopulmonary bypass Do particulate emboli from the ascending aorta in coronary bypass grafting correlate with aortic wall thickness? , 2006 .

[34]  Nature Neurology,et al.  Ischemic stroke in the elderly: an overview of evidence , 2010 .

[35]  Wolfgang A. Wall,et al.  Improved robustness and consistency of 3D contact algorithms based on a dual mortar approach , 2013 .

[36]  Charles A. Taylor,et al.  Effect of Curvature on Displacement Forces Acting on Aortic Endografts: A 3-Dimensional Computational Analysis , 2009, Journal of endovascular therapy : an official journal of the International Society of Endovascular Specialists.

[37]  F. Mohr,et al.  Control of endoaortic clamp position during Port-Access mitral valve operations using transcranial Doppler echography. , 1998, The Annals of thoracic surgery.

[38]  Video-Assisted Minimally Invasive Mitral Valve Surgery: External Aortic Clamp Versus Endoclamp Techniques , 2010, Innovations.

[39]  Wolfgang A. Wall,et al.  A finite deformation mortar contact formulation using a primal–dual active set strategy , 2009 .

[40]  A. Ranasinghe,et al.  Studies of Nonpharmacological Interventions to Reduce Brain Injury , 2011 .

[41]  D. Owen,et al.  Design of simple low order finite elements for large strain analysis of nearly incompressible solids , 1996 .

[42]  D. Stump,et al.  Neurophysiologic monitoring and outcomes in cardiovascular surgery. , 1999, Journal of cardiothoracic and vascular anesthesia.

[43]  W. Chitwood,et al.  Minimally invasive mitral valve repair using transthoracic aortic occlusion. , 1997, The Annals of thoracic surgery.