Sparsity transform k‐t principal component analysis for accelerating cine three‐dimensional flow measurements

Time‐resolved three‐dimensional flow measurements are limited by long acquisition times. Among the various acceleration techniques available, k‐t methods have shown potential as they permit significant scan time reduction even with a single receive coil by exploiting spatiotemporal correlations. In this work, an extension of k‐t principal component analysis is proposed utilizing signal differences between the velocity encodings of three‐directional flow measurements to further compact the signal representation and hence improve reconstruction accuracy. The effect of sparsity transform in k‐t principal component analysis is demonstrated using simulated and measured data of the carotid bifurcation. Deploying sparsity transform for 8‐fold undersampled simulated data, velocity root‐mean‐square errors were found to decrease by 52 ± 14%, 59 ± 11%, and 16 ± 32% in the common, external, and internal carotid artery, respectively. In vivo, errors were reduced by 15 ± 17% in the common carotid artery with sparsity transform. Based on these findings, spatial resolution of three‐dimensional flow measurements was increased to 0.8 mm isotropic resolution with prospective 8‐fold undersampling and sparsity transform k‐t principal component analysis reconstruction. Volumetric data were acquired in 6 min. Pathline visualization revealed details of helical flow patterns partially hidden at lower spatial resolution. Magn Reson Med, 2013. © 2012 Wiley Periodicals, Inc.

[1]  J. Hennig,et al.  3D blood flow characteristics in the carotid artery bifurcation assessed by flow‐sensitive 4D MRI at 3T , 2009, Magnetic resonance in medicine.

[2]  Andreas Stadlbauer,et al.  Accelerated time-resolved three-dimensional MR velocity mapping of blood flow patterns in the aorta using SENSE and k-t BLAST. , 2010, European journal of radiology.

[3]  D. Parker,et al.  Accuracy of phase‐contrast flow measurements in the presence of partial‐volume effects , 1993, Journal of magnetic resonance imaging : JMRI.

[4]  Murat Aksoy,et al.  Time‐resolved 3D quantitative flow MRI of the major intracranial vessels: Initial experience and comparative evaluation at 1.5T and 3.0T in combination with parallel imaging , 2007, Magnetic resonance in medicine.

[5]  Peter Boesiger,et al.  k‐t BLAST and k‐t SENSE: Dynamic MRI with high frame rate exploiting spatiotemporal correlations , 2003, Magnetic resonance in medicine.

[6]  P. Boesiger,et al.  The role of the carotid sinus in the reduction of arterial wall stresses due to head movements--potential implications for cervical artery dissection. , 2009, Journal of biomechanics.

[7]  P. Boesiger,et al.  Electrodynamics and ultimate SNR in parallel MR imaging , 2004, Magnetic resonance in medicine.

[8]  Peter Boesiger,et al.  Optimizing spatiotemporal sampling for k‐t BLAST and k‐t SENSE: Application to high‐resolution real‐time cardiac steady‐state free precession , 2005, Magnetic resonance in medicine.

[9]  I. Marshall Computational simulations and experimental studies of 3D phase‐contrast imaging of fluid flow in carotid bifurcation geometries , 2010, Journal of magnetic resonance imaging : JMRI.

[10]  N J Pelc,et al.  Concomitant gradient terms in phase contrast MR: Analysis and correction , 1998, Magnetic resonance in medicine.

[11]  M. Bernstein,et al.  Comparison of phase‐difference and complex‐difference processing in phase‐contrast MR angiography , 1991, Journal of magnetic resonance imaging : JMRI.

[12]  Robin M Heidemann,et al.  Generalized autocalibrating partially parallel acquisitions (GRAPPA) , 2002, Magnetic resonance in medicine.

[13]  S. Kozerke,et al.  Bayesian multipoint velocity encoding for concurrent flow and turbulence mapping , 2013, Magnetic resonance in medicine.

[14]  Michael Markl,et al.  Improved SNR in phase contrast velocimetry with five‐point balanced flow encoding , 2010, Magnetic resonance in medicine.

[15]  Kevin M. Johnson,et al.  PC HYPR flow: A technique for rapid imaging of contrast dynamics , 2010, Journal of magnetic resonance imaging : JMRI.

[16]  B. Gelfand,et al.  Spatial and spectral heterogeneity of time‐varying shear stress profiles in the carotid bifurcation by phase‐contrast MRI , 2006, Journal of magnetic resonance imaging : JMRI.

[17]  N J Pelc,et al.  Magnetic resonance velocity imaging using a fast spiral phase contrast sequence , 1994, Magnetic resonance in medicine.

[18]  F. Grosveld,et al.  Atherosclerotic Lesion Size and Vulnerability Are Determined by Patterns of Fluid Shear Stress , 2006, Circulation.

[20]  M. Viergever,et al.  MR phase‐contrast flow measurement with limited spatial resolution in small vessels: Value of model‐based image analysis , 1999, Magnetic resonance in medicine.

[21]  M. Karlsson,et al.  Accuracy and reproducibility in phase contrast imaging using SENSE , 2003, Magnetic resonance in medicine.

[22]  Adrian T. Lee,et al.  Three‐Point Phase‐Contrast Velocity Measurements with Increased Velocity‐to‐Noise Ratio , 1995, Magnetic resonance in medicine.

[23]  Tobias Schaeffter,et al.  Highly undersampled phase‐contrast flow measurements using compartment‐based k–t principal component analysis , 2013, Magnetic resonance in medicine.

[24]  F. Korosec,et al.  PC VIPR: a high-speed 3D phase-contrast method for flow quantification and high-resolution angiography. , 2005, AJNR. American journal of neuroradiology.

[25]  P. Boesiger,et al.  SENSE: Sensitivity encoding for fast MRI , 1999, Magnetic resonance in medicine.

[26]  Andreas Sigfridsson,et al.  Four‐dimensional flow MRI using spiral acquisition , 2012, Magnetic resonance in medicine.

[27]  N J Pelc,et al.  Balanced phase‐contrast steady‐state free precession (PC‐SSFP): A novel technique for velocity encoding by gradient inversion , 2003, Magnetic resonance in medicine.

[28]  Feng Huang,et al.  k‐t GRAPPA: A k‐space implementation for dynamic MRI with high reduction factor , 2005, Magnetic resonance in medicine.

[29]  D. Donoho,et al.  Sparse MRI: The application of compressed sensing for rapid MR imaging , 2007, Magnetic resonance in medicine.

[30]  Michael Markl,et al.  Optimized parallel imaging for dynamic PC‐MRI with multidirectional velocity encoding , 2010, Magnetic resonance in medicine.

[31]  Peter Boesiger,et al.  Accelerating cine phase‐contrast flow measurements using k‐t BLAST and k‐t SENSE , 2005, Magnetic resonance in medicine.

[32]  Peter Boesiger,et al.  Accelerated dynamic Fourier velocity encoding by exploiting velocity-spatio-temporal correlations , 2004, Magnetic Resonance Materials in Physics, Biology and Medicine.

[33]  D C Peters,et al.  Phase‐contrast with interleaved undersampled projections , 2000, Magnetic resonance in medicine.

[34]  A J Sederman,et al.  Reducing data acquisition times in phase-encoded velocity imaging using compressed sensing. , 2010, Journal of magnetic resonance.

[35]  Li Feng,et al.  Accelerated phase‐contrast cine MRI using k‐t SPARSE‐SENSE , 2012, Magnetic resonance in medicine.

[36]  P Börnert,et al.  Effects of gradient anisotropy in MRI , 1998, Magnetic resonance in medicine.

[37]  Christof Baltes,et al.  k‐t BLAST reconstruction from non‐Cartesian k‐t space sampling , 2006, Magnetic resonance in medicine.

[38]  B. Wranne,et al.  Temporally resolved 3D phase‐contrast imaging , 1996, Magnetic resonance in medicine.

[39]  Scott E Kasner,et al.  Carotid Artery Diameter in Men and Women and the Relation to Body and Neck Size , 2006, Stroke.

[40]  Kay Nehrke,et al.  k‐t PCA: Temporally constrained k‐t BLAST reconstruction using principal component analysis , 2009, Magnetic resonance in medicine.

[41]  D. Birchall,et al.  Analysis of haemodynamic factors involved in carotid atherosclerosis using computational fluid dynamics. , 2009, The British journal of radiology.

[42]  M. Markl,et al.  Comprehensive 4D velocity mapping of the heart and great vessels by cardiovascular magnetic resonance , 2011, Journal of cardiovascular magnetic resonance : official journal of the Society for Cardiovascular Magnetic Resonance.

[43]  Thoralf Niendorf,et al.  Highly accelerated cardiovascular MR imaging using many channel technology: concepts and clinical applications , 2007, European Radiology.

[44]  P. Walker,et al.  Semiautomated method for noise reduction and background phase error correction in MR phase velocity data , 1993, Journal of magnetic resonance imaging : JMRI.

[45]  Jon-Fredrik Nielsen,et al.  Referenceless phase velocity mapping using balanced SSFP , 2009, Magnetic resonance in medicine.