Author response to: Review of the first 1502 cases of ECG‐ST waveform analysis during labour in a teaching hospital

OBJECTIVE To assess the impact of introduction of the STAN monitoring system. STUDY DESIGN Prospective observational study. SETTING Tertiary referral labour ward, St George's Hospital, London. POPULATION High-risk term pregnancies. METHODS We report all consecutive cases of intrapartum monitoring using the STAN S 21 fetal heart monitor. Cases with adverse neonatal outcome were evaluated in relation to the ST waveform analysis and cardiotocography (CTG). MAIN OUTCOME MEASURES Cord artery metabolic acidosis, neonatal encephalopathy (NNE) and reasons behind cases with poor outcome. RESULTS Between 2002 and 2005, there were 1502 women monitored by STAN. Based on combined STAN analysis in the 1502 women, action was indicated in 358 women (23.8%), while in 1108 women (73.8%) no action was indicated. Traces were not interpretable in 36 women (2.4%). Of the 836 cases (55.7%) where cord blood gases were available, there were 23 cases (2.8%) of metabolic acidosis and 16 of these (70%) were identified by STAN. Overall, there were 14 cases of NNE monitored by STAN. Retrospective analysis of these highlights human errors, such as poor CTG interpretation, delay in taking appropriate action and not following the guidelines. CONCLUSIONS Our experience suggests the need for more intense training on interpretation of CTG and strict adherence to guidelines.

[1]  M. Vääräsmäki,et al.  A comparison of intrapartum automated fetal electrocardiography and conventional cardiotocography—a randomised controlled study , 2006, BJOG : an international journal of obstetrics and gynaecology.

[2]  O. Apantaku,et al.  Review of the first 1502 cases of ECG‐ST waveform analysis during labour in a teaching hospital , 2008, BJOG : an international journal of obstetrics and gynaecology.

[3]  K. Maršál,et al.  Fetal electrocardiogram: ST waveform analysis in intrapartum surveillance , 2007, BJOG : an international journal of obstetrics and gynaecology.

[4]  I. Amer-Wåhlin,et al.  A comparison of intrapartum automated fetal electrocardiography and conventional cardiotocography—assessment of outcome , 2006, BJOG : an international journal of obstetrics and gynaecology.

[5]  S. Blad,et al.  Fetal scalp pH and ST analysis of the fetal ECG as an adjunct to CTG. A multi-center, observational study , 2004, Journal of perinatal medicine.