A depth-of-interaction encoding PET detector module with dual-ended readout using large-area silicon photomultiplier arrays

The performance of a depth-of-interaction (DOI) encoding PET detector module with dual-ended readout of LYSO scintillator arrays using large-area SiPM arrays was evaluated. Each SiPM array, with a surface area of 50.2  ×  50.2 mm2, consists of 12  ×  12 C-series SiPMs from SensL (SensL, Inc). The LYSO array, with a total size of 46  ×  46 mm2 and a pitch size of 1.0 mm, consists of a 46  ×  46 array of 0.945  ×  0.945  ×  20 mm3 polished LYSO crystals, separated by Toray reflector. Custom front-end electronics were designed to reduce the 288 SiPM signals of one detector module to nine signals, eight for position information and 1 for timing information. Schottky diodes were used to block noise from SiPMs that did not detect a significant number of scintillation photons following a gamma interaction. Measurements of noise, signal, signal-to-noise ratio, energy resolution and flood histogram quality were obtained at different bias voltages (26.0 to 31.0 V in 0.5 V intervals) and at two temperatures (5 °C and 20 °C). Clear acrylic plates, 2.0 mm thick, were used as light guides to spread the scintillation photons. Timing resolution, depth of interaction resolution, and the effect of event rate on detector performance were measured at the bias voltage determined to be optimal for the flood histograms. Performance obtained with and without the noise-blocking Shottky diodes was also compared. The results showed that all crystals in the LYSO array can be clearly resolved, and performance improved when using diodes to block noise, and at the lower temperature. The average energy resolution, flood histogram quality, timing resolution and DOI resolution were 23.8%  ±  2.0%, 1.54  ±  0.17, 1.78  ±  0.09 ns and 2.81  ±  0.13 mm respectively, obtained at a bias voltage of 30.0 V and a temperature of 5 °C using the diode readout method. The event rate experiments showed that the flood histogram and energy resolution of the detector were not significantly degraded for an event rate of up to 150 000 counts s-1.

[1]  C. Thompson,et al.  Evaluation of a 16:3 Signal Multiplexor to Acquire Signals From a SPM Array With Dual and Single Layer LYSO Crystal Blocks , 2011, IEEE Transactions on Nuclear Science.

[2]  S. Cherry,et al.  A Simple Capacitive Charge-Division Readout for Position-Sensitive Solid-State Photomultiplier Arrays , 2013, IEEE Transactions on Nuclear Science.

[3]  L. Pierce,et al.  Characterization of highly multiplexed monolithic PET / gamma camera detector modules , 2018, Physics in medicine and biology.

[4]  V. Popov Advanced data readout technique for Multianode Position Sensitive Photomultiplier Tube applicable in radiation imaging detectors , 2011 .

[5]  Carl Jackson,et al.  Evaluation of Matrix9 silicon photomultiplier array for small-animal PET. , 2015, Medical physics.

[6]  Guido Germano,et al.  Recent Advances and Future Progress in PET Instrumentation. , 2016, Seminars in nuclear medicine.

[7]  Gerald Entine,et al.  Depth of interaction resolution measurements for a high resolution PET detector using position sensitive avalanche photodiodes , 2006, Physics in medicine and biology.

[8]  S. Cherry,et al.  Comparison of large-area position-sensitive solid-state photomultipliers for small animal PET , 2012, Physics in medicine and biology.

[9]  Jinyi Qi,et al.  A Prototype PET Scanner with DOI-Encoding Detectors , 2008, Journal of Nuclear Medicine.

[10]  Simon R. Cherry,et al.  Characterization of Large-Area SiPM Array for PET Applications , 2016, IEEE Transactions on Nuclear Science.

[11]  G. Borghi,et al.  A 32 mm  ×  32 mm  ×  22 mm monolithic LYSO:Ce detector with dual-sided digital photon counter readout for ultrahigh-performance TOF-PET and TOF-PET/MRI , 2016, Physics in medicine and biology.

[12]  S. Cherry,et al.  MicroPET II: design, development and initial performance of an improved microPET scanner for small-animal imaging. , 2003, Physics in medicine and biology.

[13]  S. Cherry,et al.  Signal and noise properties of position-sensitive avalanche photodiodes , 2011, Physics in medicine and biology.

[14]  Yu Wang,et al.  A modular low dead-time coincidence system for high-resolution PET cameras , 2002 .

[15]  Simon R. Cherry,et al.  Effects of reflector and crystal surface on the performance of a depth-encoding PET detector with dual-ended readout. , 2014, Medical physics.

[16]  Jae Sung Lee,et al.  Depth-of-interaction measurement in a single-layer crystal array with a single-ended readout using digital silicon photomultiplier , 2015, Physics in medicine and biology.

[17]  D. Renker,et al.  Development of a novel depth of interaction PET detector using highly multiplexed G-APD cross-strip encoding. , 2014, Medical physics.

[18]  Gangadhar Jaliparthi,et al.  Evaluation of Imaging Modules Based on SensL Array SB-8 for Nuclear Medicine Applications , 2013, IEEE Transactions on Nuclear Science.

[19]  S. Cherry,et al.  High-resolution PET detector design: modelling components of intrinsic spatial resolution , 2005, Physics in medicine and biology.

[20]  Kanai Shah,et al.  Continuous depth-of-interaction encoding using phosphor-coated scintillators , 2009, Physics in medicine and biology.

[21]  Yiping Shao,et al.  Development of compact DOI-measurable PET detectors for simultaneous PET/MR Imaging , 2015, EJNMMI Physics.

[22]  Tom K Lewellen,et al.  Recent developments in PET detector technology , 2008, Physics in medicine and biology.

[23]  Yonggang Wang,et al.  Physical Properties of LYSO Scintillator for NN-PET Detectors , 2009, 2009 2nd International Conference on Biomedical Engineering and Informatics.

[24]  T. Lewellen,et al.  cMiCE: a high resolution animal PET using continuous LSO with a statistics based positioning scheme☆ , 2002 .

[25]  Y. Shao,et al.  Development of a prototype PET scanner with depth-of-interaction measurement using solid-state photomultiplier arrays and parallel readout electronics , 2014, Physics in medicine and biology.

[26]  Michael Ljungberg,et al.  Evaluation of inter-departmental variability of ejection fraction and cardiac volumes in myocardial perfusion scintigraphy using simulated data , 2015, EJNMMI Physics.

[27]  R. Lecomte,et al.  Scintillation light emission studies of LSO scintillators , 1998, 1998 IEEE Nuclear Science Symposium Conference Record. 1998 IEEE Nuclear Science Symposium and Medical Imaging Conference (Cat. No.98CH36255).

[28]  S. Cherry,et al.  Evaluation of linearly-graded SiPMs for high resolution small-animal PET , 2015 .