Brain temperature measurement: A study of in vitro accuracy and stability of smart catheter temperature sensors

The injured brain is vulnerable to increases in temperature after severe head injury. Therefore, accurate and reliable measurement of brain temperature is important to optimize patient outcome. In this work, we have fabricated, optimized and characterized temperature sensors for use with a micromachined smart catheter for multimodal intracranial monitoring. Developed temperature sensors have resistance of 100.79 ± 1.19Ω and sensitivity of 67.95 mV/°C in the operating range from15–50°C, and time constant of 180 ms. Under the optimized excitation current of 500 μA, adequate signal-to-noise ratio was achieved without causing self-heating, and changes in immersion depth did not introduce clinically significant errors of measurements (<0.01°C). We evaluated the accuracy and long-term drift (5 days) of twenty temperature sensors in comparison to two types of commercial temperature probes (USB Reference Thermometer, NIST-traceable bulk probe with 0.05°C accuracy; and IT-21, type T type clinical microprobe with guaranteed 0.1°C accuracy) under controlled laboratory conditions. These in vitro experimental data showed that the temperature measurement performance of our sensors was accurate and reliable over the course of 5 days. The smart catheter temperature sensors provided accuracy and long-term stability comparable to those of commercial tissue-implantable microprobes, and therefore provide a means for temperature measurement in a microfabricated, multimodal cerebral monitoring device.

[1]  L. Heytens,et al.  Intracerebral temperature monitoring in severely head injured patients , 2005, Acta Neurochirurgica.

[2]  Vikramaditya G. Yadav,et al.  Cell and protein compatibility of parylene-C surfaces. , 2007, Langmuir : the ACS journal of surfaces and colloids.

[3]  K. Polderman,et al.  Mechanisms of action, physiological effects, and complications of hypothermia , 2009, Critical care medicine.

[4]  James L. Thomas,et al.  Precision Resistors and Their Measurement , 1948 .

[5]  Laura Mcilvoy,et al.  Comparison of brain temperature to core temperature: a review of the literature. , 2004, The Journal of neuroscience nursing : journal of the American Association of Neuroscience Nurses.

[6]  R. Normann,et al.  Thermal Impact of an Active 3-D Microelectrode Array Implanted in the Brain , 2007, IEEE Transactions on Neural Systems and Rehabilitation Engineering.

[7]  Jungyoup Han,et al.  A flexible polymer tube lab-chip integrated with microsensors for smart microcatheter , 2008, Biomedical microdevices.

[8]  M. Diringer,et al.  Temperature management in acute neurologic disorders. , 2008, Neurologic clinics.

[9]  N. Badjatia,et al.  Hyperthermia and fever control in brain injury , 2009, Critical care medicine.

[10]  Francis A. Duck,et al.  Physical properties of tissue : a comprehensive reference book , 1990 .

[11]  M. Shigemori,et al.  OPTIMAL TEMPERATURE FOR THE MANAGEMENT OF SEVERE TRAUMATIC BRAIN INJURY: EFFECT OF HYPOTHERMIA ON INTRACRANIAL PRESSURE, SYSTEMIC AND INTRACRANIAL HEMODYNAMICS, AND METABOLISM , 2007, Neurosurgery.

[12]  Chong H. Ahn,et al.  Potential of a simple lab-on-a-tube for point-of-care measurements of multiple analytes. , 2010, Lab on a chip.

[13]  Jungyoup Han,et al.  Flexible biosensors on spirally rolled micro tube for cardiovascular in vivo monitoring. , 2007, Biosensors & bioelectronics.

[14]  R. Bellomo,et al.  Therapeutic hypothermia: benefits, mechanisms and potential clinical applications in neurological, cardiac and kidney injury. , 2011, Injury.

[15]  Peng Liu,et al.  A Flexible Flow Sensor System and Its Characteristics for Fluid Mechanics Measurements , 2009, Sensors.

[16]  A. Dajani,et al.  PREVENTION OF BACTERIAL ENDOCARDITIS: RECOMMENDATIONS BY THE AMERICAN HEART ASSOCIATION , 1991, GMDA bulletin.

[17]  K. Fountas,et al.  Intracranial temperature , 2004, Neurocritical care.

[18]  Daryl R Kipke,et al.  The insulation performance of reactive parylene films in implantable electronic devices. , 2009, Biomaterials.

[19]  C. Kurth,et al.  ON KEEPING A COOL HEAD , 1978, The Lancet.

[20]  C. Deyne Therapeutic hypothermia and traumatic brain injury. , 2010 .

[21]  Chong H Ahn,et al.  A novel lab-on-a-tube for multimodality neuromonitoring of patients with traumatic brain injury (TBI). , 2009, Lab on a chip.

[22]  Béla G. Lipták Process measurement and analysis , 2003 .

[23]  C. Kurth,et al.  Brain cooling efficiency with pH-stat and alpha-stat cardiopulmonary bypass in newborn pigs. , 1997, Circulation.

[24]  C. Robertson,et al.  Brain temperature exceeds systemic temperature in head-injured patients. , 1998, Critical care medicine.

[25]  C. Kurth,et al.  Infant Cardiac SurgeryKeeping a Cool Head , 2000 .