Angular dependence of dose sensitivity of surface diodes.

Commercially available surface diodes are found to have as great as +/-12% change in sensitivity with the angle of incidence of radiation. This work is a study of the cause of angular dependence in diode sensitivity and how it can be decreased. A number of different surface diodes were used in these measurements: A commercially available diode and four prototype diodes. A number of the diodes were constructed with the silicon chip, the die, mounted on a circuit board that had a plane of copper on its back side. These diodes had angular dependence of sensitivity as great as +/- 10%. It was hypothesized that the copper plane on the circuit board was the cause of the anisotropy in sensitivity of the diodes. To test this hypothesis, diodes with a new design [Patent No. 61/035,257 (pending)], without a copper back plane, were fabricated and characterized in this work. These diodes were found to have the following characteristics: A dependence on incident angle of radiation of +/- 3.6%; after 10 kGy of pre-irradiation, a 1.6% change in sensitivity for a 260-fold change in dose per pulse; an areal density of 0.08 g/cm2.

[1]  The effect of dose rate dependence of p-type silicon detectors on linac relative dosimetry. , 1997, Medical physics.

[2]  B J Mijnheer,et al.  In vivo dosimetry during external photon beam radiotherapy. , 1999, International journal of radiation oncology, biology, physics.

[3]  Limitations of silicon diodes for clinical electron dosimetry. , 2006, Radiation protection dosimetry.

[4]  I J Das,et al.  Dose perturbations at interfaces in photon beams. , 1987, Medical physics.

[5]  D. Rogers,et al.  AAPM's TG-51 protocol for clinical reference dosimetry of high-energy photon and electron beams. , 1999, Medical physics.

[6]  D Marre,et al.  Comparison of p-type commercial electron diodes for in vivo dosimetry. , 2003, Medical physics.

[7]  G. Rikner,et al.  General specifications for silicon semiconductors for use in radiation dosimetry. , 1987, Physics in medicine and biology.

[8]  P. Jursinic,et al.  Characterization of optically stimulated luminescent dosimeters, OSLDs, for clinical dosimetric measurements. , 2007, Medical physics.

[9]  J Cygler,et al.  Evaluation of a dual bias dual metal oxide-silicon semiconductor field effect transistor detector as radiation dosimeter. , 1994, Medical physics.

[10]  D. R. White,et al.  Tissue substitutes in experimental radiation physics. , 1978, Medical physics.

[11]  Plato C. Lee,et al.  Patient dosimetry quality assurance program with a commercial diode system. , 1994, International journal of radiation oncology, biology, physics.

[12]  T. Jordan Megavoltage X-ray beams: 2-50 MV. , 1996, BJR supplement.

[13]  Linearity with dose rate of low resistivity p-type silicon semiconductor detectors , 1993 .

[14]  T Knöös,et al.  Comparative dosimetry of diode and diamond detectors in electron beams for intraoperative radiation therapy. , 2000, Medical physics.

[15]  P. Jursinic,et al.  A 2-D diode array and analysis software for verification of intensity modulated radiation therapy delivery. , 2003, Medical physics.

[16]  G. Marinello Radiothermoluminescent Dosimeters and Diodes , 2007 .

[17]  A. Morgan,et al.  Evaluation of in vivo dose measurements for patients undergoing electron boost treatments. , 2001, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[18]  Commissioning a p-type silicon diode for use in clinical electron beams. , 1999, Medical physics.

[19]  George X. Ding,et al.  Determination of percentage depth-dose curves for electron beams using different types of detectors. , 2001, Medical physics.

[20]  P. Jursinic Changes in incident photon fluence of 6 and 18 MV x rays caused by blocks and block trays. , 1999, Medical physics.

[21]  N. Jornet,et al.  In vivo dosimetry: intercomparison between p-type based and n-type based diodes for the 16-25 MV energy range. , 2000, Medical physics.

[22]  A. Beddoe,et al.  Evaluation of a PTW diamond detector for electron beam measurements , 1993 .

[23]  Beverley Rader Lugo,et al.  Clinical Experience , 2009 .

[24]  P. Jursinic Implementation of an in vivo diode dosimetry program and changes in diode characteristics over a 4-year clinical history. , 2001, Medical physics.

[25]  Correction factors for low perturbation in vivo diodes used in the determination of entrance doses in high energy photon beams. , 2007, Medical physics.

[26]  A method to improve the effectiveness of diode in vivo dosimetry. , 1998, Medical physics.

[27]  Timothy C Zhu,et al.  Modeling the instantaneous dose rate dependence of radiation diode detectors. , 2003, Medical physics.

[28]  G. Rikner,et al.  Radiation damage induced dose rate non-linearity in an n-type silicon detector. , 1984, Acta radiologica. Oncology.

[29]  R K Ten Haken,et al.  Practical methods of electron depth-dose measurement compared to use of the NACP design chamber in water. , 1987, Medical physics.

[30]  A. Brahme,et al.  Comparative dosimetry in narrow high-energy photon beams. , 2000, Physics in medicine and biology.

[31]  Timothy C Zhu,et al.  Dose rate and SDD dependence of commercially available diode detectors. , 2004, Medical physics.

[32]  R. Yaparpalvi,et al.  Clinical experience with routine diode dosimetry for electron beam radiotherapy. , 2000, International journal of radiation oncology, biology, physics.

[33]  P. Jursinic Measurement of head scatter factors of linear accelerators with columnar miniphantoms. , 2006, Medical physics.

[34]  Correlation between temperature and dose rate dependence of semiconductor response; influence of accumulated dose. , 1990, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[35]  G. Rikner,et al.  Effects of radiation damage on p-type silicon detectors , 1983 .

[36]  Frank Herbert Attix,et al.  Introduction to Radiological Physics and Radiation Dosimetry: Attix/Introduction , 2007 .

[37]  I. Das,et al.  Backscatter dose perturbation at high atomic number interfaces in megavoltage photon beams. , 1989, Medical physics.