On the determination of total mercury in natural gases using the amalgamation technique and cold vapour atomic absorption spectrometry
暂无分享,去创建一个
Studies on the use of noble metal collectors for the amalgamation of mercury from natural gas prior to thermal desorption and determination by cold vapour atomic absorption spectrometry have been performed. It is shown that collectors filled with gold/platinum wire are more efficient than those packed with gold or silver wire. Quantitative sampling efficiencies are obtained for Hg0 and (CH3)2Hg added to natural gas when the gold/platinum collector is heated to 80 °C, at sampling rates < 2 l min–1 and for volumes of at least 10 l. The precision for field measurements was established to be between 8 and 15% relative standard deviation, and the detection limit for the developed procedure is calculated to be 30 ng m–3 for a 10 l sample (3s criterion).
[1] Hideyoshi Morita,et al. Atomic fluorescence spectrometry of mercury: principles and developments , 1995 .
[2] S. Järås,et al. Catalytic Materials for High-Temperature Combustion , 1993 .
[3] P. Minkkinen. Monitoring the precision of routine analyses by using duplicate determinations , 1986 .
[4] J. Hoste,et al. Comparison of the collection and desorption efficiency of activated charcoal, silver, and gold for the determination of vapor-phase atmospheric mercury , 1985 .