Analysis of Explanted Magnetically Controlled Growing Rods From Seven UK Spinal Centers

Study Design. Analysis of explanted MAGnetic Expansion Control (MAGEC) growing rods. Objective. To analyze explanted MAGEC rods used in management of early onset scoliosis and identify the mode of failure in such cases. Summary of Background Data. Magnetically controlled growing rods are increasingly used as the option of choice for early onset scoliosis. However, being more complex than conventional growing rods they are perhaps more likely to succumb to multifarious failure modes. In addition, metallosis has been reported around failed MAGEC rods. Methods. Explanted MAGEC rods from seven UK spinal centers were obtained for independent analysis. Thirty-four MAGEC rods, from 18 children, explanted for reasons including failure of rod lengthening and maximum rod distraction reached, were cut open to allow internal components to be evaluated and assessed. Results. Externally, all MAGEC rods showed localized marks, which were termed “growth marks” as they indicated growth of the rod in vivo, on the extending bar component. After cutting open, titanium wear debris was found inside all 34 (100%) MAGEC rods. Ninety-one percent (31/34) of MAGEC rods showed measurable wear of the extending bar, towards the magnet end. Substantial damage to the radial bearing was seen inside 74% (25/34) of MAGEC rods while O-ring seal failure was seen in 53% (18/34) of cases. In 44% (15/34) of MAGEC rods the drive pin was fractured but this was felt to be an effect of rod failure, not a cause. Conclusion. The combination of high volumes of titanium wear debris alongside O-ring seal damage likely accounts for the metallosis reported clinically around some MAGEC rods. Based on this explant data, a failure mechanism in MAGEC rods due to the natural off axis loading in the spine was proposed. This is the largest data set reporting a complete analysis of explanted MAGEC rods to date. Level of Evidence: 4

[1]  G. Eschweiler,et al.  Multicenter Study. , 2019, Deutsches Arzteblatt international.

[2]  E. Roughead,et al.  Heart failure after conventional metal-on-metal hip replacements , 2016, Acta Orthopaedica.

[3]  S. Evans,et al.  Metallosis following implantation of magnetically controlled growing rods in the treatment of scoliosis: a case series. , 2016, The bone & joint journal.

[4]  I. Helenius,et al.  Magnetically controlled Growing Rods for Early-onset Scoliosis: A Multicenter Study of 23 Cases With Minimum 2 years Follow-up , 2016, Spine.

[5]  J. Wilson-Macdonald,et al.  The use of magnetically-controlled growing rods to treat children with early-onset scoliosis: early radiological results in 19 children. , 2016, The bone & joint journal.

[6]  T. Joyce,et al.  Retrospective cohort study of the performance of the Pinnacle metal on metal (MoM) total hip replacement: a single-centre investigation in combination with the findings of a national retrieval centre , 2016, BMJ Open.

[7]  A. Clarke,et al.  Actuator pin fracture in magnetically controlled growing rods: two cases. , 2016, The spine journal : official journal of the North American Spine Society.

[8]  S. Ahuja,et al.  Magnetic controlled growing rods for early-onset scoliosis: a 4-year follow-up. , 2016, The spine journal : official journal of the North American Spine Society.

[9]  T. Joyce,et al.  Shorter, rough trunnion surfaces are associated with higher taper wear rates than longer, smooth trunnion surfaces in a contemporary large head metal‐on‐metal total hip arthroplasty system , 2015, Journal of orthopaedic research : official publication of the Orthopaedic Research Society.

[10]  J. Cheung,et al.  Special Article: Update on the Magnetically Controlled Growing Rod: Tips and Pitfalls , 2015, Journal of orthopaedic surgery.

[11]  T. Joyce,et al.  Determining material loss from the femoral stem trunnion in hip arthroplasty using a coordinate measuring machine , 2015, Proceedings of the Institution of Mechanical Engineers. Part H, Journal of engineering in medicine.

[12]  Andrew J. Sims,et al.  The MAGEC System for Spinal Lengthening in Children with Scoliosis: A NICE Medical Technology Guidance , 2014, Applied Health Economics and Health Policy.

[13]  S. Kurtz,et al.  Which Design and Biomaterial Factors Affect Clinical Wear Performance of Total Disc Replacements? A Systematic Review , 2014, Clinical orthopaedics and related research.

[14]  S. Pankuweit,et al.  Cobalt intoxication diagnosed with the help of Dr House , 2014, The Lancet.

[15]  J. Rawlinson,et al.  Growth Guidance System for Early-Onset Scoliosis: Comparison of Experimental and Retrieval Wear , 2013, Spine.

[16]  H. Noordeen,et al.  Early results of a remotely-operated magnetic growth rod in early-onset scoliosis. , 2013, The bone & joint journal.

[17]  T. Joyce,et al.  Topographical analysis of the femoral components of ex vivo total knee replacements , 2013, Journal of Materials Science: Materials in Medicine.

[18]  T. J. Joyce,et al.  Taper junction failure in large-diameter metal-on-metal bearings , 2012, Bone & joint research.

[19]  B. Morrey Accelerating failure rate of the ASR total hip replacement , 2012 .

[20]  C. Cacou,et al.  Monitoring the introduction and performance of a joint replacement: the United Kingdom metal-on-metal alert. , 2011, The Journal of bone and joint surgery. American volume.

[21]  T. Joyce,et al.  Volumetric wear assessment of failed metal-on-metal hip resurfacing prostheses , 2011 .

[22]  D. Skaggs,et al.  Lengthening of Dual Growing Rods and the Law of Diminishing Returns , 2011, Spine.

[23]  O. Boachie-Adjei,et al.  Complications of growing-rod treatment for early-onset scoliosis: analysis of one hundred and forty patients. , 2010, The Journal of bone and joint surgery. American volume.

[24]  T J Joyce,et al.  Early failure of metal-on-metal bearings in hip resurfacing and large-diameter total hip replacement: A consequence of excess wear. , 2010, The Journal of bone and joint surgery. British volume.

[25]  T. Joyce Analysis of the Mechanism of Fracture of Silicone Metacarpophalangeal Prostheses , 2009, The Journal of hand surgery, European volume.

[26]  T. Joyce,et al.  Tribological analysis of failed resurfacing hip prostheses and comparison with clinical data , 2009 .

[27]  T. Joyce Examination of failed ex vivo metal-on-metal metatarsophalangeal prosthesis and comparison with theoretically determined lubrication regimes , 2007 .

[28]  P. Cripton,et al.  Wear and Corrosion in Retrieved Thoracolumbar Posterior Internal Fixation , 2006, Spine.

[29]  D. Dowson Tribological principles in metal-on-metal hip joint design , 2006, Proceedings of the Institution of Mechanical Engineers. Part H, Journal of engineering in medicine.

[30]  O. Boachie-Adjei,et al.  Dual Growing Rod Technique for the Treatment of Progressive Early-Onset Scoliosis: A Multicenter Study , 2005, Spine.

[31]  J. Pais-Ribeiro,et al.  General anesthesia, surgery and hospitalization in children and their effects upon cognitive, academic, emotional and sociobehavioral development – a review , 2004, Paediatric anaesthesia.

[32]  A. Doherty,et al.  Wear debris from hip or knee replacements causes chromosomal damage in human cells in tissue culture. , 2004, The Journal of bone and joint surgery. British volume.

[33]  P. Paavolainen,et al.  Cancer Risk After Metal on Metal and Polyethylene on Metal Total Hip Arthroplasty , 1996, Clinical orthopaedics and related research.