Epid‐based in vivo dose verification for lung stereotactic treatments delivered with multiple breath‐hold segmented volumetric modulated arc therapy

Abstract We evaluated an EPID‐based in‐vivo dosimetry (IVD) method for the dose verification and the treatment reproducibility of lung SBRT‐VMAT treatments in clinical routine. Ten patients with lung metastases treated with Elekta VMAT technique were enrolled. All patients were irradiated in five consecutive fractions, with total doses of 50 Gy. Set‐up was carried out with the Elekta stereotactic body frame. Eight patients were simulated and treated using the Active Breath Control (ABC) system, a spirometer enabling patients to maintain a breath‐hold at a predetermined lung volume. Two patients were simulated and treated in free‐breathing using an abdominal compressor. IVD was performed using the SOFTDISO software. IVD tests were evaluated by means of (a) ratio R between daily in‐vivo isocenter dose and planned dose and (b) γ‐analysis between EPID integral portal images in terms of percentage of points with γ‐value smaller than one (γ %) and mean γ‐values (γ mean) using a 3%(global)/3 mm criteria. Alert criteria of ±5% for R ratio, γ % < 90%, and γ mean > 0.67 were chosen. 50 transit EPID images were acquired. For the patients treated with ABC spirometer, the results reported a high level of accuracy in dose delivery with 100% of tests within ±5%. The γ‐analysis showed a mean value of γ mean equal to 0.21 (range: 0.04–0.56) and a mean γ % equal to 96.9 (range: 78–100). Relevant discrepancies were observed only for the two patients treated without ABC, mainly due to a blurring dose effect due to residual respiratory motion. Our method provided a fast and accurate procedure in clinical routine for verifying delivered dose as well as for detecting errors.

[1]  Marcel van Herk,et al.  Impact of daily anatomical changes on EPID-based in vivo dosimetry of VMAT treatments of head-and-neck cancer. , 2015, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[2]  P. Lambin,et al.  A literature review of electronic portal imaging for radiotherapy dosimetry. , 2008, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[3]  L. Azario,et al.  Dynamic conformal arc therapy: transmitted signal in vivo dosimetry. , 2008, Medical physics.

[4]  Pietro Mancosu,et al.  Critical appraisal of volumetric modulated arc therapy in stereotactic body radiation therapy for metastases to abdominal lymph nodes. , 2009, International journal of radiation oncology, biology, physics.

[5]  Ying Xiao,et al.  Motion management strategies and technical issues associated with stereotactic body radiotherapy of thoracic and upper abdominal tumors: A review from NRG oncology , 2017, Medical physics.

[6]  P M McCowan,et al.  An in vivo dose verification method for SBRT-VMAT delivery using the EPID. , 2015, Medical physics.

[7]  Savino Cilla,et al.  Correlation functions for Elekta aSi EPIDs used as transit dosimeter for open fields , 2010, Journal of applied clinical medical physics.

[8]  Ben Mijnheer,et al.  In vivo dosimetry in external beam radiotherapy. , 2013, Medical physics.

[9]  E. Ford,et al.  Safety Considerations in Stereotactic Body Radiation Therapy. , 2017, Seminars in radiation oncology.

[10]  Savino Cilla,et al.  Routine EPID in-vivo dosimetry in a reference point for conformal radiotherapy treatments. , 2015, Physics in medicine and biology.

[11]  Savino Cilla,et al.  Generalized EPID calibration for in vivo transit dosimetry. , 2011, Physica medica : PM : an international journal devoted to the applications of physics to medicine and biology : official journal of the Italian Association of Biomedical Physics.

[12]  Savino Cilla,et al.  In-vivo portal dosimetry by an ionization chamber. , 2005, Physica medica : PM : an international journal devoted to the applications of physics to medicine and biology : official journal of the Italian Association of Biomedical Physics.

[13]  Lei Xing,et al.  Stereotactic body radiation therapy in multiple organ sites. , 2007, Journal of clinical oncology : official journal of the American Society of Clinical Oncology.

[14]  Philip M Evans,et al.  Feasibility of the use of the Active Breathing Co ordinator (ABC) in patients receiving radical radiotherapy for non-small cell lung cancer (NSCLC). , 2009, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[15]  Marcel van Herk,et al.  Online 3D EPID-based dose verification: Proof of concept. , 2016, Medical physics.

[16]  Savino Cilla,et al.  In patient dose reconstruction using a cine acquisition for dynamic arc radiation therapy , 2009, Medical & Biological Engineering & Computing.

[17]  B Mijnheer,et al.  Error detection during VMAT delivery using EPID-based 3D transit dosimetry. , 2018, Physica medica : PM : an international journal devoted to the applications of physics to medicine and biology : official journal of the Italian Association of Biomedical Physics.

[18]  L. Azario,et al.  An in-vivo dosimetry procedure for Elekta step and shoot IMRT. , 2014, Physica medica : PM : an international journal devoted to the applications of physics to medicine and biology : official journal of the Italian Association of Biomedical Physics.

[19]  M. Schell,et al.  Stereotactic body radiation therapy: the report of AAPM Task Group 101. , 2010, Medical physics.

[20]  M van Herk,et al.  Catching errors with in vivo EPID dosimetry. , 2010, Medical physics.

[21]  Savino Cilla,et al.  A validation study of a dedicated software for an automated in vivo dosimetry control in radiotherapy , 2018, Medical & Biological Engineering & Computing.

[22]  L Grimaldi,et al.  Portal dose measurements by a 2D array. , 2007, Physica medica : PM : an international journal devoted to the applications of physics to medicine and biology : official journal of the Italian Association of Biomedical Physics.

[23]  Timothy D. Solberg,et al.  Quality and safety considerations in stereotactic radiosurgery and stereotactic body radiation therapy: Executive summary , 2012, Practical radiation oncology.

[24]  T. Solberg,et al.  Quality and safety in stereotactic radiosurgery and stereotactic body radiation therapy: can more be done? , 2011, Journal of radiosurgery and SBRT.

[25]  Martha M Matuszak,et al.  Volumetric modulated arc therapy for delivery of hypofractionated stereotactic lung radiotherapy: A dosimetric and treatment efficiency analysis. , 2010, Radiotherapy and Oncology.

[26]  S Derreumaux,et al.  Lessons from recent accidents in radiation therapy in France. , 2008, Radiation protection dosimetry.

[27]  U Oppitz,et al.  Stereotactic radiotherapy of extracranial targets: CT-simulation and accuracy of treatment in the stereotactic body frame. , 2000, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[28]  Peter M McCowan,et al.  In vivo dosimetry for lung radiotherapy including SBRT. , 2017, Physica medica : PM : an international journal devoted to the applications of physics to medicine and biology : official journal of the Italian Association of Biomedical Physics.

[29]  S. Cilla,et al.  Quasi real time in vivo dosimetry for VMAT. , 2014, Medical physics.

[30]  Todsaporn Fuangrod,et al.  First Experience With Real-Time EPID-Based Delivery Verification During IMRT and VMAT Sessions. , 2015, International journal of radiation oncology, biology, physics.

[31]  A. Piermattei,et al.  Linac-based extracranial radiosurgery with Elekta volumetric modulated arc therapy and an anatomy-based treatment planning system: Feasibility and initial experience. , 2016, Medical dosimetry : official journal of the American Association of Medical Dosimetrists.

[32]  Clinical results of an EPID-based in-vivo dosimetry method for pelvic cancers treated by intensity-modulated radiation therapy. , 2014, Physica medica : PM : an international journal devoted to the applications of physics to medicine and biology : official journal of the Italian Association of Biomedical Physics.

[33]  Savino Cilla,et al.  Initial clinical experience with Epid-based in-vivo dosimetry for VMAT treatments of head-and-neck tumors. , 2016, Physica medica : PM : an international journal devoted to the applications of physics to medicine and biology : official journal of the Italian Association of Biomedical Physics.

[34]  Savino Cilla,et al.  Breast in vivo dosimetry by EPID , 2010, Journal of applied clinical medical physics.

[35]  Savino Cilla,et al.  A National project for in vivo dosimetry procedures in radiotherapy: First results , 2012 .

[36]  Savino Cilla,et al.  Complexity index (COMIX) and not type of treatment predicts undetected errors in radiotherapy planning and delivery. , 2008, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[37]  Ping Xia,et al.  Tolerance limits and methodologies for IMRT measurement‐based verification QA , 2018, Medical physics.

[38]  M. V. van Herk,et al.  In vivo portal dosimetry for head-and-neck VMAT and lung IMRT: linking γ-analysis with differences in dose-volume histograms of the PTV. , 2014, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[39]  Savino Cilla,et al.  Real time transit dosimetry for the breath-hold radiotherapy technique: An initial experience , 2008, Acta oncologica.

[40]  Marco Esposito,et al.  Characterization of EPID software for VMAT transit dosimetry , 2018, Australasian Physical & Engineering Sciences in Medicine.

[41]  V. Valentini,et al.  Volumetric intensity modulated arc therapy for stereotactic body radiosurgery in oligometastatic breast and gynecological cancers: feasibility and clinical results. , 2014, Oncology reports.

[42]  J. Wong,et al.  The use of active breathing control (ABC) to reduce margin for breathing motion. , 1999, International journal of radiation oncology, biology, physics.