Determination of arsenic in homeopathic drugs by bromide volatilization-inductively coupled plasma-atomic emission spectrometry.
暂无分享,去创建一个
Arsenic in homeopathic drugs was determined by coupling a volatile generation with inductively coupled plasma-atomic emission spectrometry. The method is based on the chemical vaporization of arsenic(III) with bromide ions in sulfuric acid media using a batch procedure and subsequent introduction of the gaseous analyte into the plasma torch. The main and interactive effects of the experimental variables affecting this method were evaluated by a 2-level full factorial design. In optimized conditions by Simplex, the method shows an absolute detection limit (3 s) of 0.28 microg for the injection of 230 microL sample. The precision (% relative standard deviation) of the determination was 4.2% at a level of 50 microg/mL As(III) (n = 5). The interference effect of various ions on the arsenic signal was evaluated.
[1] J. Dědina,et al. Hydride Generation Atomic Absorption Spectrometry , 1995 .
[2] Akbar Montaser,et al. Inductively coupled plasmas in analytical atomic spectrometry , 1987 .
[3] A. Trotman‐Dickenson,et al. ‘Comprehensive’ Inorganic Chemistry , 1958, Nature.