Initial Accuracy of HIV Rapid Test Kits Stored in Suboptimal Conditions and Validity of Delayed Reading of Oral Fluid Tests

Objectives To evaluate the effect of storing commonly used rapid diagnostic tests above manufacturer-recommended temperature (at 37°C), and the accuracy of delayed reading of oral fluid kits with relevance to HIV self-testing programmes. Design A quality assurance study of OraQuick (OraSure), Determine HIV 1/2™ (Alere) and Uni-Gold™ (Recombigen®). Methods Consecutive adults (≥18y) attending Ndirande Health Centre in urban Blantyre, Malawi in January to April 2012 underwent HIV testing with two of each of the three rapid diagnostic test kits stored for 28 days at either 18°C (optimally-stored) or at 37°C (pre-incubated). Used OraQuick test kits were stored in a laboratory for delayed day 1 and subsequent monthly re-reading was undertaken for one year. Results Of 378 individuals who underwent parallel testing, 5 (1.3%) were dropped from the final analysis due to discordant or missing reference standard results (optimally-stored Determine and Uni-Gold). Compared to the diagnostic reference standard, OraQuick had a sensitivity of 97.2% (95% CI: 93.6–99.6). There were 7 false negative results among all test kits stored at 37°C and three false negatives among optimally stored kits. Excellent agreement between pre-incubated tests and optimally-stored tests with Kappa values of 1.00 for Determine and Uni-Gold; and 0.97 (95% CI: 0.95; 1.00) for OraQuick were observed. There was high visual stability on re-reading of OraQuick, with only 1/375 pre-incubated and 1/371 optimally-stored OraQuick kits changing from the initial result over 12 months. Conclusion Erroneous results observed during HIV testing in low income settings are likely to be due to factors other than suboptimal storage conditions. Re-reading returned OraQuick kits may offer a convenient and accurate quality assurance approach, including in HIV self-testing programmes.

[1]  R. Hayes,et al.  Uptake, Accuracy, Safety, and Linkage into Care over Two Years of Promoting Annual Self-Testing for HIV in Blantyre, Malawi: A Community-Based Prospective Study , 2015, PLoS medicine.

[2]  N. Ford,et al.  A report on the misdiagnosis of HIV status , 2015 .

[3]  N. Pearce,et al.  Evaluation of HIV testing algorithms in Ethiopia: the role of the tie-breaker algorithm and weakly reacting test lines in contributing to a high rate of false positive HIV diagnoses , 2015, BMC Infectious Diseases.

[4]  L. Hocqueloux,et al.  Finger-Stick Whole Blood HIV-1/-2 Home-Use Tests Are More Sensitive than Oral Fluid-Based In-Home HIV Tests , 2014, PloS one.

[5]  S. Staveteig,et al.  Demographic patterns of HIV testing uptake in sub-Saharan Africa , 2013 .

[6]  D. Klarkowski,et al.  False Positive HIV Diagnoses in Resource Limited Settings: Operational Lessons Learned for HIV Programmes , 2013, PloS one.

[7]  L. van Damme,et al.  Performance of a Rapid and Simple HIV Testing Algorithm in a Multicenter Phase III Microbicide Clinical Trial , 2011, Clinical and Vaccine Immunology.

[8]  Alice Mao,et al.  Acute Dengue fever causes false-positive reactivity in OraQuick rapid HIV-1/2 antibody test. , 2010, Journal of acquired immune deficiency syndromes.

[9]  C. Delaugerre,et al.  Sensitivity of Five Rapid HIV Tests on Oral Fluid or Finger-Stick Whole Blood: A Real-Time Comparison in a Healthcare Setting , 2010, PloS one.

[10]  James Mudzori,et al.  Field evaluation of diagnostic accuracy of an oral fluid rapid test for HIV, tested at point-of-service sites in rural Zimbabwe. , 2009, AIDS patient care and STDs.

[11]  J. Nelson,et al.  Comparison of Two Rapid Human Immunodeficiency Virus (HIV) Assays, Determine HIV-1/2 and OraQuick Advance Rapid HIV-1/2, for Detection of Recent HIV Seroconversion , 2008, Journal of Clinical Microbiology.

[12]  D. McPhee,et al.  Assessing Proficiency of Interpretation of Rapid Human Immunodeficiency Virus Assays in Nonlaboratory Settings: Ensuring Quality of Testing , 2008, Journal of Clinical Microbiology.

[13]  P. Sullivan,et al.  Post-marketing surveillance of OraQuick whole blood and oral fluid rapid HIV testing , 2006, AIDS.