Comparison of two respiration monitoring systems for 4D imaging with a Siemens CT using a new dynamic breathing phantom

Four-dimensional computed tomography (4D-CT) requires breathing information from the patient, and for this, several systems are available. Testing of these systems, under realistic conditions, requires a phantom with a moving target and an expandable outer contour. An anthropomorphic phantom was developed to simulate patient breathing as well as lung tumor motion. Using the phantom, an optical camera system (GateCT) and a pressure sensor (AZ-733V) were simultaneously operated, and 4D-CTs were reconstructed with a Siemens CT using the provided local-amplitude-based sorting algorithm. The comparison of the tumor trajectories of both systems revealed discrepancies up to 9.7 mm. Breathing signal differences, such as baseline drift, temporal resolution and noise level were shown not to be the reason for this. Instead, the variability of the sampling interval and the accuracy of the sampling rate value written on the header of the GateCT-signal file were identified as the cause. Interpolation to regular sampling intervals and correction of the sampling rate to the actual value removed the observed discrepancies. Consistently, the introduction of sampling interval variability and inaccurate sampling rate values into the header of the AZ-733V file distorted the tumor trajectory for this system. These results underline the importance of testing new equipment thoroughly, especially if components of different manufacturers are combined.

[1]  G. Christensen,et al.  A method for the reconstruction of four-dimensional synchronized CT scans acquired during free breathing. , 2003, Medical physics.

[2]  E. Larsen,et al.  A method for incorporating organ motion due to breathing into 3D dose calculations. , 1999, Medical physics.

[3]  Oliver Jäkel,et al.  Analysis of uncertainties in Gafchromic® EBT film dosimetry of photon beams , 2008, Physics in medicine and biology.

[4]  Tinsu Pan,et al.  Four-dimensional computed tomography: image formation and clinical protocol. , 2005, Medical physics.

[5]  Uwe Oelfke,et al.  Real-time tumor tracking: automatic compensation of target motion using the Siemens 160 MLC. , 2010, Medical physics.

[6]  M van Herk,et al.  Fusion of respiration-correlated PET and CT scans: correlated lung tumour motion in anatomical and functional scans , 2005, Physics in medicine and biology.

[7]  P J Keall,et al.  The application of the sinusoidal model to lung cancer patient respiratory motion. , 2005, Medical physics.

[8]  H. Kubo,et al.  Respiration gated radiotherapy treatment: a technical study. , 1996, Physics in medicine and biology.

[9]  Hans-Ulrich Kauczor,et al.  Measurement of tumor diameter-dependent mobility of lung tumors by dynamic MRI. , 2004, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[10]  T. Guerrero,et al.  Acquiring 4D thoracic CT scans using a multislice helical method. , 2004, Physics in medicine and biology.

[11]  Timothy D. Solberg,et al.  Phase versus amplitude sorting of 4D‐CT data , 2006, Journal of applied clinical medical physics.

[12]  Steve B. Jiang,et al.  Comparison of target registration errors for multiple image-guided techniques in accelerated partial breast irradiation. , 2008, International journal of radiation oncology, biology, physics.

[13]  D. Yan,et al.  Coupling surface cameras with on-board fluoroscopy: a feasibility study. , 2011, Medical physics.

[14]  Icru Prescribing, recording, and reporting photon beam therapy , 1993 .

[15]  George Starkschall,et al.  Verification of four-dimensional photon dose calculations. , 2009, Medical physics.

[16]  Jan Seuntjens,et al.  A deformable phantom for 4D radiotherapy verification: design and image registration evaluation. , 2008, Medical physics.

[17]  Ross Berbeco,et al.  Use of a realistic breathing lung phantom to evaluate dose delivery errorsa). , 2010, Medical physics.

[18]  Mitsuhiro Nakamura,et al.  Impact of motion velocity on four-dimensional target volumes: a phantom study. , 2009, Medical physics.

[19]  Parag J. Parikh,et al.  Development of the 4D Phantom for patient-specific end-to-end radiation therapy QA , 2007, SPIE Medical Imaging.

[20]  Steve Webb,et al.  Quantifying the effect of respiratory motion on lung tumour dosimetry with the aid of a breathing phantom with deforming lungs , 2006, Physics in medicine and biology.

[21]  R. Mohan,et al.  Acquiring a four-dimensional computed tomography dataset using an external respiratory signal. , 2003, Physics in medicine and biology.

[22]  George T. Y. Chen,et al.  A phantom evaluation of a stereo-vision surface imaging system for radiotherapy patient setup. , 2005, Medical physics.

[23]  D. Low,et al.  A comparison between amplitude sorting and phase-angle sorting using external respiratory measurement for 4D CT. , 2006, Medical physics.

[24]  Steve B. Jiang,et al.  Using surface imaging and visual coaching to improve the reproducibility and stability of deep-inspiration breath hold for left-breast-cancer radiotherapy , 2009, Physics in medicine and biology.

[25]  Steve B. Jiang,et al.  Patient-specific motion artifacts in 4DCT. , 2010, Medical physics.

[26]  Steve B. Jiang,et al.  A theoretical model for respiratory motion artifacts in free-breathing CT scans , 2008, Physics in medicine and biology.

[27]  C. Ling,et al.  Respiration-correlated spiral CT: a method of measuring respiratory-induced anatomic motion for radiation treatment planning. , 2002, Medical physics.

[28]  W. Schlegel,et al.  Accuracy of a commercial optical 3D surface imaging system for realignment of patients for radiotherapy of the thorax , 2007, Physics in medicine and biology.

[29]  Maria Francesca Spadea,et al.  Evaluation and commissioning of a surface based system for respiratory sensing in 4D CT , 2010, Journal of applied clinical medical physics.

[30]  P Rassiah-Szegedi,et al.  A proto-type design of a real-tissue phantom for the validation of deformation algorithms and 4D dose calculations. , 2010, Physics in medicine and biology.

[31]  Christoph Bert,et al.  Target motion tracking with a scanned particle beam. , 2007, Medical physics.

[32]  R Mohan,et al.  Determining parameters for respiration-gated radiotherapy. , 2001, Medical physics.