Geometric correction factor for transepithelial electrical resistance measurements in transwell and microfluidic cell cultures

Transepithelial electrical resistance (TEER) measurements are regularly used in in vitro models to quantitatively evaluate the cell barrier function. Although it would be expected that TEER values obtained with the same cell type and experimental setup were comparable, values reported in the literature show a large dispersion for unclear reasons. This work highlights a possible error in a widely used formula to calculate the TEER, in which it may be erroneously assumed that the entire cell culture area contributes equally to the measurement. In this study, we have numerically calculated this error in some cell cultures previously reported. In particular, we evidence that some TEER measurements resulted in errors when measuring low TEERs, especially when using Transwell inserts 12 mm in diameter or microfluidic systems that have small chamber heights. To correct this error, we propose the use of a geometric correction factor (GCF) for calculating the TEER. In addition, we describe a simple method to determine the GCF of a particular measurement system, so that it can be applied retrospectively. We have also experimentally validated an interdigitated electrodes (IDE) configuration where the entire cell culture area contributes equally to the measurement, and it also implements minimal electrode coverage so that the cells can be visualized alongside TEER analysis.

[1]  Hywel Morgan,et al.  On-chip epithelial barrier function assays using electrical impedance spectroscopy. , 2010, Lab on a chip.

[2]  R. Ransohoff,et al.  Human Brain Microvascular Endothelial Cells and Umbilical Vein Endothelial Cells Differentially Facilitate Leukocyte Recruitment and Utilize Chemokines for T Cell Migration , 2008, Clinical & developmental immunology.

[3]  A. Calabro,et al.  Evaluation of in vitro cytotoxicity and paracellular permeability of intact monolayers with mouse embryonic stem cells. , 2008, Toxicology in vitro : an international journal published in association with BIBRA.

[4]  C. Pfarrer,et al.  Bovine Caruncular Epithelial Cell Line (BCEC-1) Isolated from the Placenta Forms a Functional Epithelial Barrier in a Polarised Cell Culture Model , 2007, Placenta.

[5]  S. Krähenbühl,et al.  The human brain endothelial cell line hCMEC/D3 as a human blood‐brain barrier model for drug transport studies , 2008, Journal of neurochemistry.

[6]  K. Suh,et al.  A multi-layer microfluidic device for efficient culture and analysis of renal tubular cells. , 2010, Lab on a chip.

[7]  A. Berg,et al.  BBB ON CHIP: microfluidic platform to mechanically and biochemically modulate blood-brain barrier function , 2013, Biomedical microdevices.

[8]  C. Noe,et al.  Validation of in vitro cell culture models of the blood-brain barrier: tightness characterization of two promising cell lines. , 2008, Journal of pharmaceutical sciences.

[9]  M. Gye Changes in the expression of claudins and transepithelial electrical resistance of mouse Sertoli cells by Leydig cell coculture. , 2003, International journal of andrology.

[10]  Koichi Mizutani,et al.  Electrical impedance analysis of potato tissues during drying , 2014 .

[11]  R. Bragós,et al.  Real‐time and on‐line monitoring of morphological cell parameters using electrical impedance spectroscopy measurements , 2016 .

[12]  Andreas Janshoff,et al.  Automated multi-well device to measure transepithelial electrical resistances under physiological conditions. , 2004, BioTechniques.

[13]  Chia-Ching Wu,et al.  Enhancement of renal epithelial cell functions through microfluidic-based coculture with adipose-derived stem cells. , 2013, Tissue engineering. Part A.

[14]  H. Schwan,et al.  THE CONDUCTIVITY OF LIVING TISSUES , 1957, Annals of the New York Academy of Sciences.

[15]  Sverre Grimnes,et al.  Sources of error in tetrapolar impedance measurements on biomaterials and other ionic conductors , 2007 .

[16]  D. Ingber,et al.  Reconstituting Organ-Level Lung Functions on a Chip , 2010, Science.

[17]  D. Janigro,et al.  A new dynamic in vitro modular capillaries-venules modular system: Cerebrovascular physiology in a box , 2013, BMC Neuroscience.

[18]  E. Wood,et al.  Effect of Chitosan on Epithelial Cell Tight Junctions , 2004, Pharmaceutical Research.

[19]  A. Moses,et al.  Comparative study of four immortalized human brain capillary endothelial cell lines, hCMEC/D3, hBMEC, TY10, and BB19, and optimization of culture conditions, for an in vitro blood–brain barrier model for drug permeability studies , 2013, Fluids and Barriers of the CNS.

[20]  Pál Ormos,et al.  A versatile lab-on-a-chip tool for modeling biological barriers , 2016 .

[21]  Rosa Villa,et al.  Portable 4 Wire Bioimpedance Meter with Bluetooth Link , 2009 .

[22]  Kathrin Benson,et al.  Impedance-based cell monitoring: barrier properties and beyond , 2013, Fluids and Barriers of the CNS.

[23]  T. Kunze,et al.  Temperature corrected transepithelial electrical resistance (TEER) measurement to quantify rapid changes in paracellular permeability. , 2010, Die Pharmazie.

[24]  J. Greenwood,et al.  Blood‐brain barrier‐specific properties of a human adult brain endothelial cell line , 2005, FASEB journal : official publication of the Federation of American Societies for Experimental Biology.

[25]  Ignacio A. Romero,et al.  Development of a three-dimensional, all-human in vitro model of the blood–brain barrier using mono-, co-, and tri-cultivation Transwell models , 2011, Journal of Neuroscience Methods.

[26]  X. Illa,et al.  A Novel Modular Bioreactor to In Vitro Study the Hepatic Sinusoid , 2014, PloS one.

[27]  Z. Ablonczy,et al.  VEGF modulation of retinal pigment epithelium resistance. , 2007, Experimental eye research.

[28]  Shuvo Roy,et al.  A microfluidic bioreactor with integrated transepithelial electrical resistance (TEER) measurement electrodes for evaluation of renal epithelial cells , 2010, Biotechnology and bioengineering.

[29]  Donald E Ingber,et al.  Measuring direct current trans-epithelial electrical resistance in organ-on-a-chip microsystems. , 2015, Lab on a chip.

[30]  Kapil Pant,et al.  SyM-BBB: a microfluidic Blood Brain Barrier model. , 2013, Lab on a chip.

[31]  Experimental tools to monitor the dynamics of endothelial barrier function: a survey of in vitro approaches , 2014, Cell and Tissue Research.

[32]  C. Ringbom,et al.  Establishment and functional characterization of an in vitro model of the blood-brain barrier, comprising a co-culture of brain capillary endothelial cells and astrocytes. , 2001, European journal of pharmaceutical sciences : official journal of the European Federation for Pharmaceutical Sciences.

[33]  Masami Niwa,et al.  Permeability Studies on In Vitro Blood–Brain Barrier Models: Physiology, Pathology, and Pharmacology , 2005, Cellular and Molecular Neurobiology.

[34]  P. Bagos,et al.  Effects of Yeast and Bacterial Commensals and Pathogens of the Female Genital Tract on the Transepithelial Electrical Resistance of HeLa Cells , 2016, The open microbiology journal.

[35]  A. McBain,et al.  Strain-Dependent Augmentation of Tight-Junction Barrier Function in Human Primary Epidermal Keratinocytes by Lactobacillus and Bifidobacterium Lysates , 2013, Applied and Environmental Microbiology.

[36]  Edward J. Rapp,et al.  Immortalized Human Brain Endothelial Cells and Flow-Based Vascular Modeling: A Marriage of Convenience for Rational Neurovascular Studies , 2008, Journal of cerebral blood flow and metabolism : official journal of the International Society of Cerebral Blood Flow and Metabolism.

[37]  Dana M Spence,et al.  Microfluidic transendothelial electrical resistance measurement device that enables blood flow and postgrowth experiments. , 2011, Analytical chemistry.

[38]  Á. Kittel,et al.  A new blood–brain barrier model using primary rat brain endothelial cells, pericytes and astrocytes , 2009, Neurochemistry International.

[39]  Martin A. M. Gijs,et al.  Microfluidic chip for monitoring Ca2+ transport through a confluent layer of intestinal cells , 2014 .

[40]  Shuichi Takayama,et al.  Fabrication of two-layered channel system with embedded electrodes to measure resistance across epithelial and endothelial barriers. , 2010, Analytical chemistry.

[41]  Thomas J. Raub,et al.  In vitro models of the blood-brain barrier. , 1998, Alternatives to laboratory animals : ATLA.

[42]  Mandy B. Esch,et al.  TEER Measurement Techniques for In Vitro Barrier Model Systems , 2015, Journal of laboratory automation.

[43]  Rosa Villa,et al.  A novel strategy to monitor microfluidic in-vitro blood-brain barrier models using impedance spectroscopy , 2015, Microtechnologies for the New Millennium.

[44]  Rosa Villa,et al.  Manufacturing and full characterization of silicon carbide-based multi-sensor micro-probes for biomedical applications , 2007, Microelectron. J..

[45]  Tushar Kanti Bera,et al.  Bioelectrical Impedance Methods for Noninvasive Health Monitoring: A Review , 2014, Journal of medical engineering.

[46]  Mohammad F. Kiani,et al.  A Novel Dynamic Neonatal Blood-Brain Barrier on a Chip , 2015, PloS one.

[47]  Hanseup Kim,et al.  Characterization of a microfluidic in vitro model of the blood-brain barrier (μBBB). , 2012, Lab on a chip.

[48]  R. Marinšek-Logar,et al.  Interference ofSalmonella enteritidis andLactobacillus spp. with IL-8 levels and transepithelial electrical resistance of enterocyte-like caco-2 cells , 2008, Folia Microbiologica.