Numerical Simulations of Realistic Lead Trajectories and an Experimental Verification Support the Efficacy of Parallel Radiofrequency Transmission to Reduce Heating of Deep Brain Stimulation Implants during MRI

Patients with deep brain stimulation (DBS) implants may be subject to heating during MRI due to interaction with excitatory radiofrequency (RF) fields. Parallel RF transmit (pTx) has been proposed to minimize such RF-induced heating in preliminary proof-of-concept studies. The present work evaluates the efficacy of pTx technique on realistic lead trajectories obtained from nine DBS patients. Electromagnetic simulations were performed using 4- and 8-element pTx coils compared with a standard birdcage coil excitation using patient models and lead trajectories obtained by segmentation of computed tomography data. Numerical optimization was performed to minimize local specific absorption rate (SAR) surrounding the implant tip while maintaining spatial homogeneity of the transmitted RF magnetic field (B1+), by varying the input amplitude and phase for each coil element. Local SAR was significantly reduced at the lead tip with both 4-element and 8-element pTx (median decrease of 94% and 97%, respectively), whereas the median coefficient of spatial variation of B1+ inhomogeneity was moderately increased (30% for 4-element pTx and 20% for 8-element pTx) compared to that of the birdcage coil (17%). Furthermore, the efficacy of optimized 4-element pTx was verified experimentally by imaging a head phantom that included a wire implanted to approximate the worst-case lead trajectory for localized heating, based on the simulations. Negligible temperature elevation was observed at the lead tip, with reasonable image uniformity in the surrounding region. From this experiment and the simulations based on nine DBS patient models, optimized pTx provides a robust approach to minimizing local SAR with respect to lead trajectory.

[1]  H. H. Pennes Analysis of tissue and arterial blood temperatures in the resting human forearm. , 1948, Journal of applied physiology.

[2]  James R. Munkres,et al.  Elements of algebraic topology , 1984 .

[3]  Society of magnetic resonance in medicine , 1990 .

[4]  H. H. Pennes Analysis of tissue and arterial blood temperatures in the resting human forearm. 1948. , 1948, Journal of applied physiology.

[5]  G. P. Graber,et al.  Heating near implanted medical devices by the MRI RF-magnetic field , 1999 .

[6]  R. Doherty Long term follow up , 1999, BMJ.

[7]  T. W. Athey,et al.  Heating Near Implanted Medical Devices by the MRI RF-magnetic Field , 1999, IEEE International Magnetics Conference.

[8]  T. Ibrahim,et al.  Effect of RF coil excitation on field inhomogeneity at ultra high fields: a field optimized TEM resonator. , 2001, Magnetic resonance imaging.

[9]  Laurent Vercueil,et al.  Deep brain stimulation in the treatment of severe dystonia , 2001, Journal of Neurology.

[10]  Ashwini Sharan,et al.  Neurostimulation systems for deep brain stimulation: In vitro evaluation of magnetic resonance imaging–related heating at 1.5 tesla , 2002, Journal of magnetic resonance imaging : JMRI.

[11]  Jorge Gonzalez-Martinez,et al.  MR imaging-related heating of deep brain stimulation electrodes: in vitro study. , 2002, AJNR. American journal of neuroradiology.

[12]  P. Börnert,et al.  Transmit SENSE , 2003, Magnetic resonance in medicine.

[13]  F. Liu,et al.  Electromagnetic fields inside a lossy, multilayered spherical head phantom excited by MRI coils: models and methods. , 2004, Physics in medicine and biology.

[14]  Jean A. Tkach,et al.  Neurostimulation system used for deep brain stimulation (DBS): MR safety issues and implications of failing to follow safety recommendations. , 2004, Investigative radiology.

[15]  Yudong Zhu,et al.  Parallel excitation with an array of transmit coils , 2004, Magnetic resonance in medicine.

[16]  Frank G Shellock,et al.  Simple design changes to wires to substantially reduce MRI-induced heating at 1.5 T: implications for implanted leads. , 2005, Magnetic resonance imaging.

[17]  W. Grill Safety considerations for deep brain stimulation: review and analysis , 2005, Expert review of medical devices.

[18]  Jean A. Tkach,et al.  Reduction of Magnetic Resonance Imaging-related Heating in Deep Brain Stimulation Leads Using a Lead Management Device , 2005, Neurosurgery.

[19]  Robert E Gross,et al.  Deep brain stimulation for Parkinson's disease: Surgical technique and perioperative management , 2006, Movement disorders : official journal of the Movement Disorder Society.

[20]  M. Bikson,et al.  Bio-heat transfer model of deep brain stimulation induced temperature changes. , 2006, Conference proceedings : ... Annual International Conference of the IEEE Engineering in Medicine and Biology Society. IEEE Engineering in Medicine and Biology Society. Annual Conference.

[21]  Douglas C Noll,et al.  Spatial domain method for the design of RF pulses in multicoil parallel excitation , 2006, Magnetic resonance in medicine.

[22]  Sung-Min Park,et al.  Calculation of MRI‐induced heating of an implanted medical lead wire with an electric field transfer function , 2007, Journal of magnetic resonance imaging : JMRI.

[23]  M. Hariz,et al.  Thalamic deep brain stimulation in the treatment of essential tremor: a long-term follow-up , 2007, British journal of neurosurgery.

[24]  Alastair J. Martin,et al.  Magnetic Resonance Imaging of Implanted Deep Brain Stimulators: Experience in a Large Series , 2007, Stereotactic and Functional Neurosurgery.

[25]  W. Kainz,et al.  Complexity of MRI induced heating on metallic leads: Experimental measurements of 374 configurations , 2008, Biomedical engineering online.

[26]  M. Ladd,et al.  Measuring RF‐induced currents inside implants: Impact of device configuration on MRI safety of cardiac pacemaker leads , 2009, Magnetic resonance in medicine.

[27]  Giorgio Bonmassar,et al.  Analysis of the Role of Lead Resistivity in Specific Absorption Rate for Deep Brain Stimulator Leads at 3T MRI , 2010, IEEE Transactions on Medical Imaging.

[28]  Aviva Abosch,et al.  Effect of the extracranial deep brain stimulation lead on radiofrequency heating at 9.4 Tesla (400.2 MHz) , 2010, Journal of magnetic resonance imaging : JMRI.

[29]  Eliana Della Flora,et al.  Deep brain stimulation for essential tremor: A systematic review , 2010, Movement disorders : official journal of the Movement Disorder Society.

[30]  Yigitcan Eryaman,et al.  Reduction of implant RF heating through modification of transmit coil electric field , 2011, Magnetic resonance in medicine.

[31]  P. Starr,et al.  Treatment of dystonia with deep brain stimulation , 2008, Neurotherapeutics.

[32]  John S Witte,et al.  Of Models and Methods , 2011, Cancer Epidemiology, Biomarkers and Prevention.

[33]  V Schulz,et al.  Optimized RF shielding techniques for simultaneous PET/MR. , 2011, Medical physics.

[34]  M. Bikson,et al.  Temperature control at DBS electrodes using a heat sink: experimentally validated FEM model of DBS lead architecture , 2009, Journal of neural engineering.

[35]  E. Mohammadi,et al.  Barriers and facilitators related to the implementation of a physiological track and trigger system: A systematic review of the qualitative evidence , 2017, International journal for quality in health care : journal of the International Society for Quality in Health Care.

[36]  X.L. Chen,et al.  Deep Brain Stimulation , 2013, Interventional Neurology.

[37]  Mark Dewhirst,et al.  A heterogeneous human tissue mimicking phantom for RF heating and MRI thermal monitoring verification , 2012, Physics in medicine and biology.

[38]  Niels Kuster,et al.  Evaluation of the RF heating of a generic deep brain stimulator exposed in 1.5 T magnetic resonance scanners , 2013, Bioelectromagnetics.

[39]  Luca T. Mainardi,et al.  MRI-Based Multiscale Model for Electromagnetic Analysis in the Human Head with Implanted DBS , 2013, Comput. Math. Methods Medicine.

[40]  Esra Abaci Turk,et al.  Reduction of the radiofrequency heating of metallic devices using a dual‐drive birdcage coil , 2013, Magnetic resonance in medicine.

[41]  Greig Scott,et al.  Controlling radiofrequency‐induced currents in guidewires using parallel transmit , 2015, Magnetic resonance in medicine.

[42]  V. Hartwig Engineering for safety assurance in MRI: analytical, numerical and experimental dosimetry. , 2015, Magnetic resonance imaging.

[43]  Giorgio Bonmassar,et al.  A Novel Brain Stimulation Technology Provides Compatibility with MRI , 2015, Scientific Reports.

[44]  Kevan J T Anderson,et al.  Safely assessing radiofrequency heating potential of conductive devices using image‐based current measurements , 2015, Magnetic resonance in medicine.

[45]  Simon J. Graham,et al.  Investigation of Parallel Radiofrequency Transmission for the Reduction of Heating in Long Conductive Leads in 3 Tesla Magnetic Resonance Imaging , 2015, PloS one.

[46]  Lawrence L. Wald,et al.  Parallel transmit pulse design for patients with deep brain stimulation implants , 2015, Magnetic resonance in medicine.

[47]  Chun T. Rim Phasor Power Electronics , 2016 .

[48]  Trevor Coward,et al.  An In-Vitro Study , 2016 .

[49]  Kevan J T Anderson,et al.  Parallel radiofrequency transmission at 3 tesla to improve safety in bilateral implanted wires in a heterogeneous model , 2017, Magnetic resonance in medicine.

[50]  Cenap Çakmak Review and Analysis , 2017 .

[51]  Maria Ida Iacono,et al.  Local SAR near deep brain stimulation (DBS) electrodes at 64 and 127 MHz: A simulation study of the effect of extracranial loops , 2017, Magnetic resonance in medicine.

[52]  John Thornton,et al.  Technical challenges and safety of magnetic resonance imaging with in situ neuromodulation from spine to brain. , 2017, European journal of paediatric neurology : EJPN : official journal of the European Paediatric Neurology Society.

[53]  Azma Mareyam,et al.  Feasibility of using linearly polarized rotating birdcage transmitters and close‐fitting receive arrays in MRI to reduce SAR in the vicinity of deep brain simulation implants , 2017, Magnetic resonance in medicine.

[54]  S. Graham,et al.  A Platform for 4-Channel Parallel Transmission MRI at 3 T: Demonstration of Reduced Radiofrequency Heating in a Test Object Containing an Implanted Wire , 2019, Journal of Medical and Biological Engineering.