C-arm perfusion imaging with a fast penalized maximum-likelihood approach

Perfusion imaging is an essential method for stroke diagnostics. One of the most important factors for a successful therapy is to get the diagnosis as fast as possible. Therefore our approach aims at perfusion imaging (PI) with a cone beam C-arm system providing perfusion information directly in the interventional suite. For PI the imaging system has to provide excellent soft tissue contrast resolution in order to allow the detection of small attenuation enhancement due to contrast agent in the capillary vessels. The limited dynamic range of flat panel detectors as well as the sparse sampling of the slow rotating C-arm in combination with standard reconstruction methods results in limited soft tissue contrast. We choose a penalized maximum-likelihood reconstruction method to get suitable results. To minimize the computational load, the 4D reconstruction task is reduced to several static 3D reconstructions. We also include an ordered subset technique with transitioning to a small number of subsets, which adds sharpness to the image with less iterations while also suppressing the noise. Instead of the standard multiplicative EM correction, we apply a Newton-based optimization to further accelerate the reconstruction algorithm. The latter optimization reduces the computation time by up to 70%. Further acceleration is provided by a multi-GPU implementation of the forward and backward projection, which fulfills the demands of cone beam geometry. In this preliminary study we evaluate this procedure on clinical data. Perfusion maps are computed and compared with reference images from magnetic resonance scans. We found a high correlation between both images.

[1]  James M. Ortega,et al.  Iterative solution of nonlinear equations in several variables , 2014, Computer science and applied mathematics.

[2]  Jeffrey A. Fessler,et al.  Improved ordered subsets algorithm for 3 D X-ray CT image reconstruction , 2012 .

[3]  P Zygmanski,et al.  Evaluation of robustness of maximum likelihood cone-beam CT reconstruction with total variation regularization , 2012, Physics in medicine and biology.

[4]  Robert Frysch,et al.  Scalable OpenCL Accelerated Multi-GPU Projection for Cone Beam Computed Tomography with a Highly Accurate Separable Footprint Method , 2013 .

[5]  Jeffrey A. Fessler,et al.  3D Forward and Back-Projection for X-Ray CT Using Separable Footprints , 2010, IEEE Transactions on Medical Imaging.

[6]  R Fahrig,et al.  Cerebral CT Perfusion Using an Interventional C-Arm Imaging System: Cerebral Blood Flow Measurements , 2011, American Journal of Neuroradiology.

[7]  Thorsten M. Buzug Computed Tomography: From Photon Statistics to Modern Cone-Beam CT , 2010 .

[8]  Peter J. Huber,et al.  Robust Statistics , 2005, Wiley Series in Probability and Statistics.

[9]  B. Ripley,et al.  Robust Statistics , 2018, Encyclopedia of Mathematical Geosciences.

[10]  Jeffrey A. Fessler,et al.  Iterative Helical CT Reconstruction in the Cloud for Ten Dollars in Five Minutes , 2013 .

[11]  H. Malcolm Hudson,et al.  Accelerated image reconstruction using ordered subsets of projection data , 1994, IEEE Trans. Medical Imaging.

[12]  Robert Frysch,et al.  ML reconstruction of cone-beam projections acquired by a flat-panel rotational X-ray device , 2013, Medical Imaging.