Broadband ultrasound field mapping system using a wavelength tuned, optically scanned focused laser beam to address a Fabry Perot polymer film sensor

An optical system for rapidly mapping broadband ultrasound fields with high spatial resolution has been developed. The transduction mechanism is based upon the detection of acoustically induced changes in the optical thickness of a thin polymer film acting as a Fabry Perot sensing interferometer (FPI). By using a PC-controlled galvanometer mirror to line-scan a focused laser beam over the surface of the FPI, and a wavelength-tuned phase bias control system to optimally set the FPI working point, a notional 1D ultrasound array was synthesized. This system enabled ultrasound fields to be mapped over an aperture of 40 mm, in 50-μm steps with an optically defined element size of 50 μm and an acquisition time of 50 ms per step. The sensor comprised a 38-μm polymer film FPI which was directly vacuum-deposited onto an impedance-matched polycarbonate backing stub. The -3 dB acoustic bandwidth of the sensor was 300 kHz to 28 MHz and the peak noise-equivalent-pressure was 10 kPa over a 20-MHz measurement bandwidth. To demonstrate the system, the outputs of various planar and focused pulsed ultrasound transducers with operating frequencies in the range 3.5 to 20 MHz were mapped. It is considered that this approach offers a practical and inexpensive alternative to piezoelectric-based arrays and scanning systems for rapid transducer field characterization and biomedical and industrial ultrasonic imaging applications.

[1]  Paul C. Beard,et al.  A 2D optical ultrasound array using a polymer film sensing interferometer , 2000, 2000 IEEE Ultrasonics Symposium. Proceedings. An International Symposium (Cat. No.00CH37121).

[2]  V Wilkens,et al.  Optical multilayer detection array for fast ultrasonic field mapping. , 1999, Optics letters.

[3]  P C Beard,et al.  Extrinsic optical-fiber ultrasound sensor using a thin polymer film as a low-finesse Fabry-Perot interferometer. , 1996, Applied optics.

[4]  M. O'Donnell,et al.  High frequency ultrasound imaging with optical arrays , 1998, IEEE Transactions on Ultrasonics, Ferroelectrics and Frequency Control.

[5]  R. Witte,et al.  High frequency ultrasound imaging using Fabry-Perot optical etalon , 2005, SPIE BiOS.

[6]  Kentaro Nakamura,et al.  Pressure Sensitivity of a Fiber-Optic Microprobe for High-Frequency Ultrasonic Field , 1999 .

[7]  Paul C. Beard,et al.  Ultrahigh-sensitivity wideband Fabry-Perot ultrasound sensors as an alternative to piezoelectric PVDF transducers for biomedical photoacoustic detection , 2004, SPIE BiOS.

[8]  Volker Wilkens,et al.  Characterization of an optical multilayer hydrophone with constant frequency response in the range from 1 to 75 MHz. , 2003, The Journal of the Acoustical Society of America.

[9]  Paul C. Beard,et al.  2D imaging of ultrasound fields using CCD array to map output of Fabry-Perot polymer film sensor , 2006 .

[10]  Stefan A. Carp,et al.  Optoacoustic imaging using interferometric measurement of surface displacement , 2004 .

[11]  Edward Z. Zhang,et al.  Fabry Perot polymer film fibre-optic hydrophones and arrays for ultrasound field characterisation , 2004 .

[12]  Martin Frenz,et al.  Optical method for two-dimensional ultrasonic detection , 1999 .

[13]  F. Perennes,et al.  Transduction mechanisms of the Fabry-Perot polymer film sensing concept for wideband ultrasound detection , 1999, IEEE Transactions on Ultrasonics, Ferroelectrics and Frequency Control.

[14]  P.C. Beard Two-dimensional ultrasound receive array using an angle-tuned Fabry-Perot polymer film sensor for transducer field characterization and transmission ultrasound imaging , 2005, IEEE Transactions on Ultrasonics, Ferroelectrics and Frequency Control.

[15]  P. Beard,et al.  Characterization of a polymer film optical fiber hydrophone for use in the range 1 to 20 MHz: A comparison with PVDF needle and membrane hydrophones , 2000, IEEE Transactions on Ultrasonics, Ferroelectrics and Frequency Control.

[16]  Paul C. Beard,et al.  Characterisation of a polmer film optical fibre hydropone for the measurement of ultrasound fields for use in the range 1-20MHz: a comparsion with PVDF needle and membrane hydrophones , 2000 .

[17]  M. O'Donnell,et al.  High frequency optoacoustic arrays using etalon detection , 2000, IEEE Transactions on Ultrasonics, Ferroelectrics and Frequency Control.

[18]  C. Koch,et al.  Measurement of spatial cross sections of ultrasound pressure fields by optical scanning means , 2005, IEEE Transactions on Ultrasonics, Ferroelectrics and Frequency Control.