Single point measurements of magnetic field gradient waveform.

Pulsed magnetic field gradients are fundamental to spatial encoding and diffusion weighting in magnetic resonance. The ideal pulsed magnetic field gradient should have negligible rise and fall times, however, there are physical limits to how fast the magnetic field gradient may change with time. Finite gradient switching times, and transient, secondary, induced magnetic field gradients (eddy currents) alter the ideal gradient waveform and may introduce a variety of undesirable image artifacts. We have developed a new method to measure the complete magnetic field gradient waveform. The measurement employs a heavily doped test sample with short MR relaxation times (T(1), T(2), and T(2)(*)<100 micros) and a series of closely spaced broadband radiofrequency excitations, combined with single point data acquisition. This technique, a measure of evolving signal phase, directly determines the magnetic field gradient waveform experienced by the test sample. The measurement is sensitive to low level transient magnetic fields produced by eddy currents and other short and long time constant non-ideal gradient waveform behaviors. Data analysis is particularly facile permitting a very ready experimental check of gradient performance.

[1]  U. Klose In vivo proton spectroscopy in presence of eddy currents , 1990, Magnetic resonance in medicine.

[2]  Gradient time-shape measurement by NMR , 1986 .

[3]  H P Hetherington,et al.  A Method to measure arbitrary k‐space trajectories for rapid MR imaging , 1997, Magnetic resonance in medicine.

[4]  Vincent J Schmithorst,et al.  Automatic gradient preemphasis adjustment: A 15‐minute journey to improved diffusion‐weighted echo‐planar imaging , 2002, Magnetic resonance in medicine.

[5]  J. Frahm,et al.  Signal restoration for NMR imaging using time-dependent gradients , 1984 .

[6]  A. Caprihan,et al.  A programmable pre‐emphasis system , 1998, Magnetic resonance in medicine.

[7]  D. G. Hughes,et al.  Intensity artifacts in MRI caused by gradient switching in an animal‐size NMR magnet , 1992, Magnetic resonance in medicine.

[8]  B. Balcom,et al.  Single-Point Ramped Imaging with T1 Enhancement (SPRITE) , 1996, Journal of magnetic resonance. Series A.

[9]  I J Lowe,et al.  A novel eddy current compensation scheme for pulsed gradient systems , 1994, Magnetic resonance in medicine.

[10]  M. M Tropper,et al.  Image reconstruction for the NMR echo-planar technique, and for a proposed adaptation to allow continuous data acquisition , 1981 .

[11]  A. Takahashi,et al.  Compensation of multi‐dimensional selective excitation pulses using measured k‐space trajectories , 1995, Magnetic resonance in medicine.

[12]  N J Pelc,et al.  Gradient characterization using a Fourier‐transform technique , 1998, Magnetic resonance in medicine.

[13]  C. Boesch,et al.  Temporal and spatial analysis of fields generated by eddy currents in superconducting magnets: Optimization of corrections and quantitative characterization of magnet/gradient systems , 1991, Magnetic resonance in medicine.

[14]  Bruce J. Balcom,et al.  Single-Point Imaging of Gradient Rise, Stabilization, and Decay , 1996 .

[15]  R. E. Wysong,et al.  A simple method of measuring gradient induced eddy currents to set compensation networks , 1993, Magnetic resonance in medicine.