Dosimetric evaluation of compensation in radiotherapy of the breast: MLC intensity modulation and physical compensators.

BACKGROUND AND PURPOSE Electronic portal images may be used to design the compensation required to maximise dose uniformity in the breast from opposed tangential beams. MATERIALS AND METHODS Four methods of implementing the desired compensation have been studied: a simple wedge, a physical compensator in conjunction with a wedge; one open field plus four shaped multi-leaf-collimated (MLC) fields, and one wedged field in conjunction with three shaped MLC fields. Evaluation was performed using thermoluminescent dosimeters (TLDs) placed inside a phantom which was designed to mimic the human breast. The measured results are compared with both the prediction of the in-house compensation design software and with the dose predicted by the GE Target II planning system. The implications of each method for the time taken to plan and deliver treatment were analysed. RESULTS The dose inhomogeneity, as measured at seven points in the central plane was greatest for the simple wedge (root mean square (rms) = 4.5%) compared to an open field plus four shaped MLC fields (rms = 2.2%), a wedged field plus three shaped MLC fields (rms = 3.3%), and the physical compensator (rms = 2.4%). The times required to plan and prepare these treatments varied considerably. The standard wedged treatment required under 15 min; both MLC-based and the physical compensator treatments required approximately 50 min. Differences of treatment delivery times were up to 8 min. CONCLUSIONS These results indicate that the dose inhomogeneity can be reduced by beam intensity modulation designed using EPIDs.

[1]  J. Yarnold,et al.  Correlation of breast dose heterogeneity with breast size using 3D CT planning and dose-volume histograms , 1995 .

[2]  P M Evans,et al.  Scattered radiation in portal images: a Monte Carlo simulation and a simple physical model. , 1996, Medical physics.

[3]  P. Munro,et al.  A review of electronic portal imaging devices (EPIDs). , 1992, Medical physics.

[4]  J. R. Williams,et al.  Radiotherapy physics : in practice , 2000 .

[5]  M C Schell,et al.  Input/output characteristics of a matrix ion-chamber electronic portal imaging device. , 1994, Medical physics.

[6]  P M Evans,et al.  Image comparison techniques for use with megavoltage imaging systems. , 1992, The British journal of radiology.

[7]  M C Schell,et al.  A technique of automating compensator design for lung inhomogeneity correction using an electron portal imaging device. , 1994, Medical physics.

[8]  A L Boyer,et al.  Compensating filters for high energy x rays. , 1982, Medical physics.

[9]  J. Yarnold,et al.  The influence of breast size on late radiation effects and association with radiotherapy dose inhomogeneity , 1995 .

[10]  D P Dearnaley,et al.  A randomised trial of patient repositioning during radiotherapy using a megavoltage imaging system. , 1994, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[11]  W P Mayles,et al.  Design of compensators for breast radiotherapy using electronic portal imaging. , 1995, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[12]  C. Yu,et al.  A method for implementing dynamic photon beam intensity modulation using independent jaws and a multileaf collimator. , 1995, Physics in medicine and biology.

[13]  D. Convery,et al.  The generation of intensity-modulated fields for conformal radiotherapy by dynamic collimation , 1992 .

[14]  P M Evans,et al.  The application of transit dosimetry to precision radiotherapy. , 1996, Medical physics.

[15]  B Fisher,et al.  Eight-year results of a randomized clinical trial comparing total mastectomy and lumpectomy with or without irradiation in the treatment of breast cancer. , 1989, The New England journal of medicine.

[16]  T. Bortfeld,et al.  X-ray field compensation with multileaf collimators. , 1994, International journal of radiation oncology, biology, physics.

[17]  W P Mayles,et al.  Improved dose homogeneity in the breast using tissue compensators. , 1991, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[18]  D. G. Lewis,et al.  A linear array, scintillation crystal-photodiode detector for megavoltage imaging. , 1991, Medical physics.

[19]  B Fisher,et al.  Five-year results of a randomized clinical trial comparing total mastectomy and segmental mastectomy with or without radiation in the treatment of breast cancer. , 1985, The New England journal of medicine.