An eSheath, left IJ approach to implantation of a Micra leadless pacemaker in a 2-year-old, 10.9 kg burn victim.
暂无分享,去创建一个
A 2-year-old female was transferred emergently from rural Mexico after sustaining third-degree circumferential thoracic/abdominal burns. She was noted to have irreversible third-degree heart block with junctional escape rate in the 50–60 bpm range with signs/symptoms of low cardiac output including altered mental status/intermittent hypotension. She was started on isoproterenol at 0.02 μg/kg/min with improvement. The patient had extensive skin grafting/harvesting. Post-operatively, multiple slow attempts to wean isoproterenol over weeks resulted in bradycardia/lethargy/poor perfusion/mentation. Interdisciplinary discussions were held with paediatric cardiology/electrophysiology/cardiothoracic surgery/burn surgery/critical-care/anaesthesia teams with a transcatheter leadless pacemaker approach preferred over any incisional implantation, including of epicardial leads given degree/location of grafting/burns. Using the Seldinger technique, a 6 Fr sheath was placed in the 1.2 cm diameter left IJ (vs. smaller 0.8 cm right IJ). Serial dilation was carried out by 2-Fr size until a 16-Fr, 36 cm long eSheath/dilator (Edwards Lifesciences, Irvine, CA, USA) was advanced over a guidewire into the mid-right atrium. After heparinization, the 23-F Micra dilator (Medtronic, Minneapolis, MN, USA) was then used to expand the inner diameter of the eSheath in preparation for implantation of the Micra. The 55.7 cm Micra sheath (Medtronic) was cut/shortened to 20 cm to allow for sufficient length of the Micra catheter. Using the Micra dilator, the shortened Micra sheath was introduced through the 16-Fr eSheath valve to allow passage of the Micra into the eSheath. Subsequently, the Micra catheter was passed through the eSheath to the mid-right atrium, and the eSheath was retracted back to the innominate vein. The Micra was advanced across the tricuspid valve, under echocardiographic/fluoroscopic guidance, to an apical-septal position. Due to lack of septal engagement, a second deployment was needed for proper implantation. The catheter/sheath was removed, and a figure-of-8 stitch was placed for haemostasis. Pacing threshold was 0.38 V at 0.24 ms, R-wave of 3.5 mV, and impedance 610 ohms, programmed VDD 70 to 115 bpm, mode switch VVIR at 115 bpm with upper sensor rate of 170 bpm. Three tines were engaged. No complications occurred (Panels A and B). Four-month follow-up demonstrated threshold of 0.5 V at 0.24 ms, R-wave of 9.0 mV, and impedance of 605 ohms. A3 and A4 programming were used as the atrial accelerometer vector with 43% A-sense, V-paced rhythm.
[1] J. Steinberger,et al. Pediatric Micra leadless pacemaker implantation via internal jugular and femoral vein: a single center, US experience. , 2021, Future cardiology.
[2] Matthew Ambrose,et al. First description of a successful leadless pacemaker implantation via the left internal jugular vein (in a 20 kg patient). , 2020, Journal of electrocardiology.
[3] P. Oslizlok,et al. Leadless Micra Pacemaker Use in the Pediatric Population: Device Implantation and Short-Term Outcomes , 2019, Pediatric Cardiology.