Isokinetic sampler for continuous airborne aerosol measurements

For aerosol measurements, especially those concerned with the aerosol particle size distribution, it is important to sample in isokinetic conditions. Most available instrumentation for aerosol measurements is intended for use on the ground under light wind conditions; intake air speeds rarely exceed a few meters per second. If the same instrumentation is used onboard an aircraft, the air must be decelerated 60 or more m/sec before It is sampled by individual instruments. On The Pennsylvania State University Meteorology research aircraft, the air for all aerosol instruments is decelerated in a single isokinetic sampler located above the roof of the cabin outside the aircraft boundary layer. The air enters the sampler through a carefully designed circular intake. Its velocity is reduced as the cross section increases along a 7° conical diffuser. The expansion cone terminates in a cylindrical chamber in which the air velocity is 1/16 the aircraft speed. Behind the sampling chamber the air is accelerated in a...