Analysis of confidence level scores from an ROC study: comparison of three mammographic systems for detection of simulated calcifications

The purpose of this study is to compare the detection performance of three different mammography systems: screen/film (SF) combination, a-Si/CsI flat-panel (FP-), and charge-coupled device (CCD-) based systems. A 5-cm thick 50% adipose/50% glandular breast tissue equivalent slab phantom was used to provide an uniform background. Calcium carbonate grains of three different size groups were used to simulate microcalcifications (MCs): 112-125, 125-140, and 140-150 μm overlapping with the uniform background. Calcification images were acquired with the three mammography systems. Digital images were printed on hardcopy films. All film images were displayed on a mammographic viewer and reviewed by 5 mammographers. The visibility of the MC was rated with a 5-point confidence rating scale for each detection task, including the negative controls. Scores were averaged over all readers for various detectors and size groups. Receiver operating characteristic (ROC) analysis was performed and the areas under the ROC curves (Az’s) were computed for various imaging conditions. The results shows that (1) the FP-based system performed significantly better than the SF and CCD-based systems for individual size groups using ROC analysis (2) the FP-based system also performed significantly better than the SF and CCD-based systems for individual size groups using averaged confidence scale, and (3) the results obtained from the Az’s were largely correlated with these from confidence level scores. However, the correlation varied slightly among different imaging conditions.

[1]  Andreas Koch,et al.  Design and evaluation of a slot-scanning full-field digital mammography system , 2002, SPIE Medical Imaging.

[2]  S Suryanarayanan,et al.  Mammographic imaging with a small format CCD-based digital cassette: physical characteristics of a clinical system. , 2000, Medical physics.

[3]  M L Giger,et al.  Digital chest radiography: effect on diagnostic accuracy of hard copy, conventional video, and reversed gray scale video display formats. , 1988, Radiology.

[4]  S Suryanarayanan,et al.  Full breast digital mammography with an amorphous silicon-based flat panel detector: physical characteristics of a clinical prototype. , 2000, Medical physics.

[5]  C E Ravin,et al.  Effects of image processing on nodule detection rates in digitized chest radiographs: ROC study of observer performance. , 1988, Radiology.

[6]  Srinivasan Vedantham,et al.  Flat-panel digital mammography system: contrast-detail comparison between screen-film radiographs and hard-copy images. , 2002, Radiology.

[7]  Peter G. J. Barten,et al.  Contrast sensitivity of the human eye and its e ects on image quality , 1999 .

[8]  Chris C. Shaw,et al.  Comparison of a-Si:H/CsI flat-panel digital imaging systems with CR-and CCD-based systems: image quality measurements , 2001 .

[9]  A R Cowen,et al.  Visibility of microcalcifications in computed and screen-film mammography. , 1997, Physics in medicine and biology.

[10]  H E Rockette,et al.  Practical issues of experimental ROC analysis. Selection of controls. , 1990, Investigative radiology.

[11]  J. Rowlands,et al.  The physics of computed radiography. , 2002, Physics in medicine and biology.

[12]  Kerry T. Krugh,et al.  Microcalcification detectability for four mammographic detectors: flat-panel, CCD, CR, and screen/film). , 2002, Medical physics.