Fracture of the femoral stem of the Ring TCH hip prosthesis.

We report the clinical and operative details of seven cases of fracture of the femoral stem of the Ring TiMESH cementless hip prosthesis (two were cemented and five uncemented). Six fractures occurred in the proximal one-third of the stem and one at mid-stem. The failures are attributed mainly to two defects in stem design, the narrowness of the anteroposterior dimensions and the depth of the recess for the titanium mesh pads. Great difficulty was experienced in removing the osseo-integrated distal fragments of the broken stems.

[1]  B. M. Wroblewski,et al.  The mechanism of fracture of the femoral prosthesis in total hip replacement , 1979, International Orthopaedics.

[2]  K. Bryant,et al.  Ring Uncemented Plastic on Metal Hip Replacements-Results from an Independent Unit , 1989, Journal of the Royal Society of Medicine.

[3]  D. Nunn The Ring uncemented plastic-on-metal total hip replacement. Five-year results. , 1988, The Journal of bone and joint surgery. British volume.

[4]  R. Mollan,et al.  Failure of the femoral component in the Howse total hip arthroplasty. , 1984, Clinical orthopaedics and related research.

[5]  G. Lord,et al.  The madreporic cementless total hip arthroplasty. New experimental data and a seven-year clinical follow-up study. , 1983, Clinical orthopaedics and related research.

[6]  R. Shastri,et al.  Fracture failure of a forged vitallium prosthesis. A case report. , 1982, The Journal of bone and joint surgery. American volume.

[7]  B M Wroblewski,et al.  Fractured stem in total hip replacement. A clinical review of 120 cases. , 1982, Acta orthopaedica Scandinavica.

[8]  J O Galante,et al.  Causes of fractures of the femoral component in total hip replacement. , 1980, The Journal of bone and joint surgery. American volume.

[9]  J. Charnley Fracture of femoral prostheses in total hip replacement. A clinical study. , 1975, Clinical orthopaedics and related research.