Tonometric Condition of Cellular Polypropylene Film Sensors in Measuring Arterial Pressure Waveform

Tonometric continuous measurement of arterial pressure becomes feasible using a cellular polypropylene (Cellular PP) film sensor. A pulsatile arterial vascular phantom model was used to find the range of optimal tonometric conditions and the responsiveness to dynamic pressure changes. The optimal tonometric condition was assessed by the correlation coefficient between the hydraulic pressure and the Cellular PP output using two different types of tubes (the latex tube and the hydrogel tube) to simulate arteries. With a setting of the normal blood pressure range, the output of Cellular PP correlated strongly with the level of hydraulic pressure, 0.998 and 0.989 in the latex tube and the hydrogel tube, respectively. For maintaining the optimal tonometric condition, the depressed depths of the latex and the hydrogel tube were less than 1.2 and 0.6 mm, respectively. The phantom model also demonstrated that the Cellular PP sensor followed changes in a hydraulic pressure dynamically under the optimal tonometric conditions. The present results demonstrated the Cellular PP film sensor is applicable to the arterial tonometry in measuring the instantaneous blood pressure while the sensor is adjusted to maintain the minimal flatness of the underlying arterial wall.Clinical Relevance— To understand the physiological characteristics of blood pressure and arterial system, the instantaneous measurement of blood pressure is necessary. The present study suggests that Cellular PP films are applicable to peripheral arteries tonometrically to obtain simultaneously the respective blood pressure waveforms.

[1]  Tetsuo Shoji,et al.  Brachial-ankle pulse wave velocity as an index of central arterial stiffness. , 2010, Journal of atherosclerosis and thrombosis.

[2]  P. Alam,et al.  H , 1887, High Explosives, Propellants, Pyrotechnics.

[3]  G. Fitzgerald,et al.  'I. , 2019, Australian journal of primary health.

[4]  Danna Zhou,et al.  d. , 1840, Microbial pathogenesis.

[5]  Pascal Verdonck,et al.  Functional analysis of the common carotid artery: relative distension differences over the vessel wall measured in vivo , 2004, Journal of hypertension.

[6]  J Booth,et al.  A Short History of Blood Pressure Measurement , 1977, Proceedings of the Royal Society of Medicine.

[7]  R. Westhorpe,et al.  Blood Pressure Monitoring – Automated Non-invasive Blood Pressure Monitors , 2009, Anaesthesia and Intensive Care.

[8]  A. Tajik,et al.  Noninvasive measurement of central vascular pressures with arterial tonometry: clinical revival of the pulse pressure waveform? , 2010, Mayo Clinic proceedings.

[9]  M. O'Rourke,et al.  Pulse wave analysis. , 1996, Journal of hypertension. Supplement : official journal of the International Society of Hypertension.

[10]  P M NEWGARD,et al.  A TRANSDUCER FOR THE CONTINUOUS EXTERNAL MEASUREMENT OF ARTERIAL BLOOD PRESSURE. , 1961, IEEE transactions on bio-medical engineering.

[12]  E. Fukada,et al.  A study of time stability of piezoelectricity in porous polypropylene electrets , 2009, 2009 IEEE International Ultrasonics Symposium.

[13]  M. Pozzi,et al.  Cellular polypropylene electromechanical properties: exploring the nonlinear region , 2018, Journal of Theoretical and Applied Physics.

[14]  L. Rabelo,et al.  Radial Applanation Tonometry as an Adjuvant Tool in the Noninvasive Arterial Stiffness and Blood Pressure Assessment , 2014 .

[15]  P. Cochat,et al.  Et al , 2008, Archives de pediatrie : organe officiel de la Societe francaise de pediatrie.

[16]  Tsuyoshi Murata,et al.  {m , 1934, ACML.

[17]  Ian J. Brown,et al.  Cardiovascular disease, chronic kidney disease, and diabetes mortality burden of cardiometabolic risk factors from 1980 to 2010: a comparative risk assessment. , 2014, The lancet. Diabetes & endocrinology.

[18]  Daniel W. Jones,et al.  The Seventh Report of the Joint National Committee on Prevention, Detection, Evaluation, and Treatment of High Blood Pressure: the JNC 7 report. , 2003, JAMA.

[19]  Masamichi Nogawa,et al.  Estimation of aortic BP waveform from noninvasive radial tonometry; validation of FFT and ARX methods , 1997, Proceedings of the 19th Annual International Conference of the IEEE Engineering in Medicine and Biology Society. 'Magnificent Milestones and Emerging Opportunities in Medical Engineering' (Cat. No.97CH36136).

[20]  A. Schutte,et al.  Comparison of central pressure estimates obtained from SphygmoCor, Omron HEM-9000AI and carotid applanation tonometry , 2011, Journal of hypertension.

[21]  J. Hillenbrand,et al.  High-sensitivity piezoelectret-film accelerometers , 2010, IEEE Transactions on Dielectrics and Electrical Insulation.

[22]  Patrick Segers,et al.  ARTERY Society guidelines for validation of non-invasive haemodynamic measurement devices: Part 1, arterial pulse wave velocity , 2010 .

[23]  D. P. Engberts,et al.  Arterial Line Placement: Safety First. , 2016, Anesthesiology.

[24]  Hao-Min Cheng,et al.  Measurement accuracy of non-invasively obtained central blood pressure by applanation tonometry: a systematic review and meta-analysis. , 2013, International journal of cardiology.