Novel Calibration Method for Switched Capacitor Arrays Enables Time Measurements With Sub-Picosecond Resolution

Switched capacitor arrays (SCA) ASICs are becoming more and more popular for the readout of detector signals, since the sampling frequency of typically several gigasamples per second allows excellent pile-up rejection and time measurements. They suffer however from the fact that their sampling bins are not equidistant in time, given by limitations of the chip manufacturing. In the past, this has limited time measurements of optimal signals to standard deviations (σ) of about 4-25 ps in accuracy for the split pulse test, depending on the specific chip. This paper introduces a novel time calibration, which determines the true sampling speed of an SCA. Additionally, for two independently running SCA chips, the achieved time resolution improved to less than 3 ps (σ) independently from the delay for the split pulse test, when simply applying a linear interpolation. When using a more advanced analyzing technique for the split pulse test with a single SCA, this limit is pushed below 1 ps (σ) for delays up to 8 ns. Various test measurements with different boards based on the DRS4 ASIC indicate that the new calibration is stable over time but not over larger temperature variations.

[1]  J. Aguilar,et al.  Performance of the front-end electronics of the ANTARES neutrino telescope , 2010 .

[2]  P. W. Cattaneo,et al.  The MEG detector for μ+→e+γ decay search , 2013, 1303.2348.

[3]  G. S. Varner,et al.  The large analog bandwidth recorder and digitizer with ordered readout (LABRADOR) ASIC , 2007 .

[4]  K. Nishimura,et al.  A Correlation-based Timing Calibration and Diagnostic Technique for Fast Digitizing ASICs , 2012 .

[5]  M. Albrow,et al.  Waveform digitization for high resolution timing detectors with silicon photomultipliers , 2012 .

[6]  R. Paoletti,et al.  Analysis techniques and performance of the Domino Ring Sampler version 4 based readout for the MAGIC telescopes , 2013, 1305.1007.

[7]  B. Wooley,et al.  An analog memory integrated circuit for waveform sampling up to 900 MHz , 1994 .

[8]  E. Delagnes,et al.  High resolution photon timing with MCP-PMTs: A comparison of a commercial constant fraction discriminator (CFD) with the ASIC-based waveform digitizers TARGET and WaveCatcher , 2010, IEEE Nuclear Science Symposuim & Medical Imaging Conference.

[9]  Roberto Dinapoli,et al.  Application of the DRS chip for fast waveform digitizing , 2010 .

[10]  Eric Delagnes,et al.  SAM: A new GHz sampling ASIC for the H.E.S.S.-II front-end electronics , 2006 .

[11]  A direct time measurement technique for the two-dimensional precision coordinate detectors based on thin-walled drift tubes , 2013, 1301.6018.

[12]  F. Kirsten,et al.  A flexible 128 channel silicon strip detector instrumentation integrated circuit with sparse data readout , 1988 .

[13]  F. Beekman,et al.  LaBr3:Ce and SiPMs for time-of-flight PET: achieving 100 ps coincidence resolving time , 2010, Physics in medicine and biology.

[14]  E. Oberla,et al.  A 15 GSa/s, 1.5 GHz bandwidth waveform digitizing ASIC , 2013, 1309.4397.

[15]  Shubin Liu,et al.  Evaluation of a Fast Pulse Sampling Module With Switched-Capacitor Arrays , 2012, IEEE Transactions on Nuclear Science.

[16]  D. Townsend,et al.  Physical and clinical performance of the mCT time-of-flight PET/CT scanner , 2011, Physics in medicine and biology.