Automated 2D-3D registration of a radiograph and a cone beam CT using line-segment enhancement.

The objective of this study was to develop a fully automated two-dimensional (2D)-three-dimensional (3D) registration framework to quantify setup deviations in prostate radiation therapy from cone beam CT (CBCT) data and a single AP radiograph. A kilovoltage CBCT image and kilovoltage AP radiograph of an anthropomorphic phantom of the pelvis were acquired at 14 accurately known positions. The shifts in the phantom position were subsequently estimated by registering digitally reconstructed radiographs (DRRs) from the 3D CBCT scan to the AP radiographs through the correlation of enhanced linear image features mainly representing bony ridges. Linear features were enhanced by filtering the images with "sticks," short line segments which are varied in orientation to achieve the maximum projection value at every pixel in the image. The mean (and standard deviations) of the absolute errors in estimating translations along the three orthogonal axes in millimeters were 0.134 (0.096) AP(out-of-plane), 0.021 (0.023) ML and 0.020 (0.020) SI. The corresponding errors for rotations in degrees were 0.011 (0.009) AP, 0.029 (0.016) ML (out-of-plane), and 0.030 (0.028) SI (out-of-plane). Preliminary results with megavoltage patient data have also been reported. The results suggest that it may be possible to enhance anatomic features that are common to DRRs from a CBCT image and a single AP radiography of the pelvis for use in a completely automated and accurate 2D-3D registration framework for setup verification in prostate radiotherapy. This technique is theoretically applicable to other rigid bony structures such as the cranial vault or skull base and piecewise rigid structures such as the spine.

[1]  William T. Rhodes,et al.  Scale- and rotation-invariant pattern recognition by a rotating kernel min-max transformation , 1990, Optics & Photonics.

[2]  G T Chen,et al.  Correlation of projection radiographs in radiation therapy using open curve segments and points. , 1992, Medical physics.

[3]  J Bijhold,et al.  Three-dimensional verification of patient placement during radiotherapy using portal images. , 1993, Medical physics.

[4]  M van Herk,et al.  An algorithm for automatic analysis of portal images: clinical evaluation for prostate treatments. , 1993, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[5]  K. Eilertsen,et al.  Methods for fully automated verification of patient set-up in external beam radiotherapy with polygon shaped fields. , 1994, Physics in medicine and biology.

[6]  G J Kutcher,et al.  The effects of out-of-plane rotations on two dimensional portal image registration in conformal radiotherapy of the prostate. , 1995, International journal of radiation oncology, biology, physics.

[7]  A Boxwala,et al.  Core-based portal image registration for automatic radiotherapy treatment verification. , 1995, International journal of radiation oncology, biology, physics.

[8]  M van Herk,et al.  Automatic three-dimensional inspection of patient setup in radiation therapy using portal images, simulator images, and computed tomography data. , 1996, Medical physics.

[9]  J Z Wang,et al.  Investigation of a phase-only correlation technique for anatomical alignment of portal images in radiation therapy. , 1996, Physics in medicine and biology.

[10]  Jürgen Weese,et al.  Voxel-based 2-D/3-D registration of fluoroscopy images and CT scans for image-guided surgery , 1997, IEEE Transactions on Information Technology in Biomedicine.

[11]  Gaudenz Danuser,et al.  Deformable multi template matching with application to portal images , 1997, Proceedings of IEEE Computer Society Conference on Computer Vision and Pattern Recognition.

[12]  Douglas L. Jones,et al.  Line and boundary detection in speckle images , 1998, IEEE Trans. Image Process..

[13]  Olaf Dössel,et al.  A structure-based method for on-line matching of portal images for an optimal patient set-up in radiotherapy , 1998 .

[14]  A Zaccarin,et al.  Automatic setup deviation measurements with electronic portal images for pelvic fields. , 1998, Medical physics.

[15]  K W Leszczynski,et al.  An image registration scheme applied to verification of radiation therapy. , 1998, The British journal of radiology.

[16]  Zhe Chen,et al.  A Novel Approach for the Registration of 2D Portal and 3D CT Images for Treatment Setup Verification in Radiotherapy , 1998, MICCAI.

[17]  Guido Gerig,et al.  Motion Measurements in Low-Contrast X-ray Imagery , 1998, MICCAI.

[18]  Jürgen Weese,et al.  A comparison of similarity measures for use in 2-D-3-D medical image registration , 1998, IEEE Transactions on Medical Imaging.

[19]  J M Balter,et al.  Determination of rotations in three dimensions using two-dimensional portal image registration. , 1998, Medical physics.

[20]  J. Chu,et al.  A simple algorithm for planar image registration in radiation therapy. , 1998, Medical physics.

[21]  Douglas L. Jones,et al.  Detection of lines and boundaries in speckle images-application to medical ultrasound , 1999, IEEE Transactions on Medical Imaging.

[22]  Zhe Chen,et al.  Entropy-Based, Multiple-Portal-to-3DCT Registration for Prostate Radiotherapy Using Iteratively Estimated Segmentation , 1999, MICCAI.

[23]  J Knisely,et al.  A minimax entropy registration framework for patient setup verification in radiotherapy. , 1999, Computer aided surgery : official journal of the International Society for Computer Aided Surgery.

[24]  D Verellen,et al.  Automatic on-line electronic portal image analysis with a wavelet-based edge detector. , 2000, Medical physics.

[25]  M van Herk,et al.  3-D portal image analysis in clinical practice: an evaluation of 2-D and 3-D analysis techniques as applied to 30 prostate cancer patients. , 2000, International journal of radiation oncology, biology, physics.

[26]  J Kim,et al.  A feasibility study of mutual information based setup error estimation for radiotherapy. , 2001, Medical physics.

[27]  D L Hill,et al.  Validation of a two- to three-dimensional registration algorithm for aligning preoperative CT images and intraoperative fluoroscopy images. , 2001, Medical physics.

[28]  An automated image registration algorithm for quantifying three-dimensional setup variations in prostate radiotherapy , 2002 .

[29]  Zhe Chen,et al.  Entropy-based dual-portal-to-3-DCT registration incorporating pixel correlation , 2003, IEEE Transactions on Medical Imaging.

[30]  Serge Miguet,et al.  Patient setup error measurement using 3D intensity-based image registration techniques. , 2003, International journal of radiation oncology, biology, physics.

[31]  Mark Jobse,et al.  A new algorithm for the registration of portal images to planning images in the verification of radiotherapy, as validated in prostate treatments. , 2003, Medical physics.

[32]  Marcel van Herk,et al.  Evaluation of cost functions for gray value matching of two-dimensional images in radiotherapy. , 2003, Medical physics.

[33]  Automated three dimensional quantification of setup deviations in radiation therapy using a single radiograph and cone beam computed tomography data , 2004 .