Impact of deformable image registration accuracy on thoracic images with different regularization weight parameter settings.

PURPOSE We assessed the deformable image registration (DIR) accuracy of thoracic images under different regularization weights using commercially available DIR software. METHODS The thoracic 4-dimensional (4D) CT images of 10 patients were used. The datasets for these patients were provided by DIR-lab (www.dir-lab.com) and included a coordinate list of 300 anatomic landmarks that had been manually identified. The ANAtomically CONstrained Deformation Algorithm (ANACONDA) of RayStation (RaySearch Laboratories, Stockholm, Sweden) was used to deform the peak-inhale to peak-exhale images under different regularization weights (4, 40, 400-default setting, 1500, 4000, 10,000, 15,000, 20,000, 30,000, and 40,000). The regularization weights were changed using a script. The registration error (RE) was determined by calculating the difference at each landmark point between the displacement calculated by the DIR software and that calculated by the landmark. We measured the computation time for each regularization weight setting. RESULTS High regularization weights resulted in a smaller RE than that observed with lower regularization weights. The RE decreases rapidly with increase in regularization weight before reaching a plateau. No significant difference was found between a regularization weight of 400 and regularization weights of 4, 40, 4000 or 40,000 (P value >0.05). The range of the average time was 8.4-12.2s. CONCLUSIONS We concluded that the default setting for ANACONDA is stable with respect to regularization weight in the thoracic region.

[1]  K. Brock,et al.  Use of image registration and fusion algorithms and techniques in radiotherapy: Report of the AAPM Radiation Therapy Committee Task Group No. 132 , 2017, Medical physics.

[2]  Suguru Dobashi,et al.  Evaluation of various deformable image registration algorithms for thoracic images , 2013, Journal of radiation research.

[3]  Hideo Kawabata,et al.  Method of evaluating respiratory induced organ motion by vector volume histogram. , 2016, Physica medica : PM : an international journal devoted to the applications of physics to medicine and biology : official journal of the Italian Association of Biomedical Physics.

[4]  Jeffrey V Siebers,et al.  Estimation of three-dimensional intrinsic dosimetric uncertainties resulting from using deformable image registration for dose mapping. , 2011, Medical physics.

[5]  M W Vannier,et al.  Image-based dose planning of intracavitary brachytherapy: registration of serial-imaging studies using deformable anatomic templates. , 2001, International journal of radiation oncology, biology, physics.

[6]  Michael Velec,et al.  A hybrid biomechanical intensity based deformable image registration of lung 4DCT. , 2015, Physics in medicine and biology.

[7]  N. Schwenzer,et al.  A strategy for multimodal deformable image registration to integrate PET/MR into radiotherapy treatment planning , 2013, Acta oncologica.

[8]  Cristian Lorenz,et al.  Four-dimensional computed tomography pulmonary ventilation images vary with deformable image registration algorithms and metrics. , 2011, Medical physics.

[9]  Kengo Ito,et al.  Evaluation of accuracy of B-spline transformation-based deformable image registration with different parameter settings for thoracic images , 2014, Journal of radiation research.

[10]  R. Castillo,et al.  A framework for evaluation of deformable image registration spatial accuracy using large landmark point sets , 2009, Physics in medicine and biology.

[11]  Michael Velec,et al.  Effect of deformable registration uncertainty on lung SBRT dose accumulation. , 2015, Medical physics.

[12]  K. Brock Results of a multi-institution deformable registration accuracy study (MIDRAS). , 2010, International journal of radiation oncology, biology, physics.

[13]  Michael Velec,et al.  Effect of breathing motion on radiotherapy dose accumulation in the abdomen using deformable registration. , 2011, International journal of radiation oncology, biology, physics.

[14]  Rafael García-Mollá,et al.  Validation of a deformable image registration produced by a commercial treatment planning system in head and neck. , 2015, Physica medica : PM : an international journal devoted to the applications of physics to medicine and biology : official journal of the Italian Association of Biomedical Physics.

[15]  Stina Svensson,et al.  The ANACONDA algorithm for deformable image registration in radiotherapy. , 2014, Medical physics.

[16]  Thomas Guerrero,et al.  A reference dataset for deformable image registration spatial accuracy evaluation using the COPDgene study archive , 2013, Physics in medicine and biology.

[17]  Cristian Lorenz,et al.  Impact of four-dimensional computed tomography pulmonary ventilation imaging-based functional avoidance for lung cancer radiotherapy. , 2011, International journal of radiation oncology, biology, physics.

[18]  Yuki Miyabe,et al.  Multi-institutional Validation Study of Commercially Available Deformable Image Registration Software for Thoracic Images. , 2016, International journal of radiation oncology, biology, physics.