Magnetic resonance imaging with adiabatic pulses using a single surface coil for RF transmission and signal detection

In order to overcome the problems that arise from nonuniform B1, fields, there has been interest in developing pulses that are insensitive to large variations in RF power. Pulses derived from adiabatic passage principles that can execute spin inversion, excitation, and 90°and 180° plane rotations in the presence of B1, inhomogeneities have recently been described. When driven with optimized modulation functions, these pulses can execute uniform excitation, refocusing, and slice‐selective inversion over a 10‐fold or greater variation in B1 magnitude. This insensitivity to B1 strength enables the execution of T  1 − and/or T  2 − weighted spin‐echo imaging experiments using coils, such as the surface coil, with extremely inhomogeneous B1 profiles. We have successfully acquired images with these pulses at 200 MHz using a single surface coil as the transmitter and receiver. Images of the slice definition, the region over which the excitation and refocusing pulses operate with a surface coil, and brain images obtained with slice planes perpendicular to the plane of the surface coil are presented. Results demonstrate that these pulses can be transmitted with a surface coil to yield high‐quality T  1 − and/or T  2 − weighted images without B1, artifacts. © 1989 Academic Press, Inc.