Approaching Ultimate Intrinsic SNR in a Uniform Spherical Sample with Finite Arrays of Loop Coils.

We investigated to what degree and at what rate the ultimate intrinsic (UI) signal-to-noise ratio (SNR) may be approached using finite radiofrequency detector arrays. We used full-wave electromagnetic field simulations based on dyadic Green's functions to compare the SNR of arrays of loops surrounding a uniform sphere with the ultimate intrinsic SNR (UISNR), for increasing numbers of elements over a range of magnetic field strengths, voxel positions, sphere sizes, and acceleration factors. We evaluated the effect of coil conductor losses and the performance of a variety of distinct geometrical arrangements such as "helmet" and "open-pole" configurations in multiple imaging planes. Our results indicate that UISNR at the center is rapidly approached with encircling arrays and performance is substantially lower near the surface, where a quadrature detection configuration tailored to voxel position is optimal. Coil noise is negligible at high field, where sample noise dominates. Central SNR for practical array configurations such as the helmet is similar to that of close-packed arrangements. The observed trends can provide physical insights to improve coil design.

[1]  Christopher M Collins,et al.  Numerical field calculations considering the human subject for engineering and safety assurance in MRI , 2009, NMR in biomedicine.

[2]  P. Boesiger,et al.  Electrodynamics and ultimate SNR in parallel MR imaging , 2004, Magnetic resonance in medicine.

[3]  Florian Wiesinger,et al.  Parallel magnetic resonance imaging: potential and limitations at high fields , 2005 .

[4]  D. Sodickson,et al.  Ultimate intrinsic signal‐to‐noise ratio for parallel MRI: Electromagnetic field considerations , 2003, Magnetic resonance in medicine.

[5]  D. Hoult The principle of reciprocity in signal strength calculations—a mathematical guide , 2000 .

[6]  T. Ibrahim,et al.  Application of finite difference time domain method for the design of birdcage RF head coils using multi-port excitations. , 2000, Magnetic resonance imaging.

[7]  J. Polimeni,et al.  Performance evaluation of a 32‐element head array with respect to the ultimate intrinsic SNR , 2010, NMR in biomedicine.

[8]  Michael B. Smith,et al.  SAR and B1 field distributions in a heterogeneous human head model within a birdcage coil , 1998, Magnetic resonance in medicine.

[9]  Michael B. Smith,et al.  Signal‐to‐noise ratio and absorbed power as functions of main magnetic field strength, and definition of “90°” RF pulse for the head in the birdcage coil , 2001, Magnetic resonance in medicine.

[10]  Zhiyue J. Wang,et al.  Towards a complete coil array. , 2008, Magnetic resonance imaging.

[11]  H. Ermert,et al.  Ultimate signal-to-noise-ratio of surface and body antennas for magnetic resonance imaging , 2000 .

[12]  D. Sodickson,et al.  Approaching ultimate SNR and ideal current patterns with finite surface coil arrays on a dielectric cylinder , 2008 .

[13]  E. Atalar,et al.  Ultimate intrinsic signal‐to‐noise ratio in MRI , 1998, Magnetic resonance in medicine.

[14]  J. Jackson Classical Electrodynamics, 3rd Edition , 1998 .

[15]  Christina Triantafyllou,et al.  A 128‐channel receive‐only cardiac coil for highly accelerated cardiac MRI at 3 Tesla , 2008, Magnetic resonance in medicine.

[16]  J. Polimeni,et al.  96‐Channel receive‐only head coil for 3 Tesla: Design optimization and evaluation , 2009, Magnetic resonance in medicine.

[17]  C A T van den Berg,et al.  Design of a radiative surface coil array element at 7 T: The single‐side adapted dipole antenna , 2011, Magnetic resonance in medicine.

[18]  P. Roemer,et al.  The NMR phased array , 1990, Magnetic resonance in medicine.

[19]  D. Sodickson,et al.  Ideal current patterns yielding optimal signal‐to‐noise ratio and specific absorption rate in magnetic resonance imaging: Computational methods and physical insights , 2012, Magnetic resonance in medicine.

[20]  P. Boesiger,et al.  Specific coil design for SENSE: A six‐element cardiac array , 2001, Magnetic resonance in medicine.

[21]  Robin M Heidemann,et al.  Generalized autocalibrating partially parallel acquisitions (GRAPPA) , 2002, Magnetic resonance in medicine.

[22]  Riccardo Lattanzi,et al.  Coil performance evaluation based on electrodynamics : tools for hardware design and validation in magnetic resonance imaging , 2008 .

[23]  J. C. Rautio,et al.  An investigation of microstrip conductor loss , 2000 .