2D dosimetry in a proton beam with a scintillating GEM detector

A two-dimensional position-sensitive dosimetry system based on a scintillating gas detector is being developed for pre-treatment verification of dose distributions in particle therapy. The dosimetry system consists of a chamber filled with an Ar/CF(4) scintillating gas mixture, inside which two gas electron multiplier (GEM) structures are mounted (Seravalli et al 2008b Med. Phys. Biol. 53 4651-65). Photons emitted by the excited Ar/CF(4) gas molecules during the gas multiplication in the GEM holes are detected by a mirror-lens-CCD camera system. The intensity distribution of the measured light spot is proportional to the 2D dose distribution. In this work, we report on the characterization of the scintillating GEM detector in terms of those properties that are of particular importance in relative dose measurements, e.g. response reproducibility, dose dependence, dose rate dependence, spatial and time response, field size dependence, response uniformity. The experiments were performed in a 150 MeV proton beam. We found that the detector response is very stable for measurements performed in succession (sigma = 0.6%) and its response reproducibility over 2 days is about 5%. The detector response was found to be linear with the dose in the range 0.05-19 Gy. No dose rate effects were observed between 1 and 16 Gy min(-1) at the shallow depth of a water phantom and 2 and 38 Gy min(-1) at the Bragg peak depth. No field size effects were observed in the range 120-3850 mm(2). A signal rise and fall time of 2 micros was recorded and a spatial response of <or=1 mm was measured.

[1]  Leszek Ropelewski,et al.  New observations with the gas electron multiplier (GEM) , 1997 .

[2]  P. van Luijk,et al.  Collimator scatter and 2D dosimetry in small proton beams , 2001, Physics in medicine and biology.

[3]  Andrew Holmes-Siedle,et al.  Handbook of Radiation Effects , 1993 .

[4]  G. Kraft,et al.  Three-dimensional dose verification with x-ray films in conformal carbon ion therapy. , 2003, Physics in medicine and biology.

[5]  C. V. van Eijk,et al.  First Results of a Scintillating GEM Detector for 2-D Dosimetry in an Alpha Beam , 2007, IEEE Transactions on Nuclear Science.

[6]  E Pedroni,et al.  Performance of a fluorescent screen and CCD camera as a two-dimensional dosimetry system for dynamic treatment techniques. , 2000, Medical physics.

[7]  B. A. Levin,et al.  Results of irradiation tests of KU-1 and KS-4V silica glasses as ITER candidate window materials , 2003 .

[8]  J. D. Haas,et al.  Dose imaging in radiotherapy with an Ar–CF4 filled scintillating GEM , 2003 .

[9]  R. B. Murray,et al.  Scintillation Response of Activated Inorganic Crystals to Various Charged Particles , 1961 .

[10]  Sairos Safai,et al.  Development of an inorganic scintillating mixture for proton beam verification dosimetry. , 2004, Physics in medicine and biology.

[11]  J M Schippers,et al.  A scintillating gas detector for 2D dose measurements in clinical carbon beams , 2008, Physics in Medicine and Biology.

[12]  M. D. de Boer,et al.  Characterization of a scintillating GEM detector with low energy x-rays , 2008, Physics in medicine and biology.

[13]  Peter E Metcalfe,et al.  Radiochromic film for medical radiation dosimetry , 2003 .

[14]  J. Timmer,et al.  A scintillating GEM for 2D-dosimetry in radiation therapy , 2002 .

[15]  F. Sauli GEM: A new concept for electron amplification in gas detectors , 1997 .

[16]  Sjirk N. Boon,et al.  Fast 2D phantom dosimetry for scanning proton beams. , 1998, Medical physics.

[17]  G. Kraft,et al.  Experimental investigations of the response of films to heavy-ion irradiation. , 2001, Physics in medicine and biology.