A comparison of trans-tibial amputee suction and vacuum socket conditions

Daily volume loss of the stump leads to a poor fit of the prosthetic socket. A method of preventing this volume loss and maintaining a good fit was developed. A vacuum (-78 kPa) was drawn on the expulsion port of a total surface-bearing suction socket to hold the liner tightly against the socket. Stump volume of 10 trans-tibial amputees was measured prior to and immediately after a 30 minute walk with normal and vacuum socket conditions. Under the normal condition, the limb lost an average of 6.5% of its volume during the walk. In contrast, with the liner held tightly by vacuum, the limb gained an average of 3.7% in volume. It is believed that the difference observed between conditions resulted from a greater negative pressure developed during the swing phase of gait with the vacuum condition. X-rays revealed that the limb and tibia pistoned 4mm and 7mm less, respectively, under the vacuum condition. The combination of reduced pistoning and maintenance of volume is thought to account for the more symmetrical gait observed with the vacuum.

[1]  G H Koepke,et al.  Effect of liner materials on interface pressures in below-knee prostheses. , 1970, Archives of physical medicine and rehabilitation.

[2]  P K Commean,et al.  Precision of surface measurements for below-knee residua. , 1996, Archives of physical medicine and rehabilitation.

[3]  Suction socket suspension for below-knee amputees. , 1986, Archives of physical medicine and rehabilitation.

[4]  J. Sabolich,et al.  Sense of Feel for Lower‐Limb Amputees: A Phase‐One Study , 1994 .

[5]  H. Mittelstaedt,et al.  Shifts in blood volume alter the perception of posture. , 1997, International journal of psychophysiology : official journal of the International Organization of Psychophysiology.

[6]  M S Zahedi,et al.  Repeatability of kinetic and kinematic measurements in gait studies of the lower limb amputee , 1987, Prosthetics and orthotics international.

[7]  S Grevsten Ideas on the suspension of the below-knee prosthesis∗ , 1978, Prosthetics and orthotics international.

[8]  J. Dingwell,et al.  Use of an instrumented treadmill for real-time gait symmetry evaluation and feedback in normal and trans-tibial amputee subjects , 1996, Prosthetics and orthotics international.

[9]  G H Koepke,et al.  Negative pressures during swing phase in below-knee prostheses with rubber sleeve suspension. , 1975, Archives of physical medicine and rehabilitation.

[10]  D S Childress,et al.  A review of prosthetic interface stress investigations. , 1996, Journal of rehabilitation research and development.

[11]  J Mizrahi,et al.  Some gait characteristics of below-knee amputees and their reflection on the ground reaction forces. , 1986, Engineering in medicine.

[12]  L. Bennett,et al.  Shear vs pressure as causative factors in skin blood flow occlusion. , 1979, Archives of physical medicine and rehabilitation.

[13]  D. Winter,et al.  Biomechanics of below-knee amputee gait. , 1988, Journal of biomechanics.

[14]  David A. Winter,et al.  Biomechanics and Motor Control of Human Movement , 1990 .

[15]  B. Nigg,et al.  Asymmetries in ground reaction force patterns in normal human gait. , 1989, Medicine and science in sports and exercise.

[16]  Skinner Hb,et al.  Gait analysis in amputees. , 1985 .

[17]  B M Persson,et al.  A clinical standard of stump measurement and classification in lower limb amputees , 1983, Prosthetics and orthotics international.

[18]  P K Commean,et al.  Lower extremity residual limb slippage within the prosthesis. , 1997, Archives of physical medicine and rehabilitation.

[19]  P. Naylor THE SKIN SURFACE AND FRICTION. , 1955, The British journal of dermatology.

[20]  C Cheung,et al.  A microcomputer-based system for measuring temporal asymmetry in amputee gait , 1983, Prosthetics and orthotics international.

[21]  V. C. Roberts,et al.  Clinical investigation of the pressure and shear stress on the trans-tibial stump with a prosthesis. , 1998, Medical engineering & physics.

[22]  J R Pearson,et al.  Pressure variation in the below-knee, patellar tendon bearing suction socket prosthesis. , 1974, Journal of biomechanics.

[23]  Mark R Pitkin Effects of Design Variants in Lower‐Limb Prostheses on Gait Synergy , 1997, Journal of prosthetics and orthotics : JPO.

[24]  R. Wolthuis,et al.  Response of local vascular volumes to lower body negative pressure stress. , 1975, Aviation, space, and environmental medicine.

[25]  P J Holliday,et al.  Volume fluctuations in the residual limbs of lower limb amputees. , 1982, Archives of physical medicine and rehabilitation.

[26]  Tb. Staats The UCLA total surface bearing suction below-knee prosthesis , 1987 .

[27]  J. Martin Carlson,et al.  Seating Orthosis Design fop Prevention of Decubitus Ulcers , 1995 .

[28]  R L Lieber,et al.  Volume changes occurring in postoperative below-knee residual limbs. , 1988, Journal of rehabilitation research and development.

[29]  D L Bader,et al.  Effect of externally applied skin surface forces on tissue vasculature. , 1986, Archives of physical medicine and rehabilitation.

[30]  E M Burgess,et al.  Interface shear stresses during ambulation with a below-knee prosthetic limb. , 1992, Journal of rehabilitation research and development.

[31]  H. Ogata,et al.  Total surface bearing below-knee prosthesis: advantages, disadvantages, and clinical implications. , 1998, Archives of physical medicine and rehabilitation.

[32]  H. Skinner,et al.  Gait analysis in amputees. , 1985, American journal of physical medicine.

[33]  Koepke Gh,et al.  Effect of liner materials on interface pressures in below-knee prostheses. , 1970 .

[34]  V. C. Roberts,et al.  Frictional action at lower limb/prosthetic socket interface. , 1996, Medical engineering & physics.

[35]  H. Heller,et al.  Recovery from mild hypothermia can be accelerated by mechanically distending blood vessels in the hand. , 1998, Journal of applied physiology.