Evaluation of tomosynthesis elastography in a breast-mimicking phantom.

OBJECTIVE To evaluate whether measurement of strain under static compression in tomosynthesis of a breast-mimicking phantom can be used to distinguish tumor-simulating lesions of different elasticities and to compare the results to values predicted by rheometric analysis as well as results of ultrasound elastography. MATERIALS AND METHODS We prepared three soft breast-mimicking phantoms containing simulated tumors of different elasticities. The phantoms were imaged using a wide angle tomosynthesis system with increasing compression settings ranging from 0 N to 105 N in steps of 15 N. Strain of the inclusions was measured in two planes using a commercially available mammography workstation. The elasticity of the phantom matrix and inclusion material was determined by rheometric analysis. Ultrasound elastography of the inclusions was performed using two different ultrasound elastography algorithms. RESULTS Strain at maximal compression was 24.4%/24.5% in plane 1/plane 2, respectively, for the soft inclusion, 19.6%/16.9% for the intermediate inclusion, and 6.0%/10.2% for the firm inclusion. The strain ratios predicted by rheometrical testing were 0.41, 0.83 and 1.26 for the soft, intermediate, and firm inclusions, respectively. The strain ratios obtained for the soft, intermediate, and firm inclusions were 0.72±0.13, 1.02±0.21 and 2.67±1.70, respectively for tomosynthesis elastography, 0.91, 1.64 and 2.07, respectively, for ultrasound tissue strain imaging, and 0.97, 2.06 and 2.37, respectively, for ultrasound real-time elastography. CONCLUSIONS Differentiation of tumor-simulating inclusions by elasticity in a breast mimicking phantom may be possible by measuring strain in tomosynthesis. This method may be useful for assessing elasticity of breast lesions tomosynthesis of the breast.

[1]  Anders Tingberg,et al.  Breast tomosynthesis and digital mammography: a comparison of breast cancer visibility and BIRADS classification in a population of cancers with subtle mammographic findings , 2008, European Radiology.

[2]  Duane D. Meixner,et al.  Breast lesions: evaluation with US strain imaging--clinical experience of multiple observers. , 2006, Radiology.

[3]  J. Ophir,et al.  Elastography: Elasticity Imaging Using Ultrasound with Application to Muscle and Breast in Vivo , 1993, Ultrasonic imaging.

[4]  Anders Tingberg,et al.  Breast tomosynthesis: Accuracy of tumor measurement compared with digital mammography and ultrasonography , 2010, Acta radiologica.

[5]  M Fink,et al.  Imaging anisotropic and viscous properties of breast tissue by magnetic resonance‐elastography , 2005, Magnetic resonance in medicine.

[6]  Tor D Tosteson,et al.  Digital breast tomosynthesis: initial experience in 98 women with abnormal digital screening mammography. , 2007, AJR. American journal of roentgenology.

[7]  S. Zackrisson,et al.  The effect of reduced breast compression in breast tomosynthesis: human observer study using clinical cases. , 2010, Radiation protection dosimetry.

[8]  M. Hoopmann,et al.  Tissue Doppler and Strain Imaging for Evaluating Tissue Elasticity of Breast Lesions , 2007 .

[9]  Felix Diekmann,et al.  Tomosynthesis and contrast-enhanced digital mammography: recent advances in digital mammography , 2007, European Radiology.

[10]  M Heller,et al.  Breast ultrasound elastography--results of 193 breast lesions in a prospective study with histopathologic correlation. , 2011, European journal of radiology.

[11]  Andrew D. A. Maidment,et al.  Dual-energy contrast-enhanced digital breast tomosynthesis--a feasibility study. , 2010, The British journal of radiology.

[12]  R. Ehman,et al.  MR elastography of liver tumors: preliminary results. , 2008, AJR. American journal of roentgenology.

[13]  J. Baker,et al.  Can compression be reduced for breast tomosynthesis? Monte carlo study on mass and microcalcification conspicuity in tomosynthesis. , 2009, Radiology.

[14]  Jason P Fine,et al.  Differentiating Benign from Malignant Solid Breast Masses with US Strain Imaging 1 , 2007 .

[15]  J. Bamber,et al.  Ultrasound elasticity imaging: definition and technology , 1999, European Radiology.

[16]  K. Winzer,et al.  Real‐time elastography — an advanced method of ultrasound: first results in 108 patients with breast lesions , 2006, Ultrasound in obstetrics & gynecology : the official journal of the International Society of Ultrasound in Obstetrics and Gynecology.

[17]  Byung H. Han,et al.  X-ray elastography: A feasibility study , 2009, 2009 Annual International Conference of the IEEE Engineering in Medicine and Biology Society.

[18]  S. Timoshenko,et al.  Theory of elasticity , 1975 .