In vitro biomechanical analysis of three anterior thoracolumbar implants.

OBJECT The goal of this study was to evaluate the comparative efficacy of three commonly used anterior thoracolumbar implants: the anterior thoracolumbar locking plate (ATLP), the smooth-rod Kaneda (SRK), and the Z-plate. METHODS In vitro testing was performed using the T9-L3 segments of human cadaver spines. An L-1 corpectomy was performed, and stabilization was achieved using one of three anterior devices: the ATLP in nine spines, the SRK in 10, and the Z-plate in 10. Specimens were load tested with 1.5-, 3-, 4.5-, and 6-Nm in flexion and extension, right and left lateral bending, and right and left axial rotation. Angular motion was monitored using two video cameras that tracked light-emitting diodes attached to the vertebral bodies. Testing was performed in the intact state in spines stabilized with one of the three aforementioned devices after the devices had been fatigued to 5000 cycles at +/- 3 Nm and after bilateral facetectomy. There was no difference in the stability of the intact spines with use of the three devices. There were no differences between the SRK- and Z-plate-instrumented spines in any state. In extension testing, the mean angular rotation (+/- standard deviation) of spines instrumented with the SRK (4.7 +/- 3.2 degrees) and Z-plate devices (3.3 +/- 2.3 degrees) was more rigid than that observed in the ATLP-stabilized spines (9 +/- 4.8 degrees). In flexion testing after induction of fatigue, however, only the SRK (4.2 +/- 3.2 degrees) was stiffer than the ATLP (8.9 +/- 4.9 degrees). Also, in extension postfatigue, only the SRK (2.4 +/- 3.4 degrees) provided more rigid fixation than the ATLP (6.4 +/- 2.9 degrees). All three devices were equally unstable after bilateral facetectomy. The SRK and Z-plate anterior thoracolumbar implants were both more rigid than the ATLP, and of the former two the SRK was stiffer. CONCLUSIONS The authors' results suggest that in cases in which profile and ease of application are not of paramount importance, the SRK has an advantage over the other two tested implants in achieving rigid fixation immediately postoperatively.

[1]  H. An,et al.  Biomechanical Evaluation of Anterior Thoracolumbar Spinal Instrumentation , 1995, Spine.

[2]  C. Dickman,et al.  Surgical treatment alternatives for fixation of unstable fractures of the thoracic and lumbar spine. A meta-analysis. , 1994, Spine.

[3]  V. Goel,et al.  Kinematics of the Whole Lumbar Spine: Effect of Discectomy , 1985, Spine.

[4]  J. Weinstein,et al.  Biomechanics of the Spine: Clinical and Surgical Perspective , 1989 .

[5]  S. Tromanhauser,et al.  Pedicle Screw Instrumentation for Thoracolumbar Burst Fractures and Fracture‐Dislocations , 1992, Spine.

[6]  J. Torner,et al.  Biomechanical studies on two anterior thoracolumbar implants in cadaveric spines. , 1999, Spine.

[7]  E. Transfeldt,et al.  Delayed Anterior Decompression in Patients with Spinal Cord and Cauda Equina Injuries of the Thoracolumbar Spine , 1990, Spine.

[8]  B. Cunningham,et al.  Static and fatigue biomechanical properties of anterior thoracolumbar instrumentation systems. A synthetic testing model. , 1999, Spine.

[9]  P. McAfee,et al.  Anterior spinal fixators. A biomechanical in vitro study. , 1993, Spine.

[10]  R B Ashman,et al.  Effect of Spinal Construct Stiffness on Early Fusion Mass Incorporation Experimental Study , 1990, Spine.

[11]  J. Kostuik Anterior Fixation for Burst Fractures of the Thoracic and Lumbar Spine with or without Neurological Involvement , 1988, Spine.

[12]  T. Zdeblick A prospective, randomized study of lumbar fusion. Preliminary results. , 1993, Spine.

[13]  R. Chesnut,et al.  Early failure of short-segment pedicle instrumentation for thoracolumbar fractures. A preliminary report. , 1994, The Journal of bone and joint surgery. American volume.

[14]  D. Kunz,et al.  Anterior Instrumentation of the Thoracolumbar Spine: A Biomechanical Comparison , 1997, Spine.

[15]  Mcafee Pc Complications of anterior approaches to the thoracolumbar spine. Emphasis on Kaneda instrumentation. , 1994 .

[16]  P. McAfee,et al.  Anterior decompression of traumatic thoracolumbar fractures with incomplete neurological deficit using a retroperitoneal approach. , 1985, The Journal of bone and joint surgery. American volume.

[17]  J. Brantigan,et al.  The variable screw placement spinal fixation system. Report of a prospective study of 250 patients enrolled in Food and Drug Administration clinical trials. , 1993, Spine.

[18]  V K Goel,et al.  Biomechanical Studies of a Dynamized Anterior Thoracolumbar Implant , 2000, Spine.

[19]  J. Thalgott,et al.  Four year experience with the AO Anterior Thoracolumbar Locking Plate , 1997, Spinal Cord.

[20]  K. Kaneda,et al.  Burst Fractures with Neurologic Deficits of the Thoracolumbar‐Lumbar Spine Results of Anterior Decompression and Stabilization with Anterior Instrumentation , 1984, Spine.

[21]  Jae Yong Ahn,et al.  Biomechanical evaluation of anterior and posterior fixations in an unstable calf spine model. , 1997, Spine.

[22]  J. Neff,et al.  Survivorship Analysis of VSP Spine Instrumentation in the Treatment of Thoracolumbar and Lumbar Burst Fractures , 1991, Spine.

[23]  K. Kaneda,et al.  Anterior Decompression and Stabilization with the Kaneda Device for Thoracolumbar Burst Fractures Associated with Neurological Deficits*† , 1997, The Journal of bone and joint surgery. American volume.

[24]  D. Benson,et al.  Unstable thoracolumbar and lumbar burst fractures treated with the AO fixateur interne. , 1992, Journal of spinal disorders.

[25]  J N Weinstein,et al.  A Technique to Evaluate an Internal Spinal Device by Use of the Selspot System: An Application to Luque Closed Loop , 1987, Spine.

[26]  H. Dunn Anterior stabilization of thoracolumbar injuries. , 1984, Clinical orthopaedics and related research.

[27]  T. A. Zdeblick Z plate anterior thoracolumbar instrumentation , 1995 .

[28]  C. Dickman,et al.  A Historical Cohort Study of Pedicle Screw Fixation in Thoracic, Lumbar, and Sacral Spinal Fusions. , 1994, Spine.

[29]  A. Ghanayem,et al.  Anterior Instrumentation in the Management of Thoracolumbar Burst Fractures , 1997, Clinical orthopaedics and related research.

[30]  R. Gaines,et al.  Experimental Evaluation of Seven Different Spinal Fracture Internal Fixation Devices Using Nonfailure Stability Testing: The Load‐Sharing and Unstable‐Mechanism Concepts , 1991, Spine.

[31]  P. McAfee Complications of anterior approaches to the thoracolumbar spine. Emphasis on Kaneda instrumentation. , 1994, Clinical orthopaedics and related research.