Wireless Magnetoelastic Monitoring of Biliary Stents

This paper presents a system for wirelessly monitoring the accumulation of sludge in a biliary stent. Two generations of the system are detailed. The first-generation system utilizes a 2 × 37.5-mm ribbon sensor with a mass of 18 mg, along with 0.8-mm-thick × 1.6-mm-diameter neodymium magnets to bias the sensor. Both components are integrated with a 4-mm-diameter stainless steel stent. The second-generation system comprises a sensor and a magnetic layer [consisting of strontium ferrite particles suspended in polydimethylsiloxane (PDMS)] that conform to the meshed topology and tubular curvature of a 5-mm-diameter Elgiloy stent. The second-generation sensors have an active area of 7.5 × 29 mm and a mass of 9.1 mg. The sensors in both generations are fabricated from 28-¿m -thick foils of magnetoelastic 2826 MB Metglas, an amorphous Ni-Fe alloy. Analytical and finite-element models that predict sensor response in the dynamic biological environment are presented. The response of each system to viscosity changes that precede and accompany biliary sludge accumulation is tested, with resonant frequency changes of 2.8% and 6.5% over a 10-cP range for each respective generation. Sludge accumulation is simulated with successive coatings of either paraffin or an acrylate terpolymer. Resonant frequency response to this mass loading effect is similar for both generations of the system, showing a 40% decrease after applying a mass load of 2.5× the mass of the sensor.

[1]  T. O'dell Measurement of Magnetomechanical Coupling factor in Amorphous Ribbons , 1982 .

[2]  Magnetoelastic wireless sensing of tissue growth for self-expanding biliary stents , 2007, 2007 IEEE 20th International Conference on Micro Electro Mechanical Systems (MEMS).

[3]  K. Takahata,et al.  A planar approach for manufacturing cardiac stents: design, fabrication, and mechanical evaluation , 2004, Journal of Microelectromechanical Systems.

[4]  F P T Baaijens,et al.  Mechanical properties and failure of Streptococcus mutans biofilms, studied using a microindentation device. , 2006, Journal of microbiological methods.

[5]  A. Clark,et al.  A new method of magnetostrictivity and magnetostriction measurement , 1989, International Magnetics Conference.

[6]  P. Howard,et al.  Gall bladder emptying patterns in response to a normal meal in healthy subjects and patients with gall stones: ultrasound study. , 1991, Gut.

[7]  J. G. Benatar,et al.  FEM IMPLEMENTATIONS OF MAGNETOSTRICTIVE-BASED APPLICATIONS , 2005 .

[8]  T. Perng,et al.  Embrittlement of amorphous Fe40Ni38Mo4B18 alloy by electrolytic hydrogen , 1995 .

[9]  David G. Jones,et al.  Vibration and Shock in Damped Mechanical Systems , 1968 .

[10]  W. M. Haynes CRC Handbook of Chemistry and Physics , 1990 .

[11]  Emilio Guaglianone,et al.  Plastic Biliary Stent Occlusion: Factors Involved and Possible Preventive Approaches , 2007, Clinical Medicine & Research.

[12]  P. M. Anderson Magnetomechanical coupling, ΔE effect, and permeability in FeSiB and FeNiMoB alloys , 1982 .

[13]  K. Wise,et al.  Micromachined Antenna Stents and Cuffs for Monitoring Intraluminal Pressure and Flow , 2006, Journal of Microelectromechanical Systems.

[14]  D. G. Lord,et al.  Application of the Villari effect to electric power harvesting , 2006 .

[15]  C. Grimes,et al.  A remotely interrogatable magnetochemical sensor for environmental monitoring , 1998, IEEE Aerospace Conference. Proceedings.

[16]  Arthur E. Clark,et al.  Magnetomechanical coupling and permeability in transversely annealed metglas 2605 alloys , 1981 .

[17]  C A Grimes,et al.  A remote query magnetostrictive viscosity sensor. , 2000, Sensors and actuators. A, Physical.

[18]  G. Whitesides,et al.  Solvent compatibility of poly(dimethylsiloxane)-based microfluidic devices. , 2003, Analytical chemistry.

[19]  S. Dong,et al.  Giant magnetoelectric effect in Metglas/polyvinylidene-fluoride laminates , 2006 .

[20]  Craig A. Grimes,et al.  Wireless Magnetoelastic Resonance Sensors: A Critical Review , 2002 .

[21]  R. C. Weast CRC Handbook of Chemistry and Physics , 1973 .

[22]  L. K. Lagorce,et al.  Magnetic and mechanical properties of micromachined strontium ferrite/polyimide composites , 1997 .

[23]  Craig A. Grimes,et al.  A wireless micro-sensor for simultaneous measurement of pH, temperature, and pressure , 2001 .

[24]  L. Bondemark,et al.  Long-term effects of orthodontic magnets on human buccal mucosa--a clinical, histological and immunohistochemical study. , 1998, European journal of orthodontics.

[25]  M. Ladisch,et al.  Poly(dimethylsiloxane) (PDMS) and Silicon Hybrid Biochip for Bacterial Culture , 2003 .

[26]  J. Wittenberg,et al.  Biliary pressure: manometric and perfusion studies at percutaneous transhepatic cholangiography and percutaneous biliary drainage. , 1983, Radiology.

[27]  K. G. Ong,et al.  Magnetoelastic Sensors for Biomedical Monitoring , 2005 .

[28]  Y. Gianchandani,et al.  Photochemically Patterned Biliary Stents with Integrated Permanent Magnets and Deformable Assembly Features for Wireless Magnetoelastic Tissue Growth Sensing , 2007, TRANSDUCERS 2007 - 2007 International Solid-State Sensors, Actuators and Microsystems Conference.

[29]  Kefeng Zeng,et al.  Wireless Magnetoelastic Physical, Chemical, and Biological Sensors , 2007, IEEE Transactions on Magnetics.

[30]  T. Tsang,et al.  Role of bile mucin in bacterial adherence to biliary stents. , 2002, The Journal of laboratory and clinical medicine.

[31]  S. Lim,et al.  Magnetostriction measurements of metallic glass ribbon by fiber-optic Mach–Zehnder interferometry , 1999 .

[32]  Craig A. Grimes,et al.  Simultaneous measurement of liquid density and viscosity using remote query magnetoelastic sensors , 2000 .

[33]  J. Livingston,et al.  Magnetomechanical Properties of Amorphous Metals , 1982, April 16.

[34]  Y. Gianchandani,et al.  WIRELESS BILIARY STENT SYSTEM WITH WISHBONE-ARRAY RESONANT MAGNETOELASTIC ( WARM ) SENSOR AND CONFORMAL MAGNETIC LAYER , 2008 .

[35]  M. Wilson,et al.  Corrosion of intra-oral magnets in the presence and absence of biofilms of Streptococcus sanguis. , 1995, Biomaterials.

[36]  D. Jüngst,et al.  Mucin and phospholipids determine viscosity of gallbladder bile in patients with gallstones. , 2001, World journal of gastroenterology.