Quantitative tensile-tilting stages for the high voltage electron microscope.
暂无分享,去创建一个
This paper describes a soft and a hard tensile stage for the high voltage electron microscope. Both stages can be combined with a top entry double tilting stage. The devices are driven by thermal expansion elements operating against water cooled parts. The drive mechanism shows smooth action with a relatively low response time and good long-time stability. The stages are equipped with strain gauge bridges for force and elongation measurements. The soft stage has a maximum load of 13 g, and a hard one reaches 1.5 kg.
[1] T. Imura,et al. Hydraulic-Type Tensile Device for Transmission Electron Microscopy , 1972 .
[2] H. Saka,et al. Measurement of the Stress and Strain on Specimens in an Electron Microscope , 1971 .