Computational fluid and mass transport on macro and micro scales in an artificial kidney

The evaluation of dialyzer geometry was obtained by investigating transport processes and fluid properties inside the dialyzer. With computational fluid dynamics (CFD), a macroscopic as well as a microscopic model of the dialyzer was developed. Blood and dialysate flow distributions were calculated in a macroscopic numerical model of a low flux hollow fiber dialyzer. Computer simulations were compared to flow field SPECT visualizations (Single Photon Emission Computerized Tomography). The derived local fluid velocities were further implemented in a three-dimensional microscopic computer model of a single hollow fiber with its surrounding membrane and dialysate compartment. Different equations govern blood and dialysate flow (Navier-Stokes), radial filtration flow (Darcy), and solute transport (convection-diffusion). Blood was modeled as a non-Newtonian fluid with a viscosity varying both in radial and axial direction. Dialysate flow was assumed as a laminar Newtonian flow with constant viscosity. The permeability characteristics of the asymmetrical polysulphone membrane were calculated from laboratory tests for forward and backfiltration. The model was calibrated and validated in order to accurately calculate water removal as well as mass transport of small (urea) and middle molecules (vitamin B12 and inulin). Furthermore, the present microscopic CFD model was used to investigate the impact on the mass removal of dialyzer geometry and flow distributions.

[1]  P. Verdonck,et al.  Computational flow modeling in hollow-fiber dialyzers. , 2002, Artificial organs.

[2]  E. Leonard,et al.  Theoretical assessment of the effect of flow maldistributions on the mass transfer efficiency of artificial organs , 1985, Medical and Biological Engineering and Computing.

[3]  P. Gaehtgens,et al.  Flow of blood through narrow capillaries: rheological mechanisms determining capillary hematocrit and apparent viscosity. , 1980, Biorheology.

[4]  S. Eloot Experimental and numerical modeling of dialysis , 2004 .

[5]  Frank A. Gotch,et al.  Principles and Biophysics of Dialysis , 1996 .

[6]  Avinash C. Kak,et al.  Principles of computerized tomographic imaging , 2001, Classics in applied mathematics.

[7]  R. Vanholder,et al.  New insights in uremic toxins. , 2003, Kidney international. Supplement.

[8]  M. Jaffrin,et al.  Simultaneous convective and diffusive mass transfers in a hemodialyser. , 1990, Journal of biomechanical engineering.

[9]  P. Verdonck,et al.  Combining SPECT Medical Imaging and Computational Fluid Dynamics for Analyzing Blood and Dialysate Flow in Hemodialyzers , 2005, The International journal of artificial organs.

[10]  M J Lysaght,et al.  Kinetics of hemodiafiltration. I. In vitro transport characteristics of a hollow-fiber blood ultrafilter. , 1975, The Journal of laboratory and clinical medicine.

[11]  C. Ronco,et al.  Haemodialysers and Associated Devices , 1996 .

[12]  L. Henderson,et al.  Kinetics of hemodiafiltration. II. Clinical characterization of a new blood cleansing modality. , 1975, The Journal of laboratory and clinical medicine.

[13]  L. Mockros,et al.  Compact cross-flow tubular oxygenators. , 1985, Transactions - American Society for Artificial Internal Organs.

[14]  Raymond Vanholder,et al.  Review on uremic toxins: classification, concentration, and interindividual variability. , 2003, Kidney international.

[15]  P. Verdonck,et al.  In Vitro Evaluation of the Hydraulic Permeability of Polysulfone Dialysers , 2002, The International journal of artificial organs.